What should be done if the vacuum levels in the spectrometer chamber and the sample chamber of an RoHS analyzer fail to meet the specified requirements?
发布时间: 2022-12-12 10:06
RoHS detector It is a scanning instrument. During operation, many components are in motion, such as the goniometer, crystal changer, and collimator. Parts that undergo frequent movement are prone to failure, and the electronic circuit boards that control and monitor these components can also develop issues. The RoHS tester is equipped with diagnostic software; sensors distributed throughout the instrument transmit status signals to a computer, enabling operators and service engineers to determine whether the device is functioning properly and to pinpoint the location of any faults. However, certain problems that arise during measurement cannot be detected by the diagnostic software—such software merely indicates the presence of a fault. To identify the root cause, maintenance personnel must be familiar with the instrument’s architecture and possess relevant repair experience. This article presents an analysis of, and solutions for, a malfunction where the vacuum levels in the RoHS tester’s spectrometer chamber and sample chamber fail to meet specified requirements. Readers interested in this topic are encouraged to review it.
1. Fault Analysis
X-ray fluorescence spectroscopy typically operates in a vacuum optical path; however, numerous components of the spectrometer chamber and sample chamber are connected to the external environment, and many of these can develop leaks. When troubleshooting vacuum issues, categorize potential problems into three areas—vacuum pump, sample chamber, and spectrometer chamber—and systematically inspect each section to narrow down the source of the fault.
2. Solution
Remove the interface between the vacuum pump and the spectrometer chamber and the sample chamber, seal it with a rubber stopper, and then evacuate. If the specified pressure can be reached within a few seconds, the possibility of a vacuum pump malfunction can be ruled out. If it takes a long time to reach the target pressure, the pump’s efficiency may have declined. This typically occurs in instruments that frequently analyze pelletized samples and oils, where powders or oils are drawn into the pump oil, altering its viscosity. In such cases, the vacuum pump oil should be replaced.
A common source of air leakage in the sample chamber is the sealing ring on the sample‑rotation assembly. During measurements, the sample typically rotates at 0.5 rpm. After several years of operation, the sealing ring at the sample‑rotation interface wears out, leading to a deterioration in sealing performance.
A common source of air leakage in the spectrometer chamber is the flow meter, which is installed inside the chamber and connected to the external environment via inlet and outlet tubes. The flow meter’s window membrane is very thin; any leakage through this membrane can compromise the chamber’s vacuum level. Method: Connect the inlet and outlet tubes with flexible hoses to isolate the flow meter from the external atmosphere, then perform a vacuum evacuation.
Inspect for vacuum leaks. Exercise caution during disassembly and installation. Avoid allowing soot or hair to settle on the sealing ring, as this could create new leak points. When installing, apply a small amount of vacuum‑grade grease to the sealing surfaces.
The above is an overview of the fault analysis and troubleshooting methods for cases where the vacuum levels in the spectrometer chamber and sample chamber of a RoHS detector fail to meet the specified requirements. For more information, please continue browsing our website; if you would like to learn more about our products, please contact us by phone or via online consultation.
Latest News
Recommended Products
Frequently Asked Questions
What is RoHS?
RoHS is a mandatory standard established by EU legislation, officially known as the “Directive on the Restriction of the Use of Certain Hazardous Substances in Electrical and Electronic Equipment.” The standard entered into force on July 1, 2006, and primarily aims to regulate the materials and manufacturing processes of electrical and electronic products, thereby promoting human health and environmental protection.
The purpose of this standard is to eliminate six substances—lead, mercury, cadmium, hexavalent chromium, polybrominated biphenyls, and polybrominated diphenyl ethers (note: the correct Chinese term for PBDE is “polybrominated diphenyl ethers”; “polybrominated biphenyl ethers” is an incorrect designation)—and specifically stipulates that the cadmium content must not exceed 0.01%.
The latest RoHS Directive has added four new phthalate‑ester testing items: DEHP—di(2‑ethylhexyl) phthalate; BBP—butyl benzyl phthalate; DBP—dibutyl phthalate; and DIBP—diisobutyl phthalate.
What are the limit values for the ten hazardous substances under RoHS?
1. Cadmium: less than 100 ppm.
2. Lead: less than 1000 ppm.
3. Mercury: less than 1000 ppm.
4. Hexavalent chromium: less than 1000 ppm.
5. Polybrominated biphenyls (PBB): less than 1000 ppm.
6. Polybrominated diphenyl ethers (PBDEs): less than 1000 ppm.
7. Di(2-ethylhexyl) phthalate (DEHP): less than 1000 ppm.
8. Butyl benzyl phthalate (BBP): less than 1000 ppm.
9. Dibutyl phthalate (DBP): less than 1000 ppm.
10. Diisobutyl phthalate (DIBP): less than 1000 ppm.
What instruments are typically used for RoHS testing?
The commonly used analytical instrument for RoHS 1.0 is X-ray fluorescence spectroscopy (XRF). XRF analyzers are available in standalone, benchtop, and handheld configurations. Typically, a handheld XRF analyzer is employed first for portable on-site testing, with a focus on components of the product that pose a higher risk of containing restricted substances.
With the introduction of RoHS 2.0, four phthalate esters have been added to the RoHS Directive, necessitating additional testing to determine the levels of these compounds, which are extracted using solvents. The solvent extracts are then analyzed for the presence of phthalates by gas chromatography–mass spectrometry (GC/MS) or gas chromatography with flame ionization detection (GC/FID). Other elements and polybrominated diphenyl ethers/alkylated elastomers are analyzed using comparative methods and GC/MS, respectively.
Do the ten RoHS items require two instruments for testing?
The ten items specified in RoHS require two instruments for testing.
For the ten‑item RoHS testing, two instruments are required to complete the analysis; a single instrument cannot achieve this. Even if one instrument is used to measure heavy metal content, the other must be employed to analyze organic compounds.
X-ray fluorescence spectrometers are used to detect RoHS 1.0, while chromatographs, pyrolyzers, or hyphenated techniques such as GC–MS are employed to analyze the four phthalates specified in RoHS 2.0. At present, no single instrument can independently perform all the tests required for RoHS 2.0.
Can ROHS testing instruments detect metallic elements?
ROHS testing instruments are specialized devices used to detect the presence of hazardous substances in electronic components, household appliances, toys, and other products. These harmful substances can pose long-term risks to both human health and the environment, making the use of ROHS testing equipment of paramount importance.
Regarding whether ROHS testing instruments can detect metallic elements, the answer is yes. This is because ROHS testing covers not only hazardous substances but also the analysis of metallic constituents in compliant products. These metals include lead, cadmium, mercury, and others, which are widely used during the manufacturing process. Although these elements are essential, their excessive use poses significant risks to human health and the environment. Consequently, the advent of ROHS testing instruments has greatly facilitated the detection of metallic elements.