The reason gas chromatographs use peak area or peak height for quantification is!
发布时间: 2022-11-14 10:44
Gas chromatograph How do we perform quantification? Naturally, we use peak area or peak height. But why can peak area or peak height be used for quantification? Those of us who work in chromatography all know that this is the standard practice, and the results are highly accurate. During my studies, my instructors always answered my questions in this way. In fact, there are two key reasons why chromatography can achieve such precise quantification.
I. Linear Response of the Detector.
We discussed this issue in our lecture on FID. Among all chromatographic detectors, with the exception of the FPD, all detectors exhibit a linear response, expressed as m = KS. Here, m denotes the amount of analyte (either mass or moles) arriving at the detector per unit time, K is the linear response coefficient, and S is the detector’s response signal. In other words, the response signal of a gas chromatography detector is directly proportional to the total amount of analyte entering the detector per unit time. Of course, we also recognize that this relationship is limited: if the analyte quantity is too large or too small, linearity will be lost.
II. Plate Theory.
The plate theory fully elucidates the relationship between peak height and injection volume, namely that they are directly proportional. This can be understood by referring to my explanation of the plate theory.
Based on these two points, it can be confidently stated that the peak height of the analyte is directly proportional to the total amount of the sample being analyzed. Alternatively, using calculus, one can infer that the peak area of the analyte is proportional to the total sample quantity. Since both peak height and peak area are proportional to the total injected amount, why do we prefer using peak area over the simpler peak height? The answer is straightforward: the plate theory is not entirely accurate, and peak shapes often deviate from a perfect normal distribution, making peak height less representative. When can peak height provide a reliable characterization? Clearly, when the peak shape is well‑defined and symmetrical, conforming to a neat normal distribution—or, more generally, when the peak is sharp and symmetric. What if chromatographic peaks frequently exhibit tailing?
It’s very simple: just quantify the peak area. In most cases, the peak area is highly representative. So, should we finally acknowledge that peak areas can be insufficiently representative? Indeed, when the GC detector is overloaded, or the analog-to-digital converter is saturated, or peak‑area integration is inaccurate, or the injection is not sufficiently representative—along with certain other special circumstances—using peak area alone will not yield correct results. Nevertheless, quantifying by peak area remains the most precise method available to us, so we have little choice but to rely on it.
Latest News
Recommended Products
Frequently Asked Questions
What is RoHS?
RoHS is a mandatory standard established by EU legislation, officially known as the “Directive on the Restriction of the Use of Certain Hazardous Substances in Electrical and Electronic Equipment.” The standard entered into force on July 1, 2006, and primarily aims to regulate the materials and manufacturing processes of electrical and electronic products, thereby promoting human health and environmental protection.
The purpose of this standard is to eliminate six substances—lead, mercury, cadmium, hexavalent chromium, polybrominated biphenyls, and polybrominated diphenyl ethers (note: the correct Chinese term for PBDE is “polybrominated diphenyl ethers”; “polybrominated biphenyl ethers” is an incorrect designation)—and specifically stipulates that the cadmium content must not exceed 0.01%.
The latest RoHS Directive has added four new phthalate‑ester testing items: DEHP—di(2‑ethylhexyl) phthalate; BBP—butyl benzyl phthalate; DBP—dibutyl phthalate; and DIBP—diisobutyl phthalate.
What are the limit values for the ten hazardous substances under RoHS?
1. Cadmium: less than 100 ppm.
2. Lead: less than 1000 ppm.
3. Mercury: less than 1000 ppm.
4. Hexavalent chromium: less than 1000 ppm.
5. Polybrominated biphenyls (PBB): less than 1000 ppm.
6. Polybrominated diphenyl ethers (PBDEs): less than 1000 ppm.
7. Di(2-ethylhexyl) phthalate (DEHP): less than 1000 ppm.
8. Butyl benzyl phthalate (BBP): less than 1000 ppm.
9. Dibutyl phthalate (DBP): less than 1000 ppm.
10. Diisobutyl phthalate (DIBP): less than 1000 ppm.
What instruments are typically used for RoHS testing?
The commonly used analytical instrument for RoHS 1.0 is X-ray fluorescence spectroscopy (XRF). XRF analyzers are available in standalone, benchtop, and handheld configurations. Typically, a handheld XRF analyzer is employed first for portable on-site testing, with a focus on components of the product that pose a higher risk of containing restricted substances.
With the introduction of RoHS 2.0, four phthalate esters have been added to the RoHS Directive, necessitating additional testing to determine the levels of these compounds, which are extracted using solvents. The solvent extracts are then analyzed for the presence of phthalates by gas chromatography–mass spectrometry (GC/MS) or gas chromatography with flame ionization detection (GC/FID). Other elements and polybrominated diphenyl ethers/alkylated elastomers are analyzed using comparative methods and GC/MS, respectively.
Do the ten RoHS items require two instruments for testing?
The ten items specified in RoHS require two instruments for testing.
For the ten‑item RoHS testing, two instruments are required to complete the analysis; a single instrument cannot achieve this. Even if one instrument is used to measure heavy metal content, the other must be employed to analyze organic compounds.
X-ray fluorescence spectrometers are used to detect RoHS 1.0, while chromatographs, pyrolyzers, or hyphenated techniques such as GC–MS are employed to analyze the four phthalates specified in RoHS 2.0. At present, no single instrument can independently perform all the tests required for RoHS 2.0.
Can ROHS testing instruments detect metallic elements?
ROHS testing instruments are specialized devices used to detect the presence of hazardous substances in electronic components, household appliances, toys, and other products. These harmful substances can pose long-term risks to both human health and the environment, making the use of ROHS testing equipment of paramount importance.
Regarding whether ROHS testing instruments can detect metallic elements, the answer is yes. This is because ROHS testing covers not only hazardous substances but also the analysis of metallic constituents in compliant products. These metals include lead, cadmium, mercury, and others, which are widely used during the manufacturing process. Although these elements are essential, their excessive use poses significant risks to human health and the environment. Consequently, the advent of ROHS testing instruments has greatly facilitated the detection of metallic elements.