Calibration method for ROHS analyzers!
发布时间: 2022-10-24 10:13
With the development of society and the improvement of people’s living standards, the requirements for electronic products and their applications are becoming increasingly diverse. ROHS testing industry used in ROHS analyzer As the number of such instruments continues to grow, calibration methods for ROHS analyzers are receiving increasing attention. Below, we’ll introduce some common calibration procedures for ROHS analyzers—interested readers are encouraged to take a look!
I. Repeated Measurement Methods
When using an ROHS analyzer, ensure that the instrument is in normal operating condition, then continuously emit X-rays to acquire and generate the X-ray calibration curve. Next, select one sample for testing and perform at least 10 consecutive measurements to ensure that the recorded mass fractions of the target elements are accurate to the percentage level. Finally, calculate the standard relative deviation of these 10 measurements; this represents the repeatability of the ROHS analyzer.
II. Methods for Indicating Errors
The recommended data for ROHS analyzer testing represent the calibration curve between the instrument’s indicated values and the known reference values, established as a functional relationship between the X-ray intensity emitted by the X-ray tube and the concentrations of several elements. First, preheat the instrument for 30 minutes, then continuously irradiate standard samples. While the instrument is in operation, perform at least ten consecutive measurements, calculate the resulting measurement data, and determine the relative error—the difference between the measured relative concentration and the reference value.
III. Ionizing Radiation Methods
During testing, the X-ray radiation emitted by the ROHS analyzer can be measured at six distinct spatial locations—front, rear, top, bottom, left, and right. With the instrument operating under normal conditions, a dosimeter is used to measure the X‑ray energy release rate at each of these six external positions, with measurements taken at a distance of 5 cm. The maximum value among these measurements is then recorded.
IV. Stability Methods
Once the instrument is powered on and its performance has stabilized, it continuously emits standard X‑rays to generate a calibration curve. Subsequently, when analyzing samples, measurements should be repeated at least six times over a period of no less than two hours, ideally with intervals of approximately 15 minutes, to ensure that the reported elemental concentrations are accurate to within the specified percentage range. Finally, the relative standard deviation of the six measurement results is calculated; this value represents the analytical precision of the ROHS analyzer.
In summary, with the rapid advancement of science and technology, ROHS analyzers have become an indispensable tool—truly a choice of our times. This article aims to provide readers with a comprehensive understanding of ROHS analyzer calibration methods. For more information, please continue browsing. ROHS Analyzer Information Column If you’d like to learn more about our products, please leave us a message online or contact us by phone!
Latest News
Recommended Products
Frequently Asked Questions
What is RoHS?
RoHS is a mandatory standard established by EU legislation, officially known as the “Directive on the Restriction of the Use of Certain Hazardous Substances in Electrical and Electronic Equipment.” The standard entered into force on July 1, 2006, and primarily aims to regulate the materials and manufacturing processes of electrical and electronic products, thereby promoting human health and environmental protection.
The purpose of this standard is to eliminate six substances—lead, mercury, cadmium, hexavalent chromium, polybrominated biphenyls, and polybrominated diphenyl ethers (note: the correct Chinese term for PBDE is “polybrominated diphenyl ethers”; “polybrominated biphenyl ethers” is an incorrect designation)—and specifically stipulates that the cadmium content must not exceed 0.01%.
The latest RoHS Directive has added four new phthalate‑ester testing items: DEHP—di(2‑ethylhexyl) phthalate; BBP—butyl benzyl phthalate; DBP—dibutyl phthalate; and DIBP—diisobutyl phthalate.
What are the limit values for the ten hazardous substances under RoHS?
1. Cadmium: less than 100 ppm.
2. Lead: less than 1000 ppm.
3. Mercury: less than 1000 ppm.
4. Hexavalent chromium: less than 1000 ppm.
5. Polybrominated biphenyls (PBB): less than 1000 ppm.
6. Polybrominated diphenyl ethers (PBDEs): less than 1000 ppm.
7. Di(2-ethylhexyl) phthalate (DEHP): less than 1000 ppm.
8. Butyl benzyl phthalate (BBP): less than 1000 ppm.
9. Dibutyl phthalate (DBP): less than 1000 ppm.
10. Diisobutyl phthalate (DIBP): less than 1000 ppm.
What instruments are typically used for RoHS testing?
The commonly used analytical instrument for RoHS 1.0 is X-ray fluorescence spectroscopy (XRF). XRF analyzers are available in standalone, benchtop, and handheld configurations. Typically, a handheld XRF analyzer is employed first for portable on-site testing, with a focus on components of the product that pose a higher risk of containing restricted substances.
With the introduction of RoHS 2.0, four phthalate esters have been added to the RoHS Directive, necessitating additional testing to determine the levels of these compounds, which are extracted using solvents. The solvent extracts are then analyzed for the presence of phthalates by gas chromatography–mass spectrometry (GC/MS) or gas chromatography with flame ionization detection (GC/FID). Other elements and polybrominated diphenyl ethers/alkylated elastomers are analyzed using comparative methods and GC/MS, respectively.
Do the ten RoHS items require two instruments for testing?
The ten items specified in RoHS require two instruments for testing.
For the ten‑item RoHS testing, two instruments are required to complete the analysis; a single instrument cannot achieve this. Even if one instrument is used to measure heavy metal content, the other must be employed to analyze organic compounds.
X-ray fluorescence spectrometers are used to detect RoHS 1.0, while chromatographs, pyrolyzers, or hyphenated techniques such as GC–MS are employed to analyze the four phthalates specified in RoHS 2.0. At present, no single instrument can independently perform all the tests required for RoHS 2.0.
Can ROHS testing instruments detect metallic elements?
ROHS testing instruments are specialized devices used to detect the presence of hazardous substances in electronic components, household appliances, toys, and other products. These harmful substances can pose long-term risks to both human health and the environment, making the use of ROHS testing equipment of paramount importance.
Regarding whether ROHS testing instruments can detect metallic elements, the answer is yes. This is because ROHS testing covers not only hazardous substances but also the analysis of metallic constituents in compliant products. These metals include lead, cadmium, mercury, and others, which are widely used during the manufacturing process. Although these elements are essential, their excessive use poses significant risks to human health and the environment. Consequently, the advent of ROHS testing instruments has greatly facilitated the detection of metallic elements.