For a full-spectrum direct-reading spectrometer, is it better to evacuate the chamber or to purge it with argon?

发布时间: 2022-09-30 11:40

With the growing demands of casting enterprises, Full-spectrum direct-reading spectrometer It boasts a more compact footprint, convenient matrix and channel addition, and accuracy that meets the needs of most customers. Currently, full-spectrum direct-reading optical spectrometers on the market are available in either argon‑filled or vacuum‑sealed configurations. When selecting a direct‑reading optical spectrometer, many users wonder: for a full‑spectrum model, is it better to use a vacuum system or an argon‑filled one?

To explain why a full-spectrum direct-reading spectrometer must be purged with argon and evacuated prior to analysis: Direct-reading spectrometers are used to analyze various metallic materials—such as cast iron, stainless steel, low-alloy steel, and copper alloys—and they also need to detect nonmetallic elements like carbon, phosphorus, sulfur, boron, arsenic, and nitrogen. The spectral lines of these elements lie in the vacuum ultraviolet region, where oxygen, nitrogen, and water vapor in the air strongly absorb UV radiation, causing a sharp attenuation of the measured UV spectral intensity. This, in turn, compromises the accuracy and stability of the analytical results for the target elements. Therefore, the instrument must operate under an inert, non‑air environment to ensure relatively stable detection of UV‑region elements.

At present, there are two primary approaches to achieving non‑air‑filled operation in the vacuum ultraviolet (VUV) spectral region: 1) vacuum pumping within the optical chamber; and 2) filling the chamber with argon gas. From the perspective of pure UV transmission, both methods can be used to measure elements such as carbon (C), phosphorus (P), sulfur (S), boron (B), arsenic (As), and nitrogen (N). These two approaches reflect the distinct technological strategies adopted by different spectrometer manufacturers. The following sections provide a detailed explanation of the operating principles underlying each of these techniques.

1. Vacuum‑pumping mode. In this mode, a vacuum pump evacuates the air from the optical chamber, creating a vacuum inside and enabling the measurement of elements in the ultraviolet region. The vacuum‑pumping optical system comprises the optical chamber, a vacuum pump, solenoid valves, optical components, and electrical control units. This type of spectrometer features a relatively large optical chamber, with both the diffraction grating and the optical detector housed within it. Because vacuum pumping generates negative pressure, the chamber walls are made thicker and reinforced with stiffening ribs to prevent deformation. At present, most full‑spectrum direct‑reading spectrometers manufactured domestically employ a vacuum‑pumping design.

Vacuum‑pumping technology for spectrometers is relatively mature; however, the addition of a vacuum system increases costs compared with argon‑filled systems. During vacuum pumping, the following issues may arise: 1) The optical chamber may deform due to insufficient pressure resistance, resulting in poor data stability; 2) Some manufacturers, in an effort to reduce costs, omit differential‑pressure valves, allowing oil and gas to be drawn back into the optical chamber and causing contamination; 3) Vacuum pumps can fail after prolonged operation, necessitating maintenance or replacement.

2. Argon‑purging mode. In this mode, an inert gas—typically argon—is introduced into the optical chamber to displace the air, thereby establishing optimal analytical conditions for elements in the ultraviolet region. The argon‑purged optical system comprises the optical chamber, a gas‑delivery subsystem, optical components, and electrical controls. Argon purging eliminates the need for a mechanical pump, and the pressure differential between the chamber and the ambient environment is negligible, so the chamber’s pressure‑resistance requirements are less stringent. Moreover, since the grating and detector can be mounted externally, the optical chamber is significantly smaller than that of a vacuum‑type instrument; for a given focal length, an argon‑purged instrument is both more compact and lighter.

Argon‑filled direct‑reading spectrometers offer numerous advantages, but they also have certain drawbacks compared with vacuum‑type spectrometers: 1) Since the optical chamber must be purged with argon, argon consumption is relatively higher; 2) If the purity of the introduced argon is insufficient, it can lead to contamination of the light‑element chamber.

At present, mainstream imported full-spectrum spectrometers, owing to their mature technology and cost-effectiveness, are mostly purged with argon.

Latest News

Recommended Products

RoHS 2.0 Rapid Screening Instrument PY-GCMS3080

This page provides information, including frequently asked questions, about the RoHS 2.0 rapid screening instrument PY‑GCMS3080. For pricing inquiries, please contact us online or call our hotline at 19926823034!

Learn more+

RoHS 2.0 Rapid Screening Instrument PY-GCMS3080

RoHS 1.0 Rapid Testing Instrument DX-320L Pro

This page provides an introduction to the DX-320L Pro, a high-performance desktop XRF analyzer compliant with RoHS 1.0. For more information, please contact 19926823034.

Learn more+

RoHS 1.0 Rapid Testing Instrument DX-320L Pro

RoHS 1.0 Detector – IDP8300L Fluorescence Spectrometer

This page provides specifications and FAQs for the DP8300L fluorescence spectrometer. For pricing inquiries, please contact us online or call 19926823034!

