Causes of Gas-Path Contamination in Chromatographs and Methods for Diagnosis and Troubleshooting!

发布时间: 2022-11-30 10:43

In Chromatograph On the premise that contamination in the gas path has been confirmed, a series of measures have been implemented to address it. There are three primary causes of gas‑path contamination in a chromatograph: stationary phase bleed, contamination of the gas‑line tubing by impurities, and impure carrier gas. To further pinpoint the root cause of the problem, proceed according to the following troubleshooting steps:

(1) Lower the column temperature. Since the amount of stationary phase loss from the chromatographic column is exponentially related to the column temperature, reducing the temperature can significantly decrease this loss. If the baseline becomes more stable upon lowering the column temperature, it indicates that the initial column bleed was excessive and requires further optimization based on the specific analytical conditions.

(2) Whether substantial column bleed is permissible. Under the constraints of certain analytical methods, some degree of column bleed may be unavoidable; in such cases, one may consider appropriately enhancing the stability of other components of the instrument to ensure the overall analytical method remains viable.

(3) Address significant column bleed. First, suspect that the column may not be sufficiently aged; this can be confirmed by further aging the column at an elevated temperature and then checking whether the baseline improves at the operating temperature. If aging does not yield a noticeable improvement, inject several injections of distilled water at a column temperature above 150°C as a cleaning procedure (with each injection volume around 10–20 µL). After steam‑cleaning with water, if the treatment is effective, it indicates that the column is contaminated with impurities; if steam cleaning proves ineffective, consider replacing the column with a new one.

(4) Post-column gas‑line leak testing. The tubing between the chromatographic column and the thermal conductivity detector, including the detector’s own gas lines, must be leak‑free. If a leak is present, oxygen from the ambient air can ingress into the gas stream, compromising baseline stability; in severe cases, it may even corrode the tungsten filament, causing irreversible damage. The post‑column leak test is straightforward: simply block the outlet of the thermal conductivity cell and observe whether the rotor of the corresponding flow meter drops to zero.

(5) Replace the filters and purifiers. The filters and purifiers in the carrier‑gas line of the chromatograph should be activated or replaced after a certain period of use. If the carrier‑gas supply is impure, they should be replaced even more promptly. After replacing the filters and purifiers, observe any changes in baseline stability. If the baseline improves markedly, this indicates that the carrier gas is not sufficiently pure or that the filters/purifiers have lost their effectiveness.

(6) Impure carrier gas: Although a gas source of lower purity can, after passing through a high-quality filter and purifier, be used as a higher‑grade gas with reduced impurity levels, this approach shortens the service life of the filter and purifier. Moreover, the greater the impurity content of the gas source, the shorter the usable lifespan of the filtration–purification system. Therefore, the most reliable solution is to select a carrier gas of high purity and equip it with an effective filter and purifier. This ensures maximum baseline stability and extends the instrument’s normal operational life to approximately one year.

(7) Cleaning Contaminants from the Gas‑Line System. When cleaning contaminants from the gas‑line system, begin by flushing with distilled water or ethanol. To do this, heat the entire system to above 150°C, then repeatedly inject 10–20 µL of distilled water or ethanol into the injector using a syringe. After the corresponding peaks have eluted, assess baseline stability. If the baseline improves markedly, the line is likely only lightly contaminated and can remain in service; if baseline stability remains unchanged or changes only slightly, consider a thorough cleaning of the line. In the gas‑chromatography system, the inlet, the tubing connecting the column to the thermal conductivity detector (TCD), and the TCD cell itself are particularly prone to contamination, so these sections should be given special attention during cleaning.

(8) Air ingress into the detector. Even minor leaks in the post‑column gas line are the primary cause of oxygen from the ambient air entering the thermal conductivity detector. Such leaks most often occur at connection fittings and at the mounting points of the tungsten filament assembly. For methods to repair these leaks, refer to the procedures for inspecting and eliminating gas‑line leaks described earlier.

The above outlines the causes of gas‑path contamination in chromatographs, along with methods for troubleshooting and resolution. For more information, please continue browsing our website!

Latest News

Recommended Products

RoHS 2.0 Rapid Screening Instrument PY-GCMS3080

This page provides information, including frequently asked questions, about the RoHS 2.0 rapid screening instrument PY‑GCMS3080. For pricing inquiries, please contact us online or call our hotline at 19926823034!

Learn more+

RoHS 2.0 Rapid Screening Instrument PY-GCMS3080

RoHS 1.0 Rapid Testing Instrument DX-320L Pro

This page provides an introduction to the DX-320L Pro, a high-performance desktop XRF analyzer compliant with RoHS 1.0. For more information, please contact 19926823034.

Learn more+

RoHS 1.0 Rapid Testing Instrument DX-320L Pro

RoHS 1.0 Detector – IDP8300L Fluorescence Spectrometer

This page provides specifications and FAQs for the DP8300L fluorescence spectrometer. For pricing inquiries, please contact us online or call 19926823034!

