The working principle of a circularly polarized fluorescence spectrometer!
发布时间: 2022-10-25 10:31
Circularly polarized fluorescence spectroscopy boasts a wide range of applications and promising prospects in fields such as luminescent materials, biomolecular proteins, information display and storage, electronics, and nonlinear optics, attracting considerable attention and interest from scientists. This technique can provide structural information about molecular excited states and characterize polymer architectures, making it an essential tool for investigating the stereochemistry of organic compounds. So, what is the underlying principle of circularly polarized fluorescence spectroscopy? Let’s take a brief look below.
Working principle: Light is an electromagnetic wave, describable in terms of a vibrating electric field and a magnetic field perpendicular to it. If, at a given point along the propagation path, the light wave exhibits a single vibration direction that rotates systematically by a fixed angle as the wave propagates—while the amplitude remains constant—the trajectory traced by the tip of the electric-field vector forms a helix with a circular cross-section; such light is termed circularly polarized. Building on the concept of circular dichroism, it has been observed that the intensities of left- and right-circularly polarized fluorescence differ. Typically, the degree of circular polarization fluorescence is quantified as the intensity difference between left- and right-circularly polarized fluorescence, denoted by CPL = ΔF = FL − FR.
Previously, reports on circularly polarized fluorescence detection were all conducted using instruments designed and built by the researchers themselves. It was not until 1972 that Steinberg and Gafni (SG) at the Weizmann Institute of Science in Israel proposed the circularly polarized fluorescence modulation measurement method shown in Figure 1A. The basic components include an excitation source, a monochromator, the sample, an optical elastic modulator, a polarizer, an emission monochromator, a photomultiplier tube, a lock-in amplifier, and a computer. In this approach, the modulated optoelectronic signal and the PEM optical elastic modulation signal are fed into the lock-in amplifier; by locking the frequencies and phases of these two signals, circularly polarized fluorescence is extracted from the overall fluorescence.
In 1982, Schippers, van den Beukle, and de Kruif (SBD) at Leiden University in the Netherlands introduced a circular dichroism fluorescence measurement method that replaced the lock-in amplifier with photon counting, thereby addressing the issue of unstable lock-in amplifier output. Subsequently, this approach was primarily employed to characterize complex protein structures; however, detecting weak circular dichroism signals remained exceedingly slow. Between 1992 and 1995, with advances in electronic technologies such as time-to-digital converters (TDCs), Schauerte, Steel, and Gafni (SSG) at the University of Michigan further developed a direct phase‑subtraction method for circular dichroism fluorescence. This technique utilizes a DGG delay‑gated pulse generator to separately measure the left‑circular dichroism signal (FL) and the right‑circular dichroism signal (FR) as defined by the equation ΔF = FL − FR, then directly subtracts these values to obtain the true circular dichroism signal ΔF. The asymmetry factor is calculated via the formula glum = 2(FL − FR)/(FL + FR). This method simultaneously resolved the problems of unstable lock-in loop output and slow measurement speeds inherent in the earlier approaches.
The above is an introduction to the working principle of the circularly polarized fluorescence spectrometer. If you would like to learn more, please continue browsing. Information Column on Circularly Polarized Fluorescence Spectrometers !
Latest News
Recommended Products
Frequently Asked Questions
What is RoHS?
RoHS is a mandatory standard established by EU legislation, officially known as the “Directive on the Restriction of the Use of Certain Hazardous Substances in Electrical and Electronic Equipment.” The standard entered into force on July 1, 2006, and primarily aims to regulate the materials and manufacturing processes of electrical and electronic products, thereby promoting human health and environmental protection.
The purpose of this standard is to eliminate six substances—lead, mercury, cadmium, hexavalent chromium, polybrominated biphenyls, and polybrominated diphenyl ethers (note: the correct Chinese term for PBDE is “polybrominated diphenyl ethers”; “polybrominated biphenyl ethers” is an incorrect designation)—and specifically stipulates that the cadmium content must not exceed 0.01%.
The latest RoHS Directive has added four new phthalate‑ester testing items: DEHP—di(2‑ethylhexyl) phthalate; BBP—butyl benzyl phthalate; DBP—dibutyl phthalate; and DIBP—diisobutyl phthalate.
What are the limit values for the ten hazardous substances under RoHS?
1. Cadmium: less than 100 ppm.
2. Lead: less than 1000 ppm.
3. Mercury: less than 1000 ppm.
4. Hexavalent chromium: less than 1000 ppm.
5. Polybrominated biphenyls (PBB): less than 1000 ppm.
6. Polybrominated diphenyl ethers (PBDEs): less than 1000 ppm.
7. Di(2-ethylhexyl) phthalate (DEHP): less than 1000 ppm.
8. Butyl benzyl phthalate (BBP): less than 1000 ppm.
9. Dibutyl phthalate (DBP): less than 1000 ppm.
10. Diisobutyl phthalate (DIBP): less than 1000 ppm.
What instruments are typically used for RoHS testing?
The commonly used analytical instrument for RoHS 1.0 is X-ray fluorescence spectroscopy (XRF). XRF analyzers are available in standalone, benchtop, and handheld configurations. Typically, a handheld XRF analyzer is employed first for portable on-site testing, with a focus on components of the product that pose a higher risk of containing restricted substances.
With the introduction of RoHS 2.0, four phthalate esters have been added to the RoHS Directive, necessitating additional testing to determine the levels of these compounds, which are extracted using solvents. The solvent extracts are then analyzed for the presence of phthalates by gas chromatography–mass spectrometry (GC/MS) or gas chromatography with flame ionization detection (GC/FID). Other elements and polybrominated diphenyl ethers/alkylated elastomers are analyzed using comparative methods and GC/MS, respectively.
Do the ten RoHS items require two instruments for testing?
The ten items specified in RoHS require two instruments for testing.
For the ten‑item RoHS testing, two instruments are required to complete the analysis; a single instrument cannot achieve this. Even if one instrument is used to measure heavy metal content, the other must be employed to analyze organic compounds.
X-ray fluorescence spectrometers are used to detect RoHS 1.0, while chromatographs, pyrolyzers, or hyphenated techniques such as GC–MS are employed to analyze the four phthalates specified in RoHS 2.0. At present, no single instrument can independently perform all the tests required for RoHS 2.0.
Can ROHS testing instruments detect metallic elements?
ROHS testing instruments are specialized devices used to detect the presence of hazardous substances in electronic components, household appliances, toys, and other products. These harmful substances can pose long-term risks to both human health and the environment, making the use of ROHS testing equipment of paramount importance.
Regarding whether ROHS testing instruments can detect metallic elements, the answer is yes. This is because ROHS testing covers not only hazardous substances but also the analysis of metallic constituents in compliant products. These metals include lead, cadmium, mercury, and others, which are widely used during the manufacturing process. Although these elements are essential, their excessive use poses significant risks to human health and the environment. Consequently, the advent of ROHS testing instruments has greatly facilitated the detection of metallic elements.