What is the purpose of calibrating a handheld alloy analyzer?
发布时间: 2022-09-20 11:26
We all know Handheld alloy analyzer Test data serve as an essential basis for product validation. However, if instruments are left uncalibrated for an extended period, the accuracy of the test data cannot be guaranteed. And if the data are inaccurate, the equipment loses its practical value. So, handheld… Alloy analyzer What is the purpose of calibration? Let’s take a look!
In simple terms, the purpose of instrument calibration is to determine whether the equipment is functioning properly and whether its measurement results fall within the acceptable range of deviation. Calibration involves comparing measurements against established standards, assessing the indication error of the measuring device, and ensuring the accuracy of the measured values. It constitutes a bottom-up traceability process that establishes the source of measurement values. Such evaluation of indication errors shall be conducted in accordance with the organization’s calibration procedures, at prescribed calibration intervals, and documented through calibration records and calibration marks.
1. Handheld alloy analyzers can automatically perform self‑checks to correct for environmental factors that may affect analytical results. In particular, it is recommended to run a self‑check on the instrument system before use in the following situations:
1. When the ambient temperature of the instrument varies significantly;
2. When the test data deviates significantly from previous results;
3. If the instrument has not been used for an extended period, when it is used again;
Over time, the appropriateness of a metrological analyzer’s calibration interval depends on the calibration pass rate and historical calibration records, which can serve as the primary basis. However, changes in time, operating conditions, or the manner and circumstances of the analyzer’s use may lead to drift. Therefore, when the scheduled calibration interval arrives, the analyzer should be calibrated without delay. Furthermore, during the valid calibration period, the analyzer’s bias should be periodically checked at random. Based on this information, the calibration interval should be adjusted accordingly—either extended or shortened as needed.
II. The calibration interval is determined according to the following formula:
(1) Regarding usage frequency: As a commonly used alloy analyzer, its metrological performance tends to degrade over time; this issue can be mitigated by shortening the calibration interval. Of course, this approach also serves as an important means of enhancing raw material properties, refining manufacturing processes, and extending service life.
(2) Measurement accuracy requirements. For high‑precision equipment, the calibration interval may be appropriately shortened. Each organization should determine its desired level of accuracy based on its specific circumstances and select an appropriate accuracy grade. Avoid blindly pursuing excessively high accuracy, as this may lead to unnecessary costs. Conversely, setting the accuracy too low—so that it fails to meet operational requirements—would also be undesirable, as it could result in losses for the work at hand.
(3) If the user has strong maintenance capabilities, the calibration interval should be appropriately shortened; otherwise, the calibration interval will be longer.
(4) The performance of the handheld alloy analyzer was tested, with particular emphasis on its long-term stability and reliability. Even if analyzers of the same type exhibit poor stability and reliability, the calibration interval should be shortened.
(5) It is closely related to product quality and has specific requirements. The calibration interval is relatively short in such cases, and vice versa.
III. The steps for calibrating the touch screen are as follows:
1. Click the “Correct Touchscreen” icon.
2. The screen will prompt you to calibrate the touch screen; select “Yes.”
3. The screen will display “Please perform screen calibration” and show a cross symbol. Use the stylus to tap the center of the cross in sequence. Once all taps are complete, the touchscreen calibration is finished.
Note: If the touchscreen malfunctions, use the four-way navigation keys on the control panel to access the screen calibration menu and recalibrate the display.
The above is an introduction to the role of calibration for handheld alloy analyzers. If you would like to learn more about… Handheld Alloy Analyzer Information Yes, you can continue browsing our website or contact us!
Latest News
Recommended Products
Frequently Asked Questions
What is RoHS?
RoHS is a mandatory standard established by EU legislation, officially known as the “Directive on the Restriction of the Use of Certain Hazardous Substances in Electrical and Electronic Equipment.” The standard entered into force on July 1, 2006, and primarily aims to regulate the materials and manufacturing processes of electrical and electronic products, thereby promoting human health and environmental protection.
The purpose of this standard is to eliminate six substances—lead, mercury, cadmium, hexavalent chromium, polybrominated biphenyls, and polybrominated diphenyl ethers (note: the correct Chinese term for PBDE is “polybrominated diphenyl ethers”; “polybrominated biphenyl ethers” is an incorrect designation)—and specifically stipulates that the cadmium content must not exceed 0.01%.
The latest RoHS Directive has added four new phthalate‑ester testing items: DEHP—di(2‑ethylhexyl) phthalate; BBP—butyl benzyl phthalate; DBP—dibutyl phthalate; and DIBP—diisobutyl phthalate.
What are the limit values for the ten hazardous substances under RoHS?
1. Cadmium: less than 100 ppm.
2. Lead: less than 1000 ppm.
3. Mercury: less than 1000 ppm.
4. Hexavalent chromium: less than 1000 ppm.
5. Polybrominated biphenyls (PBB): less than 1000 ppm.
6. Polybrominated diphenyl ethers (PBDEs): less than 1000 ppm.
7. Di(2-ethylhexyl) phthalate (DEHP): less than 1000 ppm.
8. Butyl benzyl phthalate (BBP): less than 1000 ppm.
9. Dibutyl phthalate (DBP): less than 1000 ppm.
10. Diisobutyl phthalate (DIBP): less than 1000 ppm.
What instruments are typically used for RoHS testing?
The commonly used analytical instrument for RoHS 1.0 is X-ray fluorescence spectroscopy (XRF). XRF analyzers are available in standalone, benchtop, and handheld configurations. Typically, a handheld XRF analyzer is employed first for portable on-site testing, with a focus on components of the product that pose a higher risk of containing restricted substances.
With the introduction of RoHS 2.0, four phthalate esters have been added to the RoHS Directive, necessitating additional testing to determine the levels of these compounds, which are extracted using solvents. The solvent extracts are then analyzed for the presence of phthalates by gas chromatography–mass spectrometry (GC/MS) or gas chromatography with flame ionization detection (GC/FID). Other elements and polybrominated diphenyl ethers/alkylated elastomers are analyzed using comparative methods and GC/MS, respectively.
Do the ten RoHS items require two instruments for testing?
The ten items specified in RoHS require two instruments for testing.
For the ten‑item RoHS testing, two instruments are required to complete the analysis; a single instrument cannot achieve this. Even if one instrument is used to measure heavy metal content, the other must be employed to analyze organic compounds.
X-ray fluorescence spectrometers are used to detect RoHS 1.0, while chromatographs, pyrolyzers, or hyphenated techniques such as GC–MS are employed to analyze the four phthalates specified in RoHS 2.0. At present, no single instrument can independently perform all the tests required for RoHS 2.0.
Can ROHS testing instruments detect metallic elements?
ROHS testing instruments are specialized devices used to detect the presence of hazardous substances in electronic components, household appliances, toys, and other products. These harmful substances can pose long-term risks to both human health and the environment, making the use of ROHS testing equipment of paramount importance.
Regarding whether ROHS testing instruments can detect metallic elements, the answer is yes. This is because ROHS testing covers not only hazardous substances but also the analysis of metallic constituents in compliant products. These metals include lead, cadmium, mercury, and others, which are widely used during the manufacturing process. Although these elements are essential, their excessive use poses significant risks to human health and the environment. Consequently, the advent of ROHS testing instruments has greatly facilitated the detection of metallic elements.