Causes of ICP Spectrometer Ignition Failure and Troubleshooting Methods!
发布时间: 2022-11-25 10:48
ICP spectrometer Common malfunctions fall into two main categories—ignition failure and reduced sensitivity—with ignition failure being particularly complex in its causes. Today, we’ll discuss the causes of ignition failure in ICP spectrometers and the corresponding troubleshooting methods. Ignition failure is one of the most frequent issues with ICP spectrometers, occurring when the system is unable to properly ignite the plasma.
1. Cause: Impure argon gas or low pressure.
Solution:
Replace the argon with fresh gas to ensure its purity. Note that each time you use an argon cylinder, continuously monitor the pressure gauge and avoid depleting the cylinder completely. Refilling an empty argon cylinder may introduce air, which can compromise its purity. Replace the online argon filter. The argon filter is mounted on the back of the instrument and should be replaced annually or when wavelength calibration becomes unreliable.
II. Cause: The sample fails to flow properly into the nebulizer.
Solution:
If the blockage is caused by a clogged pipeline, inspect the blocked section segment by segment, replace it with a new one, or remove the obstructed portion and reconnect the remaining sections. If the injection needle is clogged, clean the needle, connect the wider end to the syringe, and draw the dilute nitric acid into the wider end by pulling the syringe plunger in one direction; repeat this process several times. Note that during this procedure, do not push on the syringe or use ultrasonic cleaning, as these methods can easily damage the spray needle. To prevent incorrect connections in the peristaltic pump tubing, pay attention to the flow directions of the air inlet and drain lines. Before ignition, start the peristaltic pump and allow the air inlet line to fill with air for a period of time, observing the movement of the air column within the tubing—its flow should proceed from the air inlet line, through the injection needle, to the atomizer, and finally to the drain line.
III. Cause: Torch tube contamination
Solution:
After prolonged use, opaque deposits can be observed on the torch tube through the viewing window. In such cases, the torch should be replaced with a clean one. If cleaning the torch tube is necessary, take care to follow these steps: after soaking the torch tube in aqua regia for more than 48 hours, rinse it thoroughly with copious amounts of tap water and distilled water, then soak it overnight in distilled water. Rinse again with deionized water and dry it in an oven. Note that when replacing the torch tube, ensure it is completely free of moisture; otherwise, localized high temperatures at the moment of ignition may cause melting or deformation.
IV. Cause: Low cold air flow rate
Solution: Low cold‑air flow may be caused by the accumulation of carbonized particles inside the instrument. You can remove the instrument’s housing and clean out internal debris, such as the fan blades, plastic bends, and the air‑delivery aluminum tube. Note that before cleaning, be sure to disconnect the transfer hose from the aluminum tube, and take care not to touch the flow sensor.
The above is an overview of the causes of ignition failure in ICP spectrometers and the corresponding troubleshooting methods. For more information, please continue browsing our website!
Latest News
Recommended Products
Frequently Asked Questions
What is RoHS?
RoHS is a mandatory standard established by EU legislation, officially known as the “Directive on the Restriction of the Use of Certain Hazardous Substances in Electrical and Electronic Equipment.” The standard entered into force on July 1, 2006, and primarily aims to regulate the materials and manufacturing processes of electrical and electronic products, thereby promoting human health and environmental protection.
The purpose of this standard is to eliminate six substances—lead, mercury, cadmium, hexavalent chromium, polybrominated biphenyls, and polybrominated diphenyl ethers (note: the correct Chinese term for PBDE is “polybrominated diphenyl ethers”; “polybrominated biphenyl ethers” is an incorrect designation)—and specifically stipulates that the cadmium content must not exceed 0.01%.
The latest RoHS Directive has added four new phthalate‑ester testing items: DEHP—di(2‑ethylhexyl) phthalate; BBP—butyl benzyl phthalate; DBP—dibutyl phthalate; and DIBP—diisobutyl phthalate.
What are the limit values for the ten hazardous substances under RoHS?
1. Cadmium: less than 100 ppm.
2. Lead: less than 1000 ppm.
3. Mercury: less than 1000 ppm.
4. Hexavalent chromium: less than 1000 ppm.
5. Polybrominated biphenyls (PBB): less than 1000 ppm.
6. Polybrominated diphenyl ethers (PBDEs): less than 1000 ppm.
7. Di(2-ethylhexyl) phthalate (DEHP): less than 1000 ppm.
8. Butyl benzyl phthalate (BBP): less than 1000 ppm.
9. Dibutyl phthalate (DBP): less than 1000 ppm.
10. Diisobutyl phthalate (DIBP): less than 1000 ppm.
What instruments are typically used for RoHS testing?
The commonly used analytical instrument for RoHS 1.0 is X-ray fluorescence spectroscopy (XRF). XRF analyzers are available in standalone, benchtop, and handheld configurations. Typically, a handheld XRF analyzer is employed first for portable on-site testing, with a focus on components of the product that pose a higher risk of containing restricted substances.
With the introduction of RoHS 2.0, four phthalate esters have been added to the RoHS Directive, necessitating additional testing to determine the levels of these compounds, which are extracted using solvents. The solvent extracts are then analyzed for the presence of phthalates by gas chromatography–mass spectrometry (GC/MS) or gas chromatography with flame ionization detection (GC/FID). Other elements and polybrominated diphenyl ethers/alkylated elastomers are analyzed using comparative methods and GC/MS, respectively.
Do the ten RoHS items require two instruments for testing?
The ten items specified in RoHS require two instruments for testing.
For the ten‑item RoHS testing, two instruments are required to complete the analysis; a single instrument cannot achieve this. Even if one instrument is used to measure heavy metal content, the other must be employed to analyze organic compounds.
X-ray fluorescence spectrometers are used to detect RoHS 1.0, while chromatographs, pyrolyzers, or hyphenated techniques such as GC–MS are employed to analyze the four phthalates specified in RoHS 2.0. At present, no single instrument can independently perform all the tests required for RoHS 2.0.
Can ROHS testing instruments detect metallic elements?
ROHS testing instruments are specialized devices used to detect the presence of hazardous substances in electronic components, household appliances, toys, and other products. These harmful substances can pose long-term risks to both human health and the environment, making the use of ROHS testing equipment of paramount importance.
Regarding whether ROHS testing instruments can detect metallic elements, the answer is yes. This is because ROHS testing covers not only hazardous substances but also the analysis of metallic constituents in compliant products. These metals include lead, cadmium, mercury, and others, which are widely used during the manufacturing process. Although these elements are essential, their excessive use poses significant risks to human health and the environment. Consequently, the advent of ROHS testing instruments has greatly facilitated the detection of metallic elements.