Common Issues and Maintenance of Direct-Reading Spectrometers
发布时间: 2019-04-16 15:40
An optical emission spectrometer, also known as an OES, is an instrument that determines the elemental composition of a material by analyzing the characteristic emission spectra emitted by atoms or ions of each element when the sample is thermally or electrically excited. It is widely used in industries involving metals, such as casting, steel production, metal recycling, and aerospace. So, what common issues might this type of instrument encounter, and how can they be addressed?
I. Common Issues and Solutions
1. The test result for the content of a certain element is unexpectedly negative—what could be the reason?
a. The concentration of this element in the sample is too low—below the detection limit.
b. The data collector has encountered a problem.
c. The light path has deviated.
At this point, the light signal intensity detected by the photomultiplier tube has decreased significantly compared to normal conditions. The solution is to re‑align the instrument’s optical components to their optimal position using a trace‑based adjustment procedure, ensuring that the photomultiplier tube receives the strongest possible light signal.
d. The equipment has drifted.
Equipment can be restored to normal operation through methods such as type standardization and calibration. It is recommended to test previously measured samples one by one to systematically identify and resolve any issues.
2. Does the material of the sampler affect the accuracy of the analytical results?
Compared with iron molds, copper molds yield significantly better results for cast‑iron specimens; therefore, iron molds are not recommended for this purpose.
For laboratory analysis of copper and certain nonferrous metal samples, a graphite crucible is recommended as the mold.
3 Recently, the consumption of argon has increased significantly compared to the past—what’s going on??
What should we do? It’s recommended to start from several aspects.
a. Check that all connections along the argon gas pipeline are secure.
b. Check whether the argon purging time has become longer than before.
c. Has the intake pressure been increased?
d. Open a new cylinder of argon and check whether the gas is recent; if so, the pressure at the time of purchase may not have been as high as before. (There might be unscrupulous sellers!)
4. Do I need to construct the working curve myself?
Working curves are typically established by the manufacturer, and users generally do not re‑establish existing ones.
First, it is cumbersome; second, there is a high likelihood that our corresponding standard‑block configurations are incomplete, and without sufficient reference standards, the reliability of the calibration curve we have established is questionable; third, this is a technically demanding task—any oversight at any stage could result in test results based on your calibration curve deviating significantly from the true values.
5. When preparing spectral samples from scrap copper or copper raw materials, is it necessary to add a flux during the melting process?
No, it’s not necessary. Copper has a relatively low melting point, making sample preparation straightforward; a medium-frequency furnace can easily melt it.
6. Is direct-reading optical emission spectroscopy accurate for determining carbon and sulfur contents in pig iron?
It is common to use direct-reading optical spectroscopy to determine carbon and sulfur contents in pig iron, but the results are not very accurate. Achieving high measurement accuracy comes at a significant cost, making it economically impractical. We recommend employing chemical methods or a carbon–sulfur analyzer for these determinations.
After replacing the excitation stage, it took a long time for the equipment to stabilize. What could be the reason?
a. Does the excitation stage produce a lot of dust? If so, that’s correct.
b. When replacing the excitation stage, the vacuum isolation valve handle was not fully turned from the Open position to the Close position, allowing some air to be drawn back into the optical chamber. Should we wait for a while to allow the vacuum level to stabilize and perform several consecutive excitations on some previously analyzed samples to see if the situation improves?
8. What is the function of a relief valve?
The primary concern is to maintain a stable vacuum level in the chamber; secondly, this prevents oil from the vacuum pump from entering the equipment and contaminating components such as gratings and mirrors. Ultimately, the goal is to extend the service life of the vacuum pump by a few more days.
9. What are the fatigue lamp test and the dark current test?
The fatigue‑lamp test helps determine whether the device’s optical system is functioning properly and identifies any faults; the dark‑current test helps assess the integrity of the device’s electronic system and pinpoints potential issues. These two tests are typically performed once per quarter.
II. Routine Maintenance
1. Cleaning the ventilation filter element
Cleaning the ventilation filter isn’t as obvious as cleaning the electrodes or the excitation platform, but it’s equally important. It helps with heat dissipation and air circulation in the electronic cabinet. Typically, you can use a vacuum cleaner to remove accumulated debris; if the buildup is particularly stubborn, we recommend washing it with water mixed with a detergent. Be sure to gently squeeze the filter from top to bottom—never rub or scrub it, as this could render it unusable. Once it’s completely dry, reinstall it for continued use.
2. Regularly back up daily data
Typically, after completing slit alignment and standardization, a backup should be performed to ensure that, in the event of computer system failures or virus infections, the equipment can be quickly restored to service.
