How do you set up a fluorescence spectrometer?
发布时间: 2022-10-26 10:21
Fluorescence spectrometer It is an instrument used to scan the fluorescence spectra emitted by liquid fluorescent labels. It can provide numerous physical parameters, including excitation spectra, emission spectra, fluorescence intensity, quantum yield, fluorescence lifetime, and fluorescence polarization, offering a comprehensive view of molecular bonding and structure. So, how do you set up a fluorescence spectrometer? Let’s find out!
1. Selection of Excitation Wavelength
How do you determine the excitation and emission wavelengths? For many beginners in spectroscopy, this can be a confusing question. If the instrument is equipped with three-dimensional scanning capabilities, the process is relatively straightforward: place the sample inside, follow the instructions to perform a three‑dimensional fluorescence scan, and you can easily identify the appropriate excitation and emission wavelengths. If the instrument lacks 3D scanning, the standard approach is as follows.
First, set the instrument’s excitation wavelength (λem) to 200 nm, then perform an emission (EM) mode scan, with the emission wavelength (λEM) scanned over a range of 210–800 nm, and record all peak wavelengths. Next, adjust the excitation wavelength (λem) and repeat the scan. If, in the second emission spectrum, the position of a particular peak (or multiple peaks) remains unchanged (or shifts only slightly), that peak (or those peaks) is generally a fluorescence peak, since the position of a fluorescence peak does not vary with changes in the excitation wavelength; only the peak height (or peak area) will change.
Next, fix the wavelength of the identified fluorescence peak as the emission wavelength (λem), and then scan the excitation wavelength (Ex) over a range narrower than the emission wavelength. If only a single excitation peak is present, this approach is straightforward; once that wavelength is set, performing another scan at the true emission wavelength (λem) will yield an excellent signal-to-noise ratio. If several peaks emerge upon scanning the excitation wavelength (Ex), select the appropriate one based on experience—typically choosing the peak with a well-defined shape, high suitability, and a reasonable bandwidth as the excitation wavelength. Alternatively, the excitation wavelength can be determined by leveraging the characteristic relationship between excitation and emission spectra: the excitation spectrum is the mirror image of the emission spectrum.
2. Filter Selection
When scanning and analyzing the emission spectrum, according to Stokes’ law, the excitation wavelength is set to be shorter than the emission wavelength. For example, if the excitation wavelength λex = 260 nm, the emission should begin at a wavelength longer than 260 nm—such as 270 nm—and end at 870 nm. In this way, sharp harmonic peaks corresponding to twice (520 nm) and three times (780 nm) the excitation wavelength will appear. If these overlap with the fluorescence spectral peaks, they can interfere with the measurement; therefore, appropriate filters must be selected to eliminate them.
The second-harmonic peak is a scattering peak of the excitation light. Although it appears at twice the excitation wavelength during scanning, its actual wavelength corresponds to the excitation wavelength. The second-harmonic peak can be eliminated by inserting a long‑pass filter. Typically, such a filter is placed at the emission path, and the chip is marked with the maximum wavelength it can transmit. In other words, once a long‑pass filter is incorporated into the system, any light with a wavelength shorter than the marked cutoff—including both second‑harmonic and third‑harmonic components—will be blocked, while light with wavelengths longer than the specified cutoff will pass through unimpeded.
3. Configuration of the Slit
Fluorescence intensity depends on numerous factors, including molecular structure, the state of the analyte, solution concentration or film thickness, temperature, the selected excitation wavelength, and the width of the measurement slit. Fluorescence intensity values obtained with different instruments can vary significantly and are therefore not directly comparable; however, the position of the emission peak is relatively easy to identify and can be used to determine the identity of the substance. Quantitative fluorescence analysis requires that certain experimental conditions be kept constant.
The slit width is critically important for fluorescence intensity. It can be adjusted appropriately based on the fluorescence intensity of the sample being measured. At a given concentration, if the fluorescence signal is too strong and exceeds the instrument’s dynamic range, the slit can be narrowed. However, it should be noted that changes in fluorescence intensity depend not only on slit width but also on spectral resolution. For fluorescence scanning, increasing the slit width enhances both fluorescence intensity and reproducibility, but at the expense of resolution. Once the instrument’s other optical components—such as grating groove density, focal length, and overall linear dispersion—are fixed, slit width becomes the primary determinant of resolution. In practical applications, the slit size should be matched to the full width at half maximum (FWHM) of the emission peak; an excessively wide slit will broaden and distort the peak. Slit width also governs the light throughput: empirically, doubling the slit width increases light throughput by a factor of four.