Learn more+

RoHS 1.0 Detector – IDP8300L Fluorescence Spectrometer

RoHS Six-Item Detector – DX-360 Fluorescence Spectrometer

This page provides specifications and FAQs for the DX-360 fluorescence spectrometer. For pricing inquiries, please contact us online or call 19926823034!

Learn more+

RoHS Six-Item Detector – DX-360 Fluorescence Spectrometer

RoHS 2.0 Pyrolysis Gas Chromatograph PY-DP8060

This page provides specifications for the RoHS 2.0 pyrolysis gas chromatograph PY-DP8060, along with FAQs and other information. For a quote, please contact us online or call our hotline at 19926823034!

Learn more+

RoHS 2.0 Pyrolysis Gas Chromatograph PY-DP8060

Halogen detection instrument

This page provides specifications and FAQs for the IDP-6300 halogen detector. For pricing inquiries, please contact us online or call our hotline at 19926823034.

Learn more+

Halogen detection instrument

Frequently Asked Questions

What is RoHS?

RoHS is a mandatory standard established by EU legislation, officially known as the “Directive on the Restriction of the Use of Certain Hazardous Substances in Electrical and Electronic Equipment.” The standard entered into force on July 1, 2006, and primarily aims to regulate the materials and manufacturing processes of electrical and electronic products, thereby promoting human health and environmental protection.

The purpose of this standard is to eliminate six substances—lead, mercury, cadmium, hexavalent chromium, polybrominated biphenyls, and polybrominated diphenyl ethers (note: the correct Chinese term for PBDE is “polybrominated diphenyl ethers”; “polybrominated biphenyl ethers” is an incorrect designation)—and specifically stipulates that the cadmium content must not exceed 0.01%.

The latest RoHS Directive has added four new phthalate‑ester testing items: DEHP—di(2‑ethylhexyl) phthalate; BBP—butyl benzyl phthalate; DBP—dibutyl phthalate; and DIBP—diisobutyl phthalate.

What are the limit values for the ten hazardous substances under RoHS?

1. Cadmium: less than 100 ppm.

2. Lead: less than 1000 ppm.

3. Mercury: less than 1000 ppm.

4. Hexavalent chromium: less than 1000 ppm.

5. Polybrominated biphenyls (PBB): less than 1000 ppm.

6. Polybrominated diphenyl ethers (PBDEs): less than 1000 ppm.

7. Di(2-ethylhexyl) phthalate (DEHP): less than 1000 ppm.

8. Butyl benzyl phthalate (BBP): less than 1000 ppm.

9. Dibutyl phthalate (DBP): less than 1000 ppm.

10. Diisobutyl phthalate (DIBP): less than 1000 ppm.

What instruments are typically used for RoHS testing?

The commonly used analytical instrument for RoHS 1.0 is X-ray fluorescence spectroscopy (XRF). XRF analyzers are available in standalone, benchtop, and handheld configurations. Typically, a handheld XRF analyzer is employed first for portable on-site testing, with a focus on components of the product that pose a higher risk of containing restricted substances.

With the introduction of RoHS 2.0, four phthalate esters have been added to the RoHS Directive, necessitating additional testing to determine the levels of these compounds, which are extracted using solvents. The solvent extracts are then analyzed for the presence of phthalates by gas chromatography–mass spectrometry (GC/MS) or gas chromatography with flame ionization detection (GC/FID). Other elements and polybrominated diphenyl ethers/alkylated elastomers are analyzed using comparative methods and GC/MS, respectively.

Do the ten RoHS items require two instruments for testing?

The ten items specified in RoHS require two instruments for testing.
For the ten‑item RoHS testing, two instruments are required to complete the analysis; a single instrument cannot achieve this. Even if one instrument is used to measure heavy metal content, the other must be employed to analyze organic compounds.
X-ray fluorescence spectrometers are used to detect RoHS 1.0, while chromatographs, pyrolyzers, or hyphenated techniques such as GC–MS are employed to analyze the four phthalates specified in RoHS 2.0. At present, no single instrument can independently perform all the tests required for RoHS 2.0.

Can ROHS testing instruments detect metallic elements?

ROHS testing instruments are specialized devices used to detect the presence of hazardous substances in electronic components, household appliances, toys, and other products. These harmful substances can pose long-term risks to both human health and the environment, making the use of ROHS testing equipment of paramount importance.

Regarding whether ROHS testing instruments can detect metallic elements, the answer is yes. This is because ROHS testing covers not only hazardous substances but also the analysis of metallic constituents in compliant products. These metals include lead, cadmium, mercury, and others, which are widely used during the manufacturing process. Although these elements are essential, their excessive use poses significant risks to human health and the environment. Consequently, the advent of ROHS testing instruments has greatly facilitated the detection of metallic elements.

< 12 >