Installation Methods and Requirements for Gas Chromatographs!
发布时间: 2022-09-19 11:04
Gas chromatograph As an analytical instrument, understanding and mastering the installation procedure for a gas chromatograph is essential before use. Below, we’ll outline the installation methods and requirements—interested parties are encouraged to take a look!
1. Requirements for the Gas Chromatography Laboratory
(1) The analytical room shall be free of strong magnetic fields, flammable gases, and corrosive gases.
(2) The indoor ambient temperature should be maintained between 5°C and 35°C, with relative humidity not exceeding 80%, and adequate ventilation must be ensured. Factories with the necessary resources are advised to install air-conditioning systems.
(3) Prepare a work platform capable of supporting the entire instrument set, with appropriate width and height to facilitate operation. In most factories, a concrete platform (0.6–0.8 m high) is preferred. The platform should be positioned 0.8–1.0 m away from the wall to allow for convenient wiring and maintenance.
(4) The power cord used for the instrument should have a capacity of at least 3 kVA, and the instrument’s power supply must not share a circuit with high‑power equipment or devices that experience frequent and substantial load fluctuations. The power supply must be properly grounded; typically, a copper rod or wire approximately 0.5 to 1.0 meters in length is driven into moist soil (or immersed in a saline solution), and the power supply’s grounding terminal is connected to it. In short, the grounding resistance should be less than 1 ohm. (Note: It is recommended to ground both the power supply and the instrument’s enclosure for improved performance.)
2. Preparation and Purification of the Gas Source
(1) Prepare in advance the high-pressure gas cylinders required for operation (typically available in large and medium-sized cities). Each cylinder is designed to contain only one specific gas, with the color of the cylinder indicating its contents; cylinders must not be interchanged. Commonly used gases include nitrogen, hydrogen, and air, and it is advisable to have two cylinders of each type. Some facilities may use hydrogen generators and oil-free air compressors; however, the air compressor must be oil-free. When the cylinder pressure drops to 1–2 MPa, the cylinder should be replaced. In general, manufacturers can only use gases of 99.99% purity; for electron capture detectors, a high-purity gas source of 99.999% or higher is required.
(2) Gas‑source purification: To remove moisture, particulate matter, and organic contaminants that may be present in various gases, the gas must undergo rigorous purification before entering the instrument. If cylinder‑supplied gases are used, some chromatographs are equipped with purifiers containing 5A molecular sieves, activated carbon, and silica gel, which generally suffice to meet the requirements. When using a standard hydrogen generator, enhanced water‑removal purification is necessary; therefore, the drying tube should have a larger internal volume (at least 450 cm³), filled with 5A molecular sieves, and a larger gas‑storage vessel should be connected downstream to minimize or eliminate the impact of fluctuations in gas‑supply pressure on the instrument’s baseline. If an air compressor is employed as the gas source, an additional air filter should be installed at the compressor’s inlet, and the capacity of the purification line should be increased. The drying tube should be packed halfway with 5A molecular sieves and halfway with activated carbon. Typically, domestically produced oil‑free air compressors can adequately meet these needs.
3. Inspect the gas chromatograph for completeness and unpack the instrument. Verify each item against the checklist in the documentation package, and properly store spare parts for vulnerable components. Then, place the instrument on the workbench according to the instructions in the user manual, connect its various modules using the wiring diagram and appropriate plugs and sockets, and finally link it to the chromatography workstation or data processor. Exercise caution to avoid incorrect connector connections.
4. Connection of the External Gas Line
(1) Installation of pressure regulators: Some instruments are equipped with pressure regulators; if not, you will need to purchase them. Use two oxygen pressure regulators and one hydrogen pressure regulator. Install two oxygen regulators on the nitrogen and air cylinders, and one hydrogen regulator on the hydrogen cylinder. (Note that the hydrogen regulator has a reverse thread; seal the connection with the supplied O‑ring gasket.) After tightening the nuts, close the regulator’s adjustment knob (i.e., loosen it), then open the cylinder’s high‑pressure valve. At this point, the regulator’s high‑pressure gauge should display a reading. Once the high‑pressure valve is closed, the indicated pressure should remain stable; otherwise, there is a leak, which must be promptly addressed (by replacing the gasket or re‑tightening). Finally, turn the adjustment knob to vent any residual air.
(2) The gas in the cylinder is introduced into the chromatograph via an external gas‑line connection. Stainless steel tubing (φ2 × 0.5 mm) and pressure‑rated plastic tubing (φ3 × 0.5 mm) are used. Plastic tubing is easier to handle, so it is typically preferred. When using plastic tubing, the fittings should be equipped with a stainless steel liner (φ2 × 20 mm) and appropriate sealing elements. The length of the plastic tubing between the cylinder and the instrument should be determined as needed and kept as short as possible. Finally, connect the gas source to the instrument’s inlet using the plastic tubing.
(3) External gas‑line leak check: Close the valves for carrier gas, hydrogen, and air on the instrument’s gas‑supply panel, then open the high‑pressure valves on each cylinder. Adjust the low‑pressure gauge on the pressure regulator to set the carrier gas, air, and hydrogen pressures at 0.3–0.4 MPa, 0.2–0.25 MPa, respectively, and finally close the high‑pressure valves. At this point, the low‑pressure gauge on the regulator should remain stable; if it drops, there is a leak in the gas‑line system, which must be located and corrected.