Shenzhen Delptek Instrument Co., Ltd. is a high-tech enterprise specializing in the R&D, manufacturing, and sales of spectroscopic (XRF), chromatographic, and mass spectrometric products. The company has two wholly owned subsidiaries—Shenzhen Shengwangda Instrument Equipment Co., Ltd. and Shenzhen Dezewei Testing Technology Co., Ltd.—and also holds controlling interests in Hong Kong Songsheng Electronics Co., Ltd. and the Longgang Wanda Industrial Park in Shenzhen. We boast an R&D team that combines X-ray fluorescence technology with expertise from German technical specialists, achieving outstanding results across diverse fields. Our spectroscopy, chromatography, and mass spectrometry instruments are already widely deployed in numerous applications. Website: http://www.delptek.cn/ Address: Building 8, Shangxiawei Entrepreneurship Industrial Park, Sha Jing Subdistrict, Bao’an District, Shenzhen Contact Information: Mobile: 18688929727 / 13751165856 Landline: 0755-66862188 Contacts: Ms. Liu, Mr. Yang E-mail: sell@delptek.cn
Latest News
Recommended Products
Frequently Asked Questions
What is RoHS?
RoHS is a mandatory standard established by EU legislation, officially known as the “Directive on the Restriction of the Use of Certain Hazardous Substances in Electrical and Electronic Equipment.” The standard entered into force on July 1, 2006, and primarily aims to regulate the materials and manufacturing processes of electrical and electronic products, thereby promoting human health and environmental protection.
The purpose of this standard is to eliminate six substances—lead, mercury, cadmium, hexavalent chromium, polybrominated biphenyls, and polybrominated diphenyl ethers (note: the correct Chinese term for PBDE is “polybrominated diphenyl ethers”; “polybrominated biphenyl ethers” is an incorrect designation)—and specifically stipulates that the cadmium content must not exceed 0.01%.
The latest RoHS Directive has added four new phthalate‑ester testing items: DEHP—di(2‑ethylhexyl) phthalate; BBP—butyl benzyl phthalate; DBP—dibutyl phthalate; and DIBP—diisobutyl phthalate.
What are the limit values for the ten hazardous substances under RoHS?
1. Cadmium: less than 100 ppm.
2. Lead: less than 1000 ppm.
3. Mercury: less than 1000 ppm.
4. Hexavalent chromium: less than 1000 ppm.
5. Polybrominated biphenyls (PBB): less than 1000 ppm.
6. Polybrominated diphenyl ethers (PBDEs): less than 1000 ppm.
7. Di(2-ethylhexyl) phthalate (DEHP): less than 1000 ppm.
8. Butyl benzyl phthalate (BBP): less than 1000 ppm.
9. Dibutyl phthalate (DBP): less than 1000 ppm.
10. Diisobutyl phthalate (DIBP): less than 1000 ppm.
What instruments are typically used for RoHS testing?
The commonly used analytical instrument for RoHS 1.0 is X-ray fluorescence spectroscopy (XRF). XRF analyzers are available in standalone, benchtop, and handheld configurations. Typically, a handheld XRF analyzer is employed first for portable on-site testing, with a focus on components of the product that pose a higher risk of containing restricted substances.
With the introduction of RoHS 2.0, four phthalate esters have been added to the RoHS Directive, necessitating additional testing to determine the levels of these compounds, which are extracted using solvents. The solvent extracts are then analyzed for the presence of phthalates by gas chromatography–mass spectrometry (GC/MS) or gas chromatography with flame ionization detection (GC/FID). Other elements and polybrominated diphenyl ethers/alkylated elastomers are analyzed using comparative methods and GC/MS, respectively.
Do the ten RoHS items require two instruments for testing?
The ten items specified in RoHS require two instruments for testing.
For the ten‑item RoHS testing, two instruments are required to complete the analysis; a single instrument cannot achieve this. Even if one instrument is used to measure heavy metal content, the other must be employed to analyze organic compounds.
X-ray fluorescence spectrometers are used to detect RoHS 1.0, while chromatographs, pyrolyzers, or hyphenated techniques such as GC–MS are employed to analyze the four phthalates specified in RoHS 2.0. At present, no single instrument can independently perform all the tests required for RoHS 2.0.
Can ROHS testing instruments detect metallic elements?
ROHS testing instruments are specialized devices used to detect the presence of hazardous substances in electronic components, household appliances, toys, and other products. These harmful substances can pose long-term risks to both human health and the environment, making the use of ROHS testing equipment of paramount importance.
Regarding whether ROHS testing instruments can detect metallic elements, the answer is yes. This is because ROHS testing covers not only hazardous substances but also the analysis of metallic constituents in compliant products. These metals include lead, cadmium, mercury, and others, which are widely used during the manufacturing process. Although these elements are essential, their excessive use poses significant risks to human health and the environment. Consequently, the advent of ROHS testing instruments has greatly facilitated the detection of metallic elements.