Proper procedures for using gas sources in gas chromatographs!

发布时间: 2022-11-07 10:56

The gas source is in Gas chromatograph It is of paramount importance; gas is the gaseous phase. Chromatograph of the gas phase. In the gas chromatographs we use, nitrogen, hydrogen, and air are the most common; other gases such as helium and argon are also employed for specialized detectors and various analytical purposes. Each gas source significantly affects the instrument’s stability and detection sensitivity. For many users, poor baseline performance is primarily attributable to issues with the gas supply. Below, we will discuss the proper methods for using gas sources in gas chromatography, along with the specific usage, management, and preventive measures for each gas, as outlined below.

I. Nitrogen

Nitrogen is the most commonly used carrier gas and can be employed with FID, FTD, FPD, TCD, and ECD detectors. Currently, there are two main sources: cylinder‑supplied nitrogen and nitrogen generators. Cylinder‑supplied nitrogen is recommended; when using it, the following points should be kept in mind:

1. Safety

Ensure cylinders are placed securely to prevent them from falling and causing secondary accidents. A cylindrical cabinet is recommended.

2. Steel cylinder

There shall be no visible deformation. The exterior coating of the nitrogen cylinder is black, and the gas‑type lettering is yellow (Nitrogen N2). Gas labels must be clear and include the following information: gas type, hazard symbols, manufacturer, and gas purity, among others.

3. Safety Valve

There are many manufacturers of pressure-reducing valves on the market; please choose carefully. Avoid selecting valves with springs that are excessively stiff. A spring that is too hard manifests itself as follows: when adjusting the pressure‑setting knob, the knob turns only a small angle, yet the set pressure increases significantly. This indicates that the valve’s spring is overly rigid. Such valves are prone to substantial pressure fluctuations under environmental shocks or temperature variations—observable by monitoring the initial pressure in the gas chromatograph—which can lead to baseline drift. In severe cases, the inlet pressure may fluctuate noticeably, preventing the system from stabilizing. After installation, the pressure‑reducing valve should undergo a pressure‑holding test, during which no leaks should be detected. Leaks in the carrier‑gas line can cause baseline drift in the gas chromatograph.

4. Nitrogen purity (see GB/T 8980-1996)

Gas chromatographs designed for high‑sensitivity analysis require high‑purity nitrogen with a purity of 99.999%. Even when connected to high‑purity nitrogen labeled at 99.999%, it is still recommended to attach Shimadzu’s ultra‑clean carrier‑gas filter (for deoxygenation, dehydration, and hydrocarbon removal) to the system. This filter is equipped with an indicator, allowing you to clearly monitor whether the carrier‑gas filter has become ineffective; by assessing the intervals between filter failures, you can evaluate the quality of the carrier gas.

In particular, for ECD detectors, using nitrogen of a purity lower than the recommended grade can cause irreversible damage. Even when employing 99.999% high-purity nitrogen in an ECD, it is essential to install molecular sieve filters and oxygen traps, since oxygen is a strongly electronegative species; its presence in the analytical system leads to elevated baseline noise and increased background signals. At elevated temperatures, excessive oxygen levels can degrade the ECD detector and deteriorate the chromatographic column.

Therefore, when using or replacing a gas chromatograph equipped with an ECD detector for the first time, it is advisable to purge the tubing after connecting a new nitrogen cylinder, thoroughly removing any residual air. Next, properly install the carrier‑gas filter and the oxygen‑removal device to extend their service life. Then connect the gas chromatograph, power it on for the first time, and allow the carrier gas and the tail gas to flow through the column oven at a low temperature for several minutes before initiating the temperature program. For gas chromatographs equipped with other detectors, it is also recommended to follow this same connection procedure when connecting a new gas cylinder, and to use appropriate carrier‑gas filtration elements. This approach better protects the chromatographic column and reduces the system’s equilibration time.

5. Nitrogen Generator

It is not recommended to use such devices, primarily because some low‑cost nitrogen generators currently marketed as “high‑purity” struggle to maintain consistent nitrogen purity, falling short of the specifications required for chromatography. Their operating principle involves generating hydrogen with a hydrogen generator, which then reacts with oxygen in compressed air—processes that are described in the section on air. The remaining gas, predominantly nitrogen, is passed through filtration stages to remove water and hydrocarbons, after which the output is referred to as high‑purity nitrogen. However, this claimed purity has not been rigorously tested, making it difficult to meet the 99.999% purity level demanded by gas chromatographs.