5. Gas-tightness check of the chromatograph. This is a critically important procedure. If there is a leak in the gas‑line system, it can not only cause instrument instability or reduced sensitivity but also pose an explosion hazard. Therefore, this check must be performed before operation (typically focusing on the carrier‑gas line; hydrogen and air lines may be left untested if they have not been disconnected). To perform the test, open the column head, remove the column from the detector, seal the column inlet, then open the carrier‑gas supply and adjust the low‑pressure output to 0.3–0.4 MPa. Next, turn on the carrier‑gas knob on the instrument panel; the pressure gauge should register a stable reading. Finally, close the carrier‑gas knob and verify that the pressure upstream of the column does not drop over a period of 30 minutes. Any noticeable drop indicates a leak, which must be addressed. If leaks are detected within the instrument’s internal gas‑line system, remove the relevant side panels and use soapy water—preferably a sodium dodecyl sulfate solution—to systematically inspect each connection for leaks (this applies to hydrogen and air lines as well). Lastly, wipe away any residual soapy water.
6. Instrument commissioning: Connect the gas lines and the instrument, etc. After completing these steps and positioning the equipment, proceed with the following checks and adjustments.
1. Before starting the instrument and inspecting the chromatographic circuit components, first open the carrier gas flow path and adjust the carrier gas knob (i.e., the carrier gas flow‑rate regulator) on the host panel.
(1) Start the instrument and turn on its main power switch. The motor inside the column oven will begin to operate; check for any unusual noises. If such noises are detected, immediately disconnect the power and perform further troubleshooting to identify and resolve the issue. Some chromatographs perform a self‑diagnostic upon startup and display the instrument’s operating status—normal or abnormal. An abnormal status may indicate which component is faulty, wiring errors, or other problems.
(2) According to the instruction manual, verify one by one whether the column temperature (including temperature programming), the injector (vaporization chamber) temperature, and the detector temperature can remain stable at high, medium, and low settings; in particular, the temperature control accuracy of the column should reach 0.01 °C.
The above is an introduction to the installation methods and requirements for gas chromatographs. If you would like to learn more about… Gas Chromatography Instrument Information Yes, you can continue browsing our website or contact us directly!
Latest News
Recommended Products
Frequently Asked Questions
What is RoHS?
RoHS is a mandatory standard established by EU legislation, officially known as the “Directive on the Restriction of the Use of Certain Hazardous Substances in Electrical and Electronic Equipment.” The standard entered into force on July 1, 2006, and primarily aims to regulate the materials and manufacturing processes of electrical and electronic products, thereby promoting human health and environmental protection.
The purpose of this standard is to eliminate six substances—lead, mercury, cadmium, hexavalent chromium, polybrominated biphenyls, and polybrominated diphenyl ethers (note: the correct Chinese term for PBDE is “polybrominated diphenyl ethers”; “polybrominated biphenyl ethers” is an incorrect designation)—and specifically stipulates that the cadmium content must not exceed 0.01%.
The latest RoHS Directive has added four new phthalate‑ester testing items: DEHP—di(2‑ethylhexyl) phthalate; BBP—butyl benzyl phthalate; DBP—dibutyl phthalate; and DIBP—diisobutyl phthalate.
What are the limit values for the ten hazardous substances under RoHS?
1. Cadmium: less than 100 ppm.
2. Lead: less than 1000 ppm.
3. Mercury: less than 1000 ppm.
4. Hexavalent chromium: less than 1000 ppm.
5. Polybrominated biphenyls (PBB): less than 1000 ppm.
6. Polybrominated diphenyl ethers (PBDEs): less than 1000 ppm.
7. Di(2-ethylhexyl) phthalate (DEHP): less than 1000 ppm.
8. Butyl benzyl phthalate (BBP): less than 1000 ppm.
9. Dibutyl phthalate (DBP): less than 1000 ppm.
10. Diisobutyl phthalate (DIBP): less than 1000 ppm.
What instruments are typically used for RoHS testing?
The commonly used analytical instrument for RoHS 1.0 is X-ray fluorescence spectroscopy (XRF). XRF analyzers are available in standalone, benchtop, and handheld configurations. Typically, a handheld XRF analyzer is employed first for portable on-site testing, with a focus on components of the product that pose a higher risk of containing restricted substances.
With the introduction of RoHS 2.0, four phthalate esters have been added to the RoHS Directive, necessitating additional testing to determine the levels of these compounds, which are extracted using solvents. The solvent extracts are then analyzed for the presence of phthalates by gas chromatography–mass spectrometry (GC/MS) or gas chromatography with flame ionization detection (GC/FID). Other elements and polybrominated diphenyl ethers/alkylated elastomers are analyzed using comparative methods and GC/MS, respectively.
Do the ten RoHS items require two instruments for testing?
The ten items specified in RoHS require two instruments for testing.
For the ten‑item RoHS testing, two instruments are required to complete the analysis; a single instrument cannot achieve this. Even if one instrument is used to measure heavy metal content, the other must be employed to analyze organic compounds.
X-ray fluorescence spectrometers are used to detect RoHS 1.0, while chromatographs, pyrolyzers, or hyphenated techniques such as GC–MS are employed to analyze the four phthalates specified in RoHS 2.0. At present, no single instrument can independently perform all the tests required for RoHS 2.0.
Can ROHS testing instruments detect metallic elements?
ROHS testing instruments are specialized devices used to detect the presence of hazardous substances in electronic components, household appliances, toys, and other products. These harmful substances can pose long-term risks to both human health and the environment, making the use of ROHS testing equipment of paramount importance.
Regarding whether ROHS testing instruments can detect metallic elements, the answer is yes. This is because ROHS testing covers not only hazardous substances but also the analysis of metallic constituents in compliant products. These metals include lead, cadmium, mercury, and others, which are widely used during the manufacturing process. Although these elements are essential, their excessive use poses significant risks to human health and the environment. Consequently, the advent of ROHS testing instruments has greatly facilitated the detection of metallic elements.