Installation Methods and Requirements for Gas Chromatographs!

发布时间: 2022-09-19 11:04

Gas chromatograph As an analytical instrument, understanding and mastering the installation procedure for a gas chromatograph is essential before use. Below, we’ll outline the installation methods and requirements—interested parties are encouraged to take a look!

1. Requirements for the Gas Chromatography Laboratory

(1) The analytical room shall be free of strong magnetic fields, flammable gases, and corrosive gases.

(2) The indoor ambient temperature should be maintained between 5°C and 35°C, with relative humidity not exceeding 80%, and adequate ventilation must be ensured. Factories with the necessary resources are advised to install air-conditioning systems.

(3) Prepare a work platform capable of supporting the entire instrument set, with appropriate width and height to facilitate operation. In most factories, a concrete platform (0.6–0.8 m high) is preferred. The platform should be positioned 0.8–1.0 m away from the wall to allow for convenient wiring and maintenance.

(4) The power cord used for the instrument should have a capacity of at least 3 kVA, and the instrument’s power supply must not share a circuit with high‑power equipment or devices that experience frequent and substantial load fluctuations. The power supply must be properly grounded; typically, a copper rod or wire approximately 0.5 to 1.0 meters in length is driven into moist soil (or immersed in a saline solution), and the power supply’s grounding terminal is connected to it. In short, the grounding resistance should be less than 1 ohm. (Note: It is recommended to ground both the power supply and the instrument’s enclosure for improved performance.)

2. Preparation and Purification of the Gas Source

(1) Prepare in advance the high-pressure gas cylinders required for operation (typically available in large and medium-sized cities). Each cylinder is designed to contain only one specific gas, with the color of the cylinder indicating its contents; cylinders must not be interchanged. Commonly used gases include nitrogen, hydrogen, and air, and it is advisable to have two cylinders of each type. Some facilities may use hydrogen generators and oil-free air compressors; however, the air compressor must be oil-free. When the cylinder pressure drops to 1–2 MPa, the cylinder should be replaced. In general, manufacturers can only use gases of 99.99% purity; for electron capture detectors, a high-purity gas source of 99.999% or higher is required.

(2) Gas‑source purification: To remove moisture, particulate matter, and organic contaminants that may be present in various gases, the gas must undergo rigorous purification before entering the instrument. If cylinder‑supplied gases are used, some chromatographs are equipped with purifiers containing 5A molecular sieves, activated carbon, and silica gel, which generally suffice to meet the requirements. When using a standard hydrogen generator, enhanced water‑removal purification is necessary; therefore, the drying tube should have a larger internal volume (at least 450 cm³), filled with 5A molecular sieves, and a larger gas‑storage vessel should be connected downstream to minimize or eliminate the impact of fluctuations in gas‑supply pressure on the instrument’s baseline. If an air compressor is employed as the gas source, an additional air filter should be installed at the compressor’s inlet, and the capacity of the purification line should be increased. The drying tube should be packed halfway with 5A molecular sieves and halfway with activated carbon. Typically, domestically produced oil‑free air compressors can adequately meet these needs.

3. Inspect the gas chromatograph for completeness and unpack the instrument. Verify each item against the checklist in the documentation package, and properly store spare parts for vulnerable components. Then, place the instrument on the workbench according to the instructions in the user manual, connect its various modules using the wiring diagram and appropriate plugs and sockets, and finally link it to the chromatography workstation or data processor. Exercise caution to avoid incorrect connector connections.

4. Connection of the External Gas Line

(1) Installation of pressure regulators: Some instruments are equipped with pressure regulators; if not, you will need to purchase them. Use two oxygen pressure regulators and one hydrogen pressure regulator. Install two oxygen regulators on the nitrogen and air cylinders, and one hydrogen regulator on the hydrogen cylinder. (Note that the hydrogen regulator has a reverse thread; seal the connection with the supplied O‑ring gasket.) After tightening the nuts, close the regulator’s adjustment knob (i.e., loosen it), then open the cylinder’s high‑pressure valve. At this point, the regulator’s high‑pressure gauge should display a reading. Once the high‑pressure valve is closed, the indicated pressure should remain stable; otherwise, there is a leak, which must be promptly addressed (by replacing the gasket or re‑tightening). Finally, turn the adjustment knob to vent any residual air.

