Some issues to keep in mind when using HPLC!
发布时间: 2022-09-30 13:52
High-performance liquid chromatograph It is one of the most commonly used analytical instruments in laboratories and is seeing increasingly widespread application. During operation, this instrument inevitably encounters various issues that can directly affect the accuracy of measurement data and the proper functioning of the device. If the operator understands the root causes of these malfunctions, they will be able to clearly identify methods for preventing and resolving them, thereby ensuring correct usage and maximizing the instrument’s performance. Today, we will share some key considerations when using HPLC, focusing on the following aspects.
I. Issues with Using Test Tubes
1. Cleanliness of the test tube
High-performance liquid chromatography (HPLC) is a highly sensitive analytical technique. If unclean test tubes are used, the accuracy of the results can be compromised. For example, when methanol is employed as the solvent to dissolve a sample, the small test tubes used are fitted with rubber stoppers. Consequently, each injection exhibits an interfering peak with a consistent retention time. Upon investigation, it was determined that this interfering peak arises from components leached from the rubber stopper into the methanol; replacing the glass test tubes eliminated the interference.
2. Dissolution of plastic test tubes
In recent years, disposable plastic test tubes have greatly facilitated sample analysis. However, during use, it is essential to account for the dissolution of organic solvents into the tube walls. When extracting samples with such tubes, certain organic solvents—such as chloroform—can dissolve into the tube material; these dissolved substances may occasionally generate signals in the detector, thereby interfering with the quantification of the analyte. In such cases, one can first run a control experiment under identical conditions to determine whether the extract, in the absence of the target analyte, produces any spurious signals on the detector. If interference persists, it is advisable to switch to glass test tubes that are resistant to organic solvents.
3. Adsorption of the sample to be tested on the tube wall
This issue also requires attention, as failure to address it may compromise the accuracy of test results. In therapeutic drug monitoring (TDM), certain analytes such as amitriptyline and imipramine tend to adsorb onto the walls of glass tubes. Therefore, polypropylene tubes should be used during sample collection. To prevent adsorption during the extraction process, a 0.5% hexamethylenediamine solution can be employed as the extracting agent, which effectively mitigates adsorption.
II. Issues with Operating the Spray Valve
Currently, the 7725 injection valve is widely used in analytical high-performance liquid chromatography. Its internal six‑way valve design makes sample injection highly convenient; however, improper use can lead to problems. For instance, during HPLC analysis, issues such as abnormal chromatographic peaks and poor reproducibility sometimes arise, primarily due to incorrect operational procedures. To address these challenges, it is necessary to take action on the following fronts.
1. Control of sample volume
The injection valve is used for sample injection, and the sample loop within the valve is volumetrically precise (typically, analytical injection valves have a loop volume of 20 µL). Because the sample solution injected into the valve exhibits a radial velocity gradient in the loop’s tubing—where the flow rate at the tube axis is higher than at the tube wall—filling the quantitative loop with the sample solution and achieving accurate quantitation via the injection valve requires an injection volume greater than twice the loop’s volume. If the injection volume is controlled by a syringe, the maximum injection volume should be limited to half the loop’s volume to prevent some of the sample from overflowing through the overflow port, which could introduce errors in quantitative analysis.
2. Cleaning of the spray valve
If residual sample from the previous injection remains in the sample loop, it will inevitably contaminate the next injection. To prevent this, the following steps should be taken: a) The injection valve has two positions—inject and load. First, with the valve in the inject position, use a syringe to flush the injection valve repeatedly with mobile phase, using approximately 40 µL each time; b) Then, manually rotate the injection valve to the load position and rinse several times with mobile phase, again using 40 µL per rinse; finally, inject the sample into the injection valve. By following these procedures, contamination originating from the injection valve can be avoided, thereby eliminating interfering peaks and improving the accuracy of the analytical results.
3. The overflow pipe of the fuel injection valve is blocked.
Sometimes, the overflow tube of the injection valve becomes clogged, making it impossible to push the syringe needle when injecting a sample. This malfunction is caused by blockage of the overflow tube. Such blockages are often due to salt solutions used in the mobile phase for dissolving samples; the salts crystallize at the drain port of the overflow tube. In this case, you can gently soak the overflow nozzle in a small beaker containing distilled water to dissolve the salt crystals at the port, thereby clearing the blockage. Additionally, regularly flushing the overflow tube with distilled water after each injection can help prevent this issue.
