How does a gas chromatograph determine the phosphorus content?
发布时间: 2022-11-02 11:05
Phosphorus is a common element, accounting for approximately 0.118% by weight in the Earth’s crust. In nature, it occurs primarily in the form of various phosphate compounds. Phosphorus is present in cells, bones, and teeth, serving as an essential component for animals, plants, and humans. A healthy individual requires about 1.4 grams of phosphorus daily, which is absorbed in the form of diverse inorganic phosphates or organic phosphorus compounds. When phosphorus exists as elemental phosphorus in water and wastewater, it can pose environmental hazards. Therefore, it is necessary to employ analytical instruments to detect its concentration; thus, Gas chromatograph For example, let’s now discuss how a gas chromatograph is used to determine the phosphorus content!
I. Method Selection
Gas chromatography can be used to directly determine elemental phosphorus in water samples extracted with organic solvents. This method is sensitive, highly reproducible, simple, and rapid, and is unaffected by various waterborne contaminants.
II. Collection and Preservation of Samples
After collecting the water sample, perform on-site extraction with an organic solvent. Slowly transfer the extract into a stoppered colorimetric tube or volumetric flask and seal it tightly, ensuring no air gaps remain. The sample should be analyzed as soon as possible. If immediate analysis is not feasible, store it in a refrigerator at approximately 4°C for up to 7 days.
III. Gas Chromatography Analysis
1. Methodological Principles
Toluene is used as the extractant to extract elemental phosphorus from water samples. After separation on a chromatographic column, the elemental phosphorus in the extract is oxidized and combusted in a flame photometric detector (FPD) to form phosphorus oxides, which are then reduced by a hydrogen-rich flame to PHO*—an excited‑state PHO fragment. The high‑temperature flame excites the PHO* fragment, causing it to emit characteristic spectral radiation with a peak detection wavelength of 526 nm. The concentration of elemental phosphorus is determined by measuring the intensity of this emission spectrum.
2. Interference and Elimination
Inorganic substances in the water sample, common organophosphorus pesticides (such as dimethoate, methyl parathion, and malathion), and other organic compounds do not interfere with the determination of elemental phosphorus.
3. Scope of Application of the Method
The method achieves a detection limit of 0.25 μg/L (based on an extraction ratio of 2:1). When the water-to-extraction‑solvent ratio is 25:1, the detection limit drops to 0.02 μg/L. This method is suitable for determining elemental phosphorus in industrial wastewater discharged from yellow‑phosphorus production facilities and in surface waters contaminated with elemental phosphorus.
4. Tools
① Gas chromatograph: equipped with a flame photometric detector (FPD) and a fluorescence filter at 526 nm ± 2 nm; ② 60 mL and 500 mL separatory funnels; ③ 1 μL microsyringe; ④ 5 mL volumetric flask.
5. Reagents
① Refined yellow phosphorus (purity 99.99%); ② Toluene: analytical grade; ③ Anhydrous ethanol: analytical grade; ④ Chromatographic stationary phase; ⑤ Chromatographic support; ⑥ Elemental phosphorus standard stock solution; ⑦ Elemental phosphorus standard solution (each milliliter of the standard solution should contain approximately 10 μg of elemental phosphorus); ⑧ Elemental phosphorus standard solution for spike addition.
6. Operating Procedure
(1) Sample Pre-treatment
① The extraction of the water sample must be performed at the sampling site. Accurately weigh 20 mL of the water sample to be analyzed into a 60-mL separatory funnel, add 10.0 mL of toluene, seal the funnel tightly, and shake vigorously for 5 minutes. Allow the mixture to stand for 5–10 minutes; after phase separation, discard the aqueous layer. Transfer the upper toluene layer to a 5-mL volumetric flask and fill to the mark, ensuring no air bubbles remain. The extract is then brought back to the laboratory for direct determination by gas chromatography.
② When the concentration of elemental phosphorus in the water sample is less than 0.001 mg/L, the sampling volume may be increased, and the ratio of water sample to extractant adjusted, until the sampling volume reaches 250 mL.