Learn more+

RoHS 1.0 Detector – IDP8300L Fluorescence Spectrometer

RoHS Six-Item Detector – DX-360 Fluorescence Spectrometer

This page provides specifications and FAQs for the DX-360 fluorescence spectrometer. For pricing inquiries, please contact us online or call 19926823034!

Learn more+

RoHS Six-Item Detector – DX-360 Fluorescence Spectrometer

RoHS 2.0 Pyrolysis Gas Chromatograph PY-DP8060

This page provides specifications for the RoHS 2.0 pyrolysis gas chromatograph PY-DP8060, along with FAQs and other information. For a quote, please contact us online or call our hotline at 19926823034!

Learn more+

RoHS 2.0 Pyrolysis Gas Chromatograph PY-DP8060

Halogen detection instrument

This page provides specifications and FAQs for the IDP-6300 halogen detector. For pricing inquiries, please contact us online or call our hotline at 19926823034.

Learn more+

Halogen detection instrument

Frequently Asked Questions

What is RoHS?

RoHS is a mandatory standard established by EU legislation, officially known as the “Directive on the Restriction of the Use of Certain Hazardous Substances in Electrical and Electronic Equipment.” The standard entered into force on July 1, 2006, and primarily aims to regulate the materials and manufacturing processes of electrical and electronic products, thereby promoting human health and environmental protection.

The purpose of this standard is to eliminate six substances—lead, mercury, cadmium, hexavalent chromium, polybrominated biphenyls, and polybrominated diphenyl ethers (note: the correct Chinese term for PBDE is “polybrominated diphenyl ethers”; “polybrominated biphenyl ethers” is an incorrect designation)—and specifically stipulates that the cadmium content must not exceed 0.01%.

The latest RoHS Directive has added four new phthalate‑ester testing items: DEHP—di(2‑ethylhexyl) phthalate; BBP—butyl benzyl phthalate; DBP—dibutyl phthalate; and DIBP—diisobutyl phthalate.

What are the limit values for the ten hazardous substances under RoHS?

1. Cadmium: less than 100 ppm.

2. Lead: less than 1000 ppm.

3. Mercury: less than 1000 ppm.

4. Hexavalent chromium: less than 1000 ppm.

5. Polybrominated biphenyls (PBB): less than 1000 ppm.

6. Polybrominated diphenyl ethers (PBDEs): less than 1000 ppm.

7. Di(2-ethylhexyl) phthalate (DEHP): less than 1000 ppm.

8. Butyl benzyl phthalate (BBP): less than 1000 ppm.

9. Dibutyl phthalate (DBP): less than 1000 ppm.

10. Diisobutyl phthalate (DIBP): less than 1000 ppm.

What instruments are typically used for RoHS testing?

The commonly used analytical instrument for RoHS 1.0 is X-ray fluorescence spectroscopy (XRF). XRF analyzers are available in standalone, benchtop, and handheld configurations. Typically, a handheld XRF analyzer is employed first for portable on-site testing, with a focus on components of the product that pose a higher risk of containing restricted substances.

With the introduction of RoHS 2.0, four phthalate esters have been added to the RoHS Directive, necessitating additional testing to determine the levels of these compounds, which are extracted using solvents. The solvent extracts are then analyzed for the presence of phthalates by gas chromatography–mass spectrometry (GC/MS) or gas chromatography with flame ionization detection (GC/FID). Other elements and polybrominated diphenyl ethers/alkylated elastomers are analyzed using comparative methods and GC/MS, respectively.

Do the ten RoHS items require two instruments for testing?

The ten items specified in RoHS require two instruments for testing.
For the ten‑item RoHS testing, two instruments are required to complete the analysis; a single instrument cannot achieve this. Even if one instrument is used to measure heavy metal content, the other must be employed to analyze organic compounds.
X-ray fluorescence spectrometers are used to detect RoHS 1.0, while chromatographs, pyrolyzers, or hyphenated techniques such as GC–MS are employed to analyze the four phthalates specified in RoHS 2.0. At present, no single instrument can independently perform all the tests required for RoHS 2.0.

Can ROHS testing instruments detect metallic elements?

ROHS testing instruments are specialized devices used to detect the presence of hazardous substances in electronic components, household appliances, toys, and other products. These harmful substances can pose long-term risks to both human health and the environment, making the use of ROHS testing equipment of paramount importance.

Regarding whether ROHS testing instruments can detect metallic elements, the answer is yes. This is because ROHS testing covers not only hazardous substances but also the analysis of metallic constituents in compliant products. These metals include lead, cadmium, mercury, and others, which are widely used during the manufacturing process. Although these elements are essential, their excessive use poses significant risks to human health and the environment. Consequently, the advent of ROHS testing instruments has greatly facilitated the detection of metallic elements.

< 12 >