III. Mobile Phase (MOBILE PHASE)
In high-performance liquid chromatography, methanol and acetonitrile are commonly used to prepare the mobile phase. For HPLC, it is preferable to use high‑grade, dedicated reagents, such as chromatographic‑grade solvents. When the requirements are less stringent, reagent‑grade or even analytical‑grade solvents may be employed. Ultraviolet detectors are widely used in HPLC; therefore, to minimize baseline noise and enhance analytical sensitivity, chromatographic‑grade reagents with low UV absorbance and low impurity levels should be selected.
1. Filtration of the mobile phase
The prepared mobile phase must be filtered through a 0.5 µm membrane filter before use. This is because the solution contains numerous microscopic particles that are difficult to detect with the naked eye. If these particles are not removed, the filters at the pump inlet and the column head can become clogged, obstructing the normal flow of the mobile phase, increasing column backpressure, and reducing column efficiency. In such cases, switch to a filtered mobile phase, remove the clogged filter, soak it in a 20% nitric acid–water solution, and sonicate it for 20 minutes to eliminate any blockages on the filter.
2. Degassing of the mobile phase
The mobile phase must be degassed prior to use to remove as much dissolved gas as possible; otherwise, these gases can degrade the performance of the chromatographic column packing and significantly interfere with detector signals. Various degassing methods are available, including ultrasonic degassing, vacuum degassing, and nitrogen sparging. Vacuum degassing and nitrogen‑sparging are currently the most widely employed techniques. When mixed with water and methanol, substantial bubble formation can occur; if used without degassing, these bubbles may enter the chromatographic column and detector, compromising analytical results.
IV. Use and Maintenance of Chromatographic Columns
The chromatographic column is the most critical component of a high-performance liquid chromatograph, and its performance plays a decisive role in the separation and quantification of the analytes. Therefore, in routine operations, special attention should be paid to the proper use and maintenance of the column to extend its service life.
1. Use a precolumn and a guard column
The guard column is installed between the pump and the injector to ensure complete equilibration of the mobile phase within the analytical column, thereby preventing components that could degrade the column packing or contaminants from entering the column. The guard column traps analytes that might strongly adsorb onto the analytical column, and it should be made of the same material as the column packing. By replacing the guard and pre‑column frequently rather than the analytical column itself, the service life of the analytical column can be extended.
2. Prevent gas from entering the chromatographic column
Some chromatographic columns, such as gel filtration columns, must not allow air bubbles to enter; otherwise, column efficiency may be reduced, and tiny air pockets that are difficult to remove can form. Therefore, to prevent air bubbles from entering the column, it is essential to use degassed mobile phase and to install the column strictly according to the following steps: a. Remove the sealing screw at the inlet of the column and check whether any solvent leaks out. b. If solvent leaks, connect the column to the tubing to prevent air ingress; if no solvent leaks, this indicates that air has already entered that end of the column. At this point, connect the outlet end of the column to the injection valve and flush the column in the reverse direction with the mobile phase to remove any trapped air. It is best to flush the column at a low flow rate of 0.2 mL/min. If the solvent flow rate is too high or the pressure rises suddenly, the column’s performance may deteriorate. d. If the flowing solvent is free of bubbles, it means that the gas inside the column has been expelled. Then reconnect the column in the correct orientation so that no further air can enter.
3. Clean the chromatographic column
To prevent the analyte and impurities from remaining in the chromatographic column, the column should be promptly cleaned after each sample analysis. First, the column must be eluted with a solvent that has strong eluting power for the analyte. Taking the commonly used reversed-phase chromatography as an example, since the first eluting compound is typically polar, 100% methanol or a slightly less polar solvent such as isopropyl alcohol and tetrahydrofuran should be employed to remove the polar substances adsorbed on the column. The volume of eluent is usually about 20 times the column volume. If the mobile phase is a buffer solution, the column should then be rinsed with distilled water to remove any salts, followed by washing with an appropriate solvent.
In gel filtration chromatography, the gel column typically uses a buffer solution as the mobile phase and must be rinsed with distilled water after use. For continuous operation, the buffer may be left in the column overnight, but it is best to maintain a low flow rate (less than 0.5 mL/min) to prevent precipitation of buffer salts. If the mobile phase contains halides, even when the column is left idle overnight, it must still be thoroughly flushed with distilled water to avoid halide‑induced corrosion.
4. Storage of the Chromatographic Column
If the chromatographic column is not in use for the time being, the following points should be observed during storage:
A. For short-term storage lasting a few days, flush the chromatographic column with a solvent (such as a gel‑filtration column or distilled water), then seal both ends of the column with sealing screws.