4. Step Length Setting
The smaller the measurement step size, the slower the measurement and the more detailed the map; the larger the step size, the faster the measurement and the coarser the spectrum. The step size should be less than one-fifth of the smallest FWHM to be measured, as it is related to the resolution. In principle, the step size should be no greater than three times the slit width.
The above is an introduction to how to set up a fluorescence spectrometer. If you would like to learn more, please continue browsing. Fluorescence Spectrometer News Section If you’d like to learn more about our products, feel free to contact us online or by phone!
Latest News
Recommended Products
Frequently Asked Questions
What is RoHS?
RoHS is a mandatory standard established by EU legislation, officially known as the “Directive on the Restriction of the Use of Certain Hazardous Substances in Electrical and Electronic Equipment.” The standard entered into force on July 1, 2006, and primarily aims to regulate the materials and manufacturing processes of electrical and electronic products, thereby promoting human health and environmental protection.
The purpose of this standard is to eliminate six substances—lead, mercury, cadmium, hexavalent chromium, polybrominated biphenyls, and polybrominated diphenyl ethers (note: the correct Chinese term for PBDE is “polybrominated diphenyl ethers”; “polybrominated biphenyl ethers” is an incorrect designation)—and specifically stipulates that the cadmium content must not exceed 0.01%.
The latest RoHS Directive has added four new phthalate‑ester testing items: DEHP—di(2‑ethylhexyl) phthalate; BBP—butyl benzyl phthalate; DBP—dibutyl phthalate; and DIBP—diisobutyl phthalate.
What are the limit values for the ten hazardous substances under RoHS?
1. Cadmium: less than 100 ppm.
2. Lead: less than 1000 ppm.
3. Mercury: less than 1000 ppm.
4. Hexavalent chromium: less than 1000 ppm.
5. Polybrominated biphenyls (PBB): less than 1000 ppm.
6. Polybrominated diphenyl ethers (PBDEs): less than 1000 ppm.
7. Di(2-ethylhexyl) phthalate (DEHP): less than 1000 ppm.
8. Butyl benzyl phthalate (BBP): less than 1000 ppm.
9. Dibutyl phthalate (DBP): less than 1000 ppm.
10. Diisobutyl phthalate (DIBP): less than 1000 ppm.
What instruments are typically used for RoHS testing?
The commonly used analytical instrument for RoHS 1.0 is X-ray fluorescence spectroscopy (XRF). XRF analyzers are available in standalone, benchtop, and handheld configurations. Typically, a handheld XRF analyzer is employed first for portable on-site testing, with a focus on components of the product that pose a higher risk of containing restricted substances.
With the introduction of RoHS 2.0, four phthalate esters have been added to the RoHS Directive, necessitating additional testing to determine the levels of these compounds, which are extracted using solvents. The solvent extracts are then analyzed for the presence of phthalates by gas chromatography–mass spectrometry (GC/MS) or gas chromatography with flame ionization detection (GC/FID). Other elements and polybrominated diphenyl ethers/alkylated elastomers are analyzed using comparative methods and GC/MS, respectively.
Do the ten RoHS items require two instruments for testing?
The ten items specified in RoHS require two instruments for testing.
For the ten‑item RoHS testing, two instruments are required to complete the analysis; a single instrument cannot achieve this. Even if one instrument is used to measure heavy metal content, the other must be employed to analyze organic compounds.
X-ray fluorescence spectrometers are used to detect RoHS 1.0, while chromatographs, pyrolyzers, or hyphenated techniques such as GC–MS are employed to analyze the four phthalates specified in RoHS 2.0. At present, no single instrument can independently perform all the tests required for RoHS 2.0.
Can ROHS testing instruments detect metallic elements?
ROHS testing instruments are specialized devices used to detect the presence of hazardous substances in electronic components, household appliances, toys, and other products. These harmful substances can pose long-term risks to both human health and the environment, making the use of ROHS testing equipment of paramount importance.
Regarding whether ROHS testing instruments can detect metallic elements, the answer is yes. This is because ROHS testing covers not only hazardous substances but also the analysis of metallic constituents in compliant products. These metals include lead, cadmium, mercury, and others, which are widely used during the manufacturing process. Although these elements are essential, their excessive use poses significant risks to human health and the environment. Consequently, the advent of ROHS testing instruments has greatly facilitated the detection of metallic elements.