ROHS Analyzer Verification Procedure!
发布时间: 2022-10-26 10:13
RoHS analyzer It is an instrument used to detect compliance with the RoHS Directive and is currently widely employed across multiple industries. Recently, many have been inquiring about the verification procedures for RoHS analyzers. Taking X-ray fluorescence spectrometers as an example, we will provide a brief explanation below.
I. Visual Inspection
1. The instrument shall be marked with its name, manufacturer, date, and serial number.
2. All components are securely connected and functioning properly.
3. The instruments, indicator lights, and safety protection devices on the panel are functioning properly.
II. Laboratory Calibration Conditions
Power supply: Three-phase or single-phase, 220 V, with voltage fluctuations not exceeding (+10%).
Grounding: The resistance of a single ground electrode is less than 0.
Cooling water: water temperature below 30°C, water pressure greater than 9.8 × 10⁴ Pa/cm², flow rate greater than 41 L/min.
Room temperature: (15–28)°C ± 3°C.
Humidity: Relative humidity less than 75%
Note: Different types of instruments have varying requirements for laboratory operating conditions. Specific requirements may be specified by the instrument manufacturer.
3. Prior to instrument calibration, allow the device to warm up at the operating power level for at least 2 hours.
For calibration purposes:
a) Pure copper or brass disc
b) Pure aluminum block
c) Chromium-nickel stainless steel round bar
Note: Other calibration samples may be prepared in accordance with the specific requirements of the instrument under test.
IV. Accuracy Calibration
Accuracy is expressed as the relative standard deviation (RSD) of 20 consecutive replicate measurements. Each measurement must be performed under varying instrumental conditions, including the crystal, counter, collimator, diaphragm, filters, attenuator, and sample‑stage position. If, among the 20 consecutive measurements, any data point exceeds the mean by more than +3σ, the measurement should be repeated.
Accuracy verification shall be performed under the following conditions. Measurement condition 1: Use a pure copper or brass sample to measure copper—count rate or numerical value, low‑frequency crystal, precision collimator, no filter, no attenuator, scintillation counter, vacuum optical path, and a counting time of 10 seconds. Measurement condition 2: Use a pure aluminum sample to measure Al K—count rate or numerical value, PET crystal, coarse collimator, with filter and attenuator, flow‑gas proportional counter, vacuum optical path, and a counting time of 1 or 2 seconds.
Set the line voltage to 40 kV or 50 kV. Adjust the current so that the CUK is in state 1. Set the count rate to 100–200 kcps. Alternate between conditions 1 and 2, performing 20 cycles for each condition, with one measurement per copper sample. The sample positions on the sample stage must be changed. If fewer than 20 samples are available on the carousel, they may be reused. If the instrument is equipped with a sample‑rotation device, rotate the samples. Calculate the CUK. Determine the RSD over 20 measurements.
This test is used to evaluate the harmonic and scanning channels of both continuous‑ and composite‑X‑ray fluorescence in synchrotron X‑ray fluorescence spectrometers. Additionally, when the instrument employs a scanning path, the factors governing changes in measurement conditions depend on the specific instrument. For synchrotron X‑ray fluorescence spectrometers that do not utilize a scanning path, this option should not be selected. Measurement Condition 2 is intended solely for adjusting test parameters and does not require result calculation.
The above is an introduction to the verification procedure for RoHS analyzers. If you would like to learn more, please continue browsing. ROHS Analyzer Information Column If you’d like to learn more about our products, please leave us a message online, contact us directly, or give us a call!
Latest News
Recommended Products
Frequently Asked Questions
What is RoHS?
RoHS is a mandatory standard established by EU legislation, officially known as the “Directive on the Restriction of the Use of Certain Hazardous Substances in Electrical and Electronic Equipment.” The standard entered into force on July 1, 2006, and primarily aims to regulate the materials and manufacturing processes of electrical and electronic products, thereby promoting human health and environmental protection.
The purpose of this standard is to eliminate six substances—lead, mercury, cadmium, hexavalent chromium, polybrominated biphenyls, and polybrominated diphenyl ethers (note: the correct Chinese term for PBDE is “polybrominated diphenyl ethers”; “polybrominated biphenyl ethers” is an incorrect designation)—and specifically stipulates that the cadmium content must not exceed 0.01%.
The latest RoHS Directive has added four new phthalate‑ester testing items: DEHP—di(2‑ethylhexyl) phthalate; BBP—butyl benzyl phthalate; DBP—dibutyl phthalate; and DIBP—diisobutyl phthalate.
What are the limit values for the ten hazardous substances under RoHS?
1. Cadmium: less than 100 ppm.
2. Lead: less than 1000 ppm.
3. Mercury: less than 1000 ppm.
4. Hexavalent chromium: less than 1000 ppm.
5. Polybrominated biphenyls (PBB): less than 1000 ppm.
6. Polybrominated diphenyl ethers (PBDEs): less than 1000 ppm.
7. Di(2-ethylhexyl) phthalate (DEHP): less than 1000 ppm.
8. Butyl benzyl phthalate (BBP): less than 1000 ppm.
9. Dibutyl phthalate (DBP): less than 1000 ppm.
10. Diisobutyl phthalate (DIBP): less than 1000 ppm.
What instruments are typically used for RoHS testing?
The commonly used analytical instrument for RoHS 1.0 is X-ray fluorescence spectroscopy (XRF). XRF analyzers are available in standalone, benchtop, and handheld configurations. Typically, a handheld XRF analyzer is employed first for portable on-site testing, with a focus on components of the product that pose a higher risk of containing restricted substances.
With the introduction of RoHS 2.0, four phthalate esters have been added to the RoHS Directive, necessitating additional testing to determine the levels of these compounds, which are extracted using solvents. The solvent extracts are then analyzed for the presence of phthalates by gas chromatography–mass spectrometry (GC/MS) or gas chromatography with flame ionization detection (GC/FID). Other elements and polybrominated diphenyl ethers/alkylated elastomers are analyzed using comparative methods and GC/MS, respectively.
Do the ten RoHS items require two instruments for testing?
The ten items specified in RoHS require two instruments for testing.
For the ten‑item RoHS testing, two instruments are required to complete the analysis; a single instrument cannot achieve this. Even if one instrument is used to measure heavy metal content, the other must be employed to analyze organic compounds.
X-ray fluorescence spectrometers are used to detect RoHS 1.0, while chromatographs, pyrolyzers, or hyphenated techniques such as GC–MS are employed to analyze the four phthalates specified in RoHS 2.0. At present, no single instrument can independently perform all the tests required for RoHS 2.0.
Can ROHS testing instruments detect metallic elements?
ROHS testing instruments are specialized devices used to detect the presence of hazardous substances in electronic components, household appliances, toys, and other products. These harmful substances can pose long-term risks to both human health and the environment, making the use of ROHS testing equipment of paramount importance.
Regarding whether ROHS testing instruments can detect metallic elements, the answer is yes. This is because ROHS testing covers not only hazardous substances but also the analysis of metallic constituents in compliant products. These metals include lead, cadmium, mercury, and others, which are widely used during the manufacturing process. Although these elements are essential, their excessive use poses significant risks to human health and the environment. Consequently, the advent of ROHS testing instruments has greatly facilitated the detection of metallic elements.