B. If the chromatographic column is not used for an extended period, the above‑mentioned methods may no longer be sufficient. In such cases, the column should be filled with the solvent specified in its instruction manual. For reversed-phase columns, methanol is typically used; for normal-phase columns, n-hexane or heptane may be employed; and gel filtration columns must not be flushed with water. Since microbial growth within the column can reduce its efficiency, the column should be rinsed with a 0.05% aqueous sodium azide solution (a preservative), after which it should be tightly sealed. When storing the column for a prolonged period, both ends must be securely sealed to prevent solvent evaporation, which can cause the packing material to dry out and shrink, thereby severely compromising column performance.
C. The chromatographic column should be stored at room temperature. If exposed to temperatures below 0°C, the column may freeze, which can also reduce its efficiency.
High-Performance Liquid Chromatography (HPLC) Its structure typically comprises a high-pressure pumping system, an injection system, a separation system, a recording and detection system, and a data-processing system. Preparative instruments should also be equipped with a fraction collection system. To achieve superior analytical results, HPLC instruments impose stricter requirements on accuracy, precision, sensitivity, and reproducibility.
Latest News
Recommended Products
Frequently Asked Questions
What is RoHS?
RoHS is a mandatory standard established by EU legislation, officially known as the “Directive on the Restriction of the Use of Certain Hazardous Substances in Electrical and Electronic Equipment.” The standard entered into force on July 1, 2006, and primarily aims to regulate the materials and manufacturing processes of electrical and electronic products, thereby promoting human health and environmental protection.
The purpose of this standard is to eliminate six substances—lead, mercury, cadmium, hexavalent chromium, polybrominated biphenyls, and polybrominated diphenyl ethers (note: the correct Chinese term for PBDE is “polybrominated diphenyl ethers”; “polybrominated biphenyl ethers” is an incorrect designation)—and specifically stipulates that the cadmium content must not exceed 0.01%.
The latest RoHS Directive has added four new phthalate‑ester testing items: DEHP—di(2‑ethylhexyl) phthalate; BBP—butyl benzyl phthalate; DBP—dibutyl phthalate; and DIBP—diisobutyl phthalate.
What are the limit values for the ten hazardous substances under RoHS?
1. Cadmium: less than 100 ppm.
2. Lead: less than 1000 ppm.
3. Mercury: less than 1000 ppm.
4. Hexavalent chromium: less than 1000 ppm.
5. Polybrominated biphenyls (PBB): less than 1000 ppm.
6. Polybrominated diphenyl ethers (PBDEs): less than 1000 ppm.
7. Di(2-ethylhexyl) phthalate (DEHP): less than 1000 ppm.
8. Butyl benzyl phthalate (BBP): less than 1000 ppm.
9. Dibutyl phthalate (DBP): less than 1000 ppm.
10. Diisobutyl phthalate (DIBP): less than 1000 ppm.
What instruments are typically used for RoHS testing?
The commonly used analytical instrument for RoHS 1.0 is X-ray fluorescence spectroscopy (XRF). XRF analyzers are available in standalone, benchtop, and handheld configurations. Typically, a handheld XRF analyzer is employed first for portable on-site testing, with a focus on components of the product that pose a higher risk of containing restricted substances.
With the introduction of RoHS 2.0, four phthalate esters have been added to the RoHS Directive, necessitating additional testing to determine the levels of these compounds, which are extracted using solvents. The solvent extracts are then analyzed for the presence of phthalates by gas chromatography–mass spectrometry (GC/MS) or gas chromatography with flame ionization detection (GC/FID). Other elements and polybrominated diphenyl ethers/alkylated elastomers are analyzed using comparative methods and GC/MS, respectively.
Do the ten RoHS items require two instruments for testing?
The ten items specified in RoHS require two instruments for testing.
For the ten‑item RoHS testing, two instruments are required to complete the analysis; a single instrument cannot achieve this. Even if one instrument is used to measure heavy metal content, the other must be employed to analyze organic compounds.
X-ray fluorescence spectrometers are used to detect RoHS 1.0, while chromatographs, pyrolyzers, or hyphenated techniques such as GC–MS are employed to analyze the four phthalates specified in RoHS 2.0. At present, no single instrument can independently perform all the tests required for RoHS 2.0.
Can ROHS testing instruments detect metallic elements?
ROHS testing instruments are specialized devices used to detect the presence of hazardous substances in electronic components, household appliances, toys, and other products. These harmful substances can pose long-term risks to both human health and the environment, making the use of ROHS testing equipment of paramount importance.
Regarding whether ROHS testing instruments can detect metallic elements, the answer is yes. This is because ROHS testing covers not only hazardous substances but also the analysis of metallic constituents in compliant products. These metals include lead, cadmium, mercury, and others, which are widely used during the manufacturing process. Although these elements are essential, their excessive use poses significant risks to human health and the environment. Consequently, the advent of ROHS testing instruments has greatly facilitated the detection of metallic elements.