Seven Common Faults of Spark Direct-Reading Spectrometers and Their Troubleshooting Methods!

发布时间: 2022-10-19 11:02

Spark Direct-Reading Spectrometer It is a highly integrated testing device composed of multiple detectors. Proper operation and straightforward maintenance of the spark‑direct‑reading spectrometer can significantly extend its service life. This article presents seven methods for diagnosing and troubleshooting common malfunctions of the spark‑direct‑reading spectrometer.

1. The excitation spot on the sample is white, with no dark halo; this indicates incomplete excitation.


Troubleshooting and Analysis: The analysis indicates that impure argon gas fails to meet the test requirements. Prior to changing the argon supply, sample analysis was normal. However, even after prolonged argon purging and continuous excitation of waste samples, the issue persisted. By replacing the argon, flushing out the excitation‑induced waste, and then analyzing the sample, the fault was eliminated and the analysis returned to normal. Thus, after repeatedly addressing this type of failure, it was determined that incomplete excitation was the root cause, primarily due to insufficient argon purity.

II. ICAL calibration failed, and the band detector value is low.


Analysis and Troubleshooting: First, follow the troubleshooting procedure for the abnormal excitation spot described above. Then, clean the spectrometer, paying particular attention to the UV‑light entrance window and the air‑filled optical entrance window; clean these components with an acetone solution or replace them as necessary.

3. The sample was not fully excited, resulting in a measured carbon content that is significantly higher than the actual value.


Fault analysis and troubleshooting: During summer operations, insufficient sample excitation was observed, with C‑element concentrations measured far above the actual values. After ruling out other potential causes, the issue was attributed to excessively high ambient temperatures. The air conditioner was turned on to set the laboratory temperature to approximately 25°C, after which the samples were re‑analyzed. The measured concentrations of all elements returned to acceptable levels, resolving the problem. Therefore, during seasonal transitions and periods of significant temperature fluctuations, it is advisable to maintain a stable operating environment for the instrument and promptly establish standardized procedures to adjust calibration curves.

4. During the excitation process, the spectrometer suddenly stopped functioning.


Analysis and Troubleshooting: First, the distance between the sample‑pressing rods is too short, resulting in insufficient clamping; during excitation, the sample shifts, causing an open circuit between the light source and the circuit board. Adjust the rod spacing and compress the sample more firmly. Second, the surface of the ground sample is not sufficiently flat; during excitation, the sample tilts, leading to an open circuit between the light source and the circuit board. Grind the sample to achieve a level surface. Third, the argon inlet pressure is inadequate; increase it to 7.0 bar. Additionally, molten material generated during sample excitation may cause the metal filaments or clips of the electrode brush to detach, resulting in electrode shorting. Alternatively, clean the spark table and the fixture.

5. The sample cannot be excited, and the system prompts that the pressure rod has been raised.


Fault analysis and troubleshooting: Improper treatment of the contact surface between the sample and the pressure rod, or the presence of oxide scale or other insulating coatings, has caused a circuit blockage. Clean it thoroughly.

6. The argon gas in the spectrometer flows rapidly.


Troubleshooting: Check whether the pressure gauge is within the normal range; if not, there may be air leaks at various connections in the gas line.

7. The excitation point of the post‑steel sample is marked by a black circle.


Fault analysis and troubleshooting methods: First, check the purity of the argon gas; then inspect the argon spark‑table outlet to determine whether the exhaust tube is blocked—clean it if necessary. Blockage at the spark‑table outlet can impede proper exhaust, leading to black rings around the ignition point. If the black ring is particularly dark, it may continue to expand and become diffuse; this could also result from sample inhomogeneity or from sample leakage after being left for half a day. Additionally, changes in excitation conditions can cause abnormal coloration after excitation. At higher excitation energy, the degree of blackening may increase. Furthermore, phenomena such as carbon buildup on the excitation electrode, gas leaks, wear on the excitation stage, or an uneven sample surface can all contribute to this issue. When replacing electrodes, care should also be taken to prevent carbon deposition.

The above constitutes the complete overview of seven types of malfunctions in spark‑direct‑reading spectrometers, along with their corresponding troubleshooting methods. For more information, please continue browsing. Spark Direct-Reading Spectrometer Information Column If you’d like to learn more about our products, please leave us a message or give us a call!

Latest News

Recommended Products

RoHS 2.0 Rapid Screening Instrument PY-GCMS3080

This page provides information, including frequently asked questions, about the RoHS 2.0 rapid screening instrument PY‑GCMS3080. For pricing inquiries, please contact us online or call our hotline at 19926823034!

Learn more+

RoHS 2.0 Rapid Screening Instrument PY-GCMS3080

RoHS 1.0 Rapid Testing Instrument DX-320L Pro

This page provides an introduction to the DX-320L Pro, a high-performance desktop XRF analyzer compliant with RoHS 1.0. For more information, please contact 19926823034.