(2) Chromatographic conditions
① Chromatographic column: 10% SE-30/Chromosorb W (AW-DMCS), 60–80 mesh, packed in a glass column 2 m long with an inner diameter of 3 mm.
② Carrier gas: high-purity nitrogen at a flow rate of 40 mL/min.
③ Hydrogen: high-purity hydrogen at a flow rate of 53 mL/min.
④ Air: flow rate of 83 mL/min.
⑤ Temperature: column temperature 150°C; vaporization chamber temperature 200°C; detector temperature 230°C.
⑥ Recording paper speed: 5 mm/min.
⑦ Sample volume: 1 µL
(3) Calibration curve
① Take seven 25 mL volumetric flasks and prepare Standard Series I or Standard Series II according to the values listed in Table 1, then dilute each to the mark with toluene. Use a temporary aliquotting procedure. The standard series in Table 1 are prepared using a 10 μg/mL elemental phosphorus standard solution as an example. Based on the standard series prepared as shown in Table 1, the actual concentrations of elemental phosphorus in the prepared standards should be close to the values indicated in the table. Under the chromatographic conditions, perform chromatographic analysis in ascending order of concentration, starting from the most dilute standard solution; all standard samples are analyzed in parallel. After blank correction, obtain the linear relationship between the elemental phosphorus concentration and the measured peak heights, and construct the calibration curve. Depending on the concentration level of the sample being analyzed, select an appropriate concentration series to plot the calibration curve.
② To maximize the accuracy of elemental phosphorus determination in water samples, the preparation of the standard series should follow the same extraction procedure as that used for the water samples. In this case, the working standard solution is prepared by diluting the stock standard solution with anhydrous ethanol to ensure complete miscibility with the aqueous phase. Seven 25 mL volumetric flasks are used; according to Table 1, appropriate volumes of the standard solution are transferred into each flask and diluted to the mark with distilled water to prepare the standard series. Sequentially, 10 mL of each standard from the series is transferred into a 60 mL separatory funnel, followed by the addition of 10 mL of analytical‑grade toluene, and the stopper is securely tightened. The mixture is vigorously shaken for 5 minutes, allowed to stand for 5–10 minutes, and the aqueous phase is discarded. The extractant is then analyzed by gas chromatography, and a calibration curve is constructed.
(4) Sample Determination
Analyze the samples under the chromatographic conditions and determine each sample in parallel.
(5) Standard chromatogram of elemental phosphorus
7. Calculation
Based on the average chromatographic peak height measured for the sample, determine the corresponding concentration from the calibration curve, and calculate the concentration of elemental phosphorus in the water sample using the following equation.
In the equation: C1 — injection concentration (µg/L or mg/L); K — sample extraction and concentration factor.
You can also compare the average chromatographic peak heights of the sample and the standard series to determine the concentration of elemental phosphorus in the sample. In this case, it is important to perform comparative calculations using the peak height measurements of standard samples from the standard series that are similar to the sample’s measured peak height, in order to minimize errors. The concentration of elemental phosphorus in the sample can then be calculated according to the following formula.
In the equation: h1 — measured peak height of the sample (in mm or μV, etc.); C — concentration of the standard samples in the calibration series (in μg/L or mg/L); H — measured peak height of the standard samples in the calibration series (in mm or μV, etc.); K — extraction and concentration factor of the sample.
When the ratio of water sample to extractant is 2:1, the extraction concentration factor k = 2; when the ratio is 4:1, k = 4; and when the ratio is 25:1, k = 25.
8. Precision and Accuracy
This method was applied to six determinations of a contaminated water sample containing 2.1 μg/L phosphorus, yielding a relative standard deviation of 4.3%. At an added concentration of 1.0 μg/L, the recovery rate was 92%.