II. Hydrogen

For the FID (flame ionization detector), the FPD (flame photometric detector) serves as the combustion gas, the FTD (alkali metal thermal ionization detector) functions as the reaction gas, and the TCD (thermal conductivity detector) is used as the carrier gas. There are also two alternative sources: cylinder‑supplied hydrogen and a hydrogen generator. Likewise, cylinder‑supplied hydrogen is recommended for routine use; high‑sensitivity analyses further require high‑purity hydrogen with a purity of 99.999% (see GB/T 3436.2‑2011).

Precautions for Use

1. Safety

When using hydrogen cylinders, special attention must be paid to safety. Failure to do so could result in a major disaster. For safe storage, it is recommended to use an explosion-proof cylinder cabinet. After connecting the hydrogen pressure regulator to the hydrogen pipeline, a pressure‑holding test must be conducted to check for leaks; the system may only be put into service if no leaks are detected.

2. Steel cylinder

Pay attention to the cylinder’s appearance: no obvious deformation is permitted.

The hydrogen cylinder is coated in green, and the gas‑type label is printed in red (hydrogen). When connecting the hydrogen cylinder’s pressure regulator to the pipeline, a pressure‑holding test must be performed to verify that there are no leaks. It is also recommended to use Shimadzu’s ultra‑clean filters in the hydrogen lines leading to the cylinder to remove water and hydrocarbons. If the hydrogen quality is good, this filter element will experience virtually no consumption. In cases where the hydrogen quality is suboptimal, the filter’s indicator can promptly alert you, helping to stabilize the detector’s baseline and enhance its sensitivity.

3. Hydrogen Generator

Currently, there are two types of hydrogen generators available in China: one uses an alkaline solution as the electrolyte, while the other employs a dedicated alkali‑free electrolyzer. The alkaline‑solution‑based unit is more cost‑effective, whereas the alkali‑free version is somewhat more expensive. The primary drawback of alkaline‑solution hydrogen generators is that, in the event of a malfunction, the alkaline solution can be flushed into the hydrogen gas lines. If not detected promptly, this solution may reach the hydrogen flow‑control device (APC) or even the detector, contaminating or damaging these components—resulting in far greater losses than any savings. Therefore, when operating an alkaline‑solution hydrogen generator, it is essential to regularly monitor the indicator on the hydrogen‑line filter and the color change of the desiccant silica gel inside the generator. If the silica gel changes color, replace it without delay. Should the hydrogen generator experience a fault, shut it down for repair to prevent damage to the APC and the detector. In addition, keep a close eye on the liquid level in the generator; if it drops too low, replenish it with deionized water promptly.

We recommend using an alkali-free hydrogen generator, because even if the generator malfunctions, its electrolyte can be flushed with pure water into the tubing, the APC, and the detector, making maintenance easier.

III. Air

For the FID (flame ionization detector), the FPD (flame photometric detector) serves as the combustion gas, while the FTD (alkali metal thermal ionization detector) is used as the reaction gas. Based on experience, approximately 70% of baseline noise and drift issues in these detectors are reportedly attributable to inadequate gas‑source conditioning. Therefore, proper maintenance of the gas supply is critical in routine operations.

1. Source

Similarly, the air supply can be either cylinder air or an air compressor (also referred to as an automatic air source).

2. Cylinder Air

In fact, cylinder air can be divided into two types: compressed air and proportioned air.

Compressed air is produced by compressing atmospheric air and storing it in a steel cylinder. The oxygen–nitrogen ratio of this type of air is generally acceptable, but the filling environment must be carefully monitored. For example, one company purchased compressed air from a steel manufacturer; however, the steel plant suffered from severe air pollution, with a strong sulfur dioxide odor throughout the facility. Moreover, the compressed air supplied by that manufacturer was filled directly, without any pre‑treatment such as molecular sieve filtration. As a result, the user’s FID detector exhibited significant baseline drift, and the FPD detector showed high noise levels. After the company switched to a different air source, all parameters returned to normal.

