Causes of peak loss in gas chromatography and their solutions!
发布时间: 2022-11-15 10:23
Gas chromatograph A chromatographic peak is the curve recorded by the detector when a substance elutes through the chromatographic column. The vertical distance between the peak’s apex and the baseline is called the peak height, typically denoted as h; the width at half‑height is referred to as the half‑peak width, usually represented by x1/2. The area enclosed by the elution curve (the chromatographic peak) and the baseline is known as the peak area, indicated by a. The time from injection until the inert carrier gas reaches its maximum concentration is called the dead time, denoted as td. The time from injection to the appearance of the chromatographic peak is termed the retention time, abbreviated as tr. The difference between the retention time and the dead time is called the corrected retention time, expressed as vb. Below, we outline the causes of peak loss in gas chromatography and corresponding solutions—please refer to this brief overview.
I. Contamination in the Gas Pathway
This can be resolved by repeatedly running the system under no-load conditions and flushing the gas pathways (including the inlet and detector).
1. To minimize contamination of the gas path, the following measures can be implemented: a high-temperature purge step should be included during the final stage of temperature programming.
2. The sample injected into the inlet should be clean;
3. Reduce the use of high-boiling-point oils;
4. Use the highest possible inlet temperature, column temperature, and detector temperature.
Second, the second scenario of peak loss occurs when the peaks fail to resolve. In addition to the aforementioned causes, this may also result from decreased column efficiency due to system contamination or from column aging; however, peak loss caused by column aging is gradual and slow.
3. False peaks are typically caused by system contamination and air leaks; the solution is to inspect for air leaks and remove any contaminants. In routine operations, a stable baseline should be recorded for reference during maintenance.
The above is an overview of the causes of peak loss in gas chromatographs and their corresponding solutions. For more information, please continue browsing. Gas Chromatography Instrument Information Section If you’d like to learn more about our products, feel free to contact us online or by phone!
Latest News
Recommended Products
Frequently Asked Questions
What is RoHS?
RoHS is a mandatory standard established by EU legislation, officially known as the “Directive on the Restriction of the Use of Certain Hazardous Substances in Electrical and Electronic Equipment.” The standard entered into force on July 1, 2006, and primarily aims to regulate the materials and manufacturing processes of electrical and electronic products, thereby promoting human health and environmental protection.
The purpose of this standard is to eliminate six substances—lead, mercury, cadmium, hexavalent chromium, polybrominated biphenyls, and polybrominated diphenyl ethers (note: the correct Chinese term for PBDE is “polybrominated diphenyl ethers”; “polybrominated biphenyl ethers” is an incorrect designation)—and specifically stipulates that the cadmium content must not exceed 0.01%.
The latest RoHS Directive has added four new phthalate‑ester testing items: DEHP—di(2‑ethylhexyl) phthalate; BBP—butyl benzyl phthalate; DBP—dibutyl phthalate; and DIBP—diisobutyl phthalate.
What are the limit values for the ten hazardous substances under RoHS?
1. Cadmium: less than 100 ppm.
2. Lead: less than 1000 ppm.
3. Mercury: less than 1000 ppm.
4. Hexavalent chromium: less than 1000 ppm.
5. Polybrominated biphenyls (PBB): less than 1000 ppm.
6. Polybrominated diphenyl ethers (PBDEs): less than 1000 ppm.
7. Di(2-ethylhexyl) phthalate (DEHP): less than 1000 ppm.
8. Butyl benzyl phthalate (BBP): less than 1000 ppm.
9. Dibutyl phthalate (DBP): less than 1000 ppm.
10. Diisobutyl phthalate (DIBP): less than 1000 ppm.
What instruments are typically used for RoHS testing?
The commonly used analytical instrument for RoHS 1.0 is X-ray fluorescence spectroscopy (XRF). XRF analyzers are available in standalone, benchtop, and handheld configurations. Typically, a handheld XRF analyzer is employed first for portable on-site testing, with a focus on components of the product that pose a higher risk of containing restricted substances.
With the introduction of RoHS 2.0, four phthalate esters have been added to the RoHS Directive, necessitating additional testing to determine the levels of these compounds, which are extracted using solvents. The solvent extracts are then analyzed for the presence of phthalates by gas chromatography–mass spectrometry (GC/MS) or gas chromatography with flame ionization detection (GC/FID). Other elements and polybrominated diphenyl ethers/alkylated elastomers are analyzed using comparative methods and GC/MS, respectively.
Do the ten RoHS items require two instruments for testing?
The ten items specified in RoHS require two instruments for testing.
For the ten‑item RoHS testing, two instruments are required to complete the analysis; a single instrument cannot achieve this. Even if one instrument is used to measure heavy metal content, the other must be employed to analyze organic compounds.
X-ray fluorescence spectrometers are used to detect RoHS 1.0, while chromatographs, pyrolyzers, or hyphenated techniques such as GC–MS are employed to analyze the four phthalates specified in RoHS 2.0. At present, no single instrument can independently perform all the tests required for RoHS 2.0.
Can ROHS testing instruments detect metallic elements?
ROHS testing instruments are specialized devices used to detect the presence of hazardous substances in electronic components, household appliances, toys, and other products. These harmful substances can pose long-term risks to both human health and the environment, making the use of ROHS testing equipment of paramount importance.
Regarding whether ROHS testing instruments can detect metallic elements, the answer is yes. This is because ROHS testing covers not only hazardous substances but also the analysis of metallic constituents in compliant products. These metals include lead, cadmium, mercury, and others, which are widely used during the manufacturing process. Although these elements are essential, their excessive use poses significant risks to human health and the environment. Consequently, the advent of ROHS testing instruments has greatly facilitated the detection of metallic elements.