The above is an introduction to how a gas chromatograph determines phosphorus content. If you would like to learn more, please continue browsing. Gas Chromatography Instrument Information Section If you’d like to learn more about the product, feel free to contact us online or by phone!
Latest News
Recommended Products
Frequently Asked Questions
What is RoHS?
RoHS is a mandatory standard established by EU legislation, officially known as the “Directive on the Restriction of the Use of Certain Hazardous Substances in Electrical and Electronic Equipment.” The standard entered into force on July 1, 2006, and primarily aims to regulate the materials and manufacturing processes of electrical and electronic products, thereby promoting human health and environmental protection.
The purpose of this standard is to eliminate six substances—lead, mercury, cadmium, hexavalent chromium, polybrominated biphenyls, and polybrominated diphenyl ethers (note: the correct Chinese term for PBDE is “polybrominated diphenyl ethers”; “polybrominated biphenyl ethers” is an incorrect designation)—and specifically stipulates that the cadmium content must not exceed 0.01%.
The latest RoHS Directive has added four new phthalate‑ester testing items: DEHP—di(2‑ethylhexyl) phthalate; BBP—butyl benzyl phthalate; DBP—dibutyl phthalate; and DIBP—diisobutyl phthalate.
What are the limit values for the ten hazardous substances under RoHS?
1. Cadmium: less than 100 ppm.
2. Lead: less than 1000 ppm.
3. Mercury: less than 1000 ppm.
4. Hexavalent chromium: less than 1000 ppm.
5. Polybrominated biphenyls (PBB): less than 1000 ppm.
6. Polybrominated diphenyl ethers (PBDEs): less than 1000 ppm.
7. Di(2-ethylhexyl) phthalate (DEHP): less than 1000 ppm.
8. Butyl benzyl phthalate (BBP): less than 1000 ppm.
9. Dibutyl phthalate (DBP): less than 1000 ppm.
10. Diisobutyl phthalate (DIBP): less than 1000 ppm.
What instruments are typically used for RoHS testing?
The commonly used analytical instrument for RoHS 1.0 is X-ray fluorescence spectroscopy (XRF). XRF analyzers are available in standalone, benchtop, and handheld configurations. Typically, a handheld XRF analyzer is employed first for portable on-site testing, with a focus on components of the product that pose a higher risk of containing restricted substances.
With the introduction of RoHS 2.0, four phthalate esters have been added to the RoHS Directive, necessitating additional testing to determine the levels of these compounds, which are extracted using solvents. The solvent extracts are then analyzed for the presence of phthalates by gas chromatography–mass spectrometry (GC/MS) or gas chromatography with flame ionization detection (GC/FID). Other elements and polybrominated diphenyl ethers/alkylated elastomers are analyzed using comparative methods and GC/MS, respectively.
Do the ten RoHS items require two instruments for testing?
The ten items specified in RoHS require two instruments for testing.
For the ten‑item RoHS testing, two instruments are required to complete the analysis; a single instrument cannot achieve this. Even if one instrument is used to measure heavy metal content, the other must be employed to analyze organic compounds.
X-ray fluorescence spectrometers are used to detect RoHS 1.0, while chromatographs, pyrolyzers, or hyphenated techniques such as GC–MS are employed to analyze the four phthalates specified in RoHS 2.0. At present, no single instrument can independently perform all the tests required for RoHS 2.0.
Can ROHS testing instruments detect metallic elements?
ROHS testing instruments are specialized devices used to detect the presence of hazardous substances in electronic components, household appliances, toys, and other products. These harmful substances can pose long-term risks to both human health and the environment, making the use of ROHS testing equipment of paramount importance.
Regarding whether ROHS testing instruments can detect metallic elements, the answer is yes. This is because ROHS testing covers not only hazardous substances but also the analysis of metallic constituents in compliant products. These metals include lead, cadmium, mercury, and others, which are widely used during the manufacturing process. Although these elements are essential, their excessive use poses significant risks to human health and the environment. Consequently, the advent of ROHS testing instruments has greatly facilitated the detection of metallic elements.