Air of this composition is sometimes referred to as high-purity air. Typically, it consists of 79% high-purity nitrogen and 21% high-purity oxygen. When this ratio is maintained, it provides an excellent gas supply—free of oil, free of water, and with minimal impurities.

However, it is important to ensure that the oxygen–nitrogen ratio is accurate. The oxygen concentration significantly affects the sensitivity of both the FID and the FPD. In severe cases, the FPD may fail to reach its peak response—indeed, this can also impact the hydrogen–air ratio in the FPD—and the TCD may even suffer damage from the alkaline source. Furthermore, some users employ high-purity oxygen as the carrier gas, causing the FID nozzle to melt under the intense combustion temperatures of the hydrogen–oxygen flame.

3. Air compressor

Air compressors are categorized into large‑scale units and small, laboratory‑grade models. Large air compressors typically serve as the primary air source for an enterprise’s power‑generation workshop, integrated into the plant’s overall infrastructure. Such systems usually include an air‑treatment unit, such as a water‑separator air filter, to ensure stable pressure and minimal contaminants, making them an ideal air supply. Special attention should be paid to the use of air compressors within laboratories. Today, the market commonly refers to these devices as “oil‑free air compressors,” but in reality, there are two distinct types. One is a truly oil‑free compressor, whose compression pistons are made from self‑lubricating materials and require no oil for lubrication. For this type, it suffices to install a water‑separator at the outlet along with appropriate desiccant filtration media (such as color‑changing silica gel). During operation, monitor the system closely and replace the filtration media regularly; under normal conditions, performance remains reliable.

There is also an air compressor. Although it is nominally labeled as an oil-free, silent compressor, the compressor it uses is actually a standard refrigerator‑or‑air‑conditioner compressor. Its piston motor and other mechanical components are immersed in antifreeze lubricant, which inevitably generates oil mist during operation, compromising the quality of the compressed air. To address this, the unit is equipped with three large filters—color‑changing silica gel, activated carbon, and molecular sieve—at the outlet to purify the compressed air.

However, the purification unit will inevitably become saturated; even if it does not reach saturation, its adsorption capacity will gradually decline over time, leading to a reduction in air purity. When using an air analyzer, various baseline issues may arise. A common scenario is that, when the compressor fails to start, the gas storage tank slowly vents, causing the tank pressure to drop gradually while the baseline rises (or falls) at a relatively shallow slope. Once the air has been vented down to the compressor’s starting pressure, the compressor begins charging the tank, resulting in a rapid pressure increase and a much steeper baseline slope—indicating a swift rise (or fall). Consequently, the baseline fluctuates systematically with each compressor start-up. To eliminate this issue, the only effective solution is to thoroughly clean the entire compressor system, including the gas storage tank.

Latest News

Recommended Products

RoHS 2.0 Rapid Screening Instrument PY-GCMS3080

This page provides information, including frequently asked questions, about the RoHS 2.0 rapid screening instrument PY‑GCMS3080. For pricing inquiries, please contact us online or call our hotline at 19926823034!

Learn more+

RoHS 2.0 Rapid Screening Instrument PY-GCMS3080

RoHS 1.0 Rapid Testing Instrument DX-320L Pro

This page provides an introduction to the DX-320L Pro, a high-performance desktop XRF analyzer compliant with RoHS 1.0. For more information, please contact 19926823034.

Learn more+

RoHS 1.0 Rapid Testing Instrument DX-320L Pro

RoHS 1.0 Detector – IDP8300L Fluorescence Spectrometer

This page provides specifications and FAQs for the DP8300L fluorescence spectrometer. For pricing inquiries, please contact us online or call 19926823034!

Learn more+

RoHS 1.0 Detector – IDP8300L Fluorescence Spectrometer

RoHS Six-Item Detector – DX-360 Fluorescence Spectrometer

This page provides specifications and FAQs for the DX-360 fluorescence spectrometer. For pricing inquiries, please contact us online or call 19926823034!

Learn more+

RoHS Six-Item Detector – DX-360 Fluorescence Spectrometer

RoHS 2.0 Pyrolysis Gas Chromatograph PY-DP8060

This page provides specifications for the RoHS 2.0 pyrolysis gas chromatograph PY-DP8060, along with FAQs and other information. For a quote, please contact us online or call our hotline at 19926823034!