Learn more+

RoHS 1.0 Rapid Testing Instrument DX-320L Pro

RoHS 1.0 Detector – IDP8300L Fluorescence Spectrometer

This page provides specifications and FAQs for the DP8300L fluorescence spectrometer. For pricing inquiries, please contact us online or call 19926823034!

Learn more+

RoHS 1.0 Detector – IDP8300L Fluorescence Spectrometer

RoHS Six-Item Detector – DX-360 Fluorescence Spectrometer

This page provides specifications and FAQs for the DX-360 fluorescence spectrometer. For pricing inquiries, please contact us online or call 19926823034!

Learn more+

RoHS Six-Item Detector – DX-360 Fluorescence Spectrometer

RoHS 2.0 Pyrolysis Gas Chromatograph PY-DP8060

This page provides specifications for the RoHS 2.0 pyrolysis gas chromatograph PY-DP8060, along with FAQs and other information. For a quote, please contact us online or call our hotline at 19926823034!

Learn more+

RoHS 2.0 Pyrolysis Gas Chromatograph PY-DP8060

Halogen detection instrument

This page provides specifications and FAQs for the IDP-6300 halogen detector. For pricing inquiries, please contact us online or call our hotline at 19926823034.

Learn more+

Halogen detection instrument

Frequently Asked Questions

What is RoHS?

RoHS is a mandatory standard established by EU legislation, officially known as the “Directive on the Restriction of the Use of Certain Hazardous Substances in Electrical and Electronic Equipment.” The standard entered into force on July 1, 2006, and primarily aims to regulate the materials and manufacturing processes of electrical and electronic products, thereby promoting human health and environmental protection.

The purpose of this standard is to eliminate six substances—lead, mercury, cadmium, hexavalent chromium, polybrominated biphenyls, and polybrominated diphenyl ethers (note: the correct Chinese term for PBDE is “polybrominated diphenyl ethers”; “polybrominated biphenyl ethers” is an incorrect designation)—and specifically stipulates that the cadmium content must not exceed 0.01%.

The latest RoHS Directive has added four new phthalate‑ester testing items: DEHP—di(2‑ethylhexyl) phthalate; BBP—butyl benzyl phthalate; DBP—dibutyl phthalate; and DIBP—diisobutyl phthalate.

What are the limit values for the ten hazardous substances under RoHS?

1. Cadmium: less than 100 ppm.

2. Lead: less than 1000 ppm.

3. Mercury: less than 1000 ppm.

4. Hexavalent chromium: less than 1000 ppm.

5. Polybrominated biphenyls (PBB): less than 1000 ppm.

6. Polybrominated diphenyl ethers (PBDEs): less than 1000 ppm.

7. Di(2-ethylhexyl) phthalate (DEHP): less than 1000 ppm.

8. Butyl benzyl phthalate (BBP): less than 1000 ppm.

9. Dibutyl phthalate (DBP): less than 1000 ppm.

10. Diisobutyl phthalate (DIBP): less than 1000 ppm.

What instruments are typically used for RoHS testing?

The commonly used analytical instrument for RoHS 1.0 is X-ray fluorescence spectroscopy (XRF). XRF analyzers are available in standalone, benchtop, and handheld configurations. Typically, a handheld XRF analyzer is employed first for portable on-site testing, with a focus on components of the product that pose a higher risk of containing restricted substances.

With the introduction of RoHS 2.0, four phthalate esters have been added to the RoHS Directive, necessitating additional testing to determine the levels of these compounds, which are extracted using solvents. The solvent extracts are then analyzed for the presence of phthalates by gas chromatography–mass spectrometry (GC/MS) or gas chromatography with flame ionization detection (GC/FID). Other elements and polybrominated diphenyl ethers/alkylated elastomers are analyzed using comparative methods and GC/MS, respectively.

Do the ten RoHS items require two instruments for testing?

The ten items specified in RoHS require two instruments for testing.
For the ten‑item RoHS testing, two instruments are required to complete the analysis; a single instrument cannot achieve this. Even if one instrument is used to measure heavy metal content, the other must be employed to analyze organic compounds.
X-ray fluorescence spectrometers are used to detect RoHS 1.0, while chromatographs, pyrolyzers, or hyphenated techniques such as GC–MS are employed to analyze the four phthalates specified in RoHS 2.0. At present, no single instrument can independently perform all the tests required for RoHS 2.0.

Can ROHS testing instruments detect metallic elements?

ROHS testing instruments are specialized devices used to detect the presence of hazardous substances in electronic components, household appliances, toys, and other products. These harmful substances can pose long-term risks to both human health and the environment, making the use of ROHS testing equipment of paramount importance.

Regarding whether ROHS testing instruments can detect metallic elements, the answer is yes. This is because ROHS testing covers not only hazardous substances but also the analysis of metallic constituents in compliant products. These metals include lead, cadmium, mercury, and others, which are widely used during the manufacturing process. Although these elements are essential, their excessive use poses significant risks to human health and the environment. Consequently, the advent of ROHS testing instruments has greatly facilitated the detection of metallic elements.

< 12 >