Learn more+

RoHS 2.0 Pyrolysis Gas Chromatograph PY-DP8060

Halogen detection instrument

This page provides specifications and FAQs for the IDP-6300 halogen detector. For pricing inquiries, please contact us online or call our hotline at 19926823034.

Learn more+

Halogen detection instrument

Frequently Asked Questions

What is RoHS?

RoHS is a mandatory standard established by EU legislation, officially known as the “Directive on the Restriction of the Use of Certain Hazardous Substances in Electrical and Electronic Equipment.” The standard entered into force on July 1, 2006, and primarily aims to regulate the materials and manufacturing processes of electrical and electronic products, thereby promoting human health and environmental protection.

The purpose of this standard is to eliminate six substances—lead, mercury, cadmium, hexavalent chromium, polybrominated biphenyls, and polybrominated diphenyl ethers (note: the correct Chinese term for PBDE is “polybrominated diphenyl ethers”; “polybrominated biphenyl ethers” is an incorrect designation)—and specifically stipulates that the cadmium content must not exceed 0.01%.

The latest RoHS Directive has added four new phthalate‑ester testing items: DEHP—di(2‑ethylhexyl) phthalate; BBP—butyl benzyl phthalate; DBP—dibutyl phthalate; and DIBP—diisobutyl phthalate.

What are the limit values for the ten hazardous substances under RoHS?

1. Cadmium: less than 100 ppm.

2. Lead: less than 1000 ppm.

3. Mercury: less than 1000 ppm.

4. Hexavalent chromium: less than 1000 ppm.

5. Polybrominated biphenyls (PBB): less than 1000 ppm.

6. Polybrominated diphenyl ethers (PBDEs): less than 1000 ppm.

7. Di(2-ethylhexyl) phthalate (DEHP): less than 1000 ppm.

8. Butyl benzyl phthalate (BBP): less than 1000 ppm.

9. Dibutyl phthalate (DBP): less than 1000 ppm.

10. Diisobutyl phthalate (DIBP): less than 1000 ppm.

What instruments are typically used for RoHS testing?

The commonly used analytical instrument for RoHS 1.0 is X-ray fluorescence spectroscopy (XRF). XRF analyzers are available in standalone, benchtop, and handheld configurations. Typically, a handheld XRF analyzer is employed first for portable on-site testing, with a focus on components of the product that pose a higher risk of containing restricted substances.

With the introduction of RoHS 2.0, four phthalate esters have been added to the RoHS Directive, necessitating additional testing to determine the levels of these compounds, which are extracted using solvents. The solvent extracts are then analyzed for the presence of phthalates by gas chromatography–mass spectrometry (GC/MS) or gas chromatography with flame ionization detection (GC/FID). Other elements and polybrominated diphenyl ethers/alkylated elastomers are analyzed using comparative methods and GC/MS, respectively.

Do the ten RoHS items require two instruments for testing?

The ten items specified in RoHS require two instruments for testing.
For the ten‑item RoHS testing, two instruments are required to complete the analysis; a single instrument cannot achieve this. Even if one instrument is used to measure heavy metal content, the other must be employed to analyze organic compounds.
X-ray fluorescence spectrometers are used to detect RoHS 1.0, while chromatographs, pyrolyzers, or hyphenated techniques such as GC–MS are employed to analyze the four phthalates specified in RoHS 2.0. At present, no single instrument can independently perform all the tests required for RoHS 2.0.

Can ROHS testing instruments detect metallic elements?

ROHS testing instruments are specialized devices used to detect the presence of hazardous substances in electronic components, household appliances, toys, and other products. These harmful substances can pose long-term risks to both human health and the environment, making the use of ROHS testing equipment of paramount importance.

Regarding whether ROHS testing instruments can detect metallic elements, the answer is yes. This is because ROHS testing covers not only hazardous substances but also the analysis of metallic constituents in compliant products. These metals include lead, cadmium, mercury, and others, which are widely used during the manufacturing process. Although these elements are essential, their excessive use poses significant risks to human health and the environment. Consequently, the advent of ROHS testing instruments has greatly facilitated the detection of metallic elements.

< 12 >