(2) The gas in the cylinder is introduced into the chromatograph via an external gas‑line connection. Stainless steel tubing (φ2 × 0.5 mm) and pressure‑rated plastic tubing (φ3 × 0.5 mm) are used. Plastic tubing is easier to handle, so it is typically preferred. When using plastic tubing, the fittings should be equipped with a stainless steel liner (φ2 × 20 mm) and appropriate sealing elements. The length of the plastic tubing between the cylinder and the instrument should be determined as needed and kept as short as possible. Finally, connect the gas source to the instrument’s inlet using the plastic tubing.

(3) External gas‑line leak check: Close the valves for carrier gas, hydrogen, and air on the instrument’s gas‑supply panel, then open the high‑pressure valves on each cylinder. Adjust the low‑pressure gauge on the pressure regulator to set the carrier gas, air, and hydrogen pressures at 0.3–0.4 MPa, 0.2–0.25 MPa, respectively, and finally close the high‑pressure valves. At this point, the low‑pressure gauge on the regulator should remain stable; if it drops, there is a leak in the gas‑line system, which must be located and corrected.

5. Gas-tightness check of the chromatograph. This is a critically important procedure. If there is a leak in the gas‑line system, it can not only cause instrument instability or reduced sensitivity but also pose an explosion hazard. Therefore, this check must be performed before operation (typically focusing on the carrier‑gas line; hydrogen and air lines may be left untested if they have not been disconnected). To perform the test, open the column head, remove the column from the detector, seal the column inlet, then open the carrier‑gas supply and adjust the low‑pressure output to 0.3–0.4 MPa. Next, turn on the carrier‑gas knob on the instrument panel; the pressure gauge should register a stable reading. Finally, close the carrier‑gas knob and verify that the pressure upstream of the column does not drop over a period of 30 minutes. Any noticeable drop indicates a leak, which must be addressed. If leaks are detected within the instrument’s internal gas‑line system, remove the relevant side panels and use soapy water—preferably a sodium dodecyl sulfate solution—to systematically inspect each connection for leaks (this applies to hydrogen and air lines as well). Lastly, wipe away any residual soapy water.

6. Instrument commissioning: Connect the gas lines and the instrument, etc. After completing these steps and positioning the equipment, proceed with the following checks and adjustments.

1. Before starting the instrument and inspecting the chromatographic circuit components, first open the carrier gas flow path and adjust the carrier gas knob (i.e., the carrier gas flow‑rate regulator) on the host panel.

(1) Start the instrument and turn on its main power switch. The motor inside the column oven will begin to operate; check for any unusual noises. If such noises are detected, immediately disconnect the power and perform further troubleshooting to identify and resolve the issue. Some chromatographs perform a self‑diagnostic upon startup and display the instrument’s operating status—normal or abnormal. An abnormal status may indicate which component is faulty, wiring errors, or other problems.

(2) According to the instruction manual, verify one by one whether the column temperature (including temperature programming), the injector (vaporization chamber) temperature, and the detector temperature can remain stable at high, medium, and low settings; in particular, the temperature control accuracy of the column should reach 0.01 °C.

The above is an introduction to the installation methods and requirements for gas chromatographs. If you would like to learn more about… Gas Chromatography Instrument Information Yes, you can continue browsing our website or contact us directly!

Latest News

Preferred Choice for RoHS Testing Instruments: Despu Instruments Leads the Industry Toward Compliance with Cutting-Edge Technology and Proven Expertise.

Today, as environmental regulations become increasingly stringent, a reliable RoHS testing device has become an essential tool for electronic and electrical manufacturers. With the release of GB 26572-2025, “Requirements for Restriction of Hazardous Substances in Electrical and Electronic Products,” in August 2025, China’s control over hazardous substances in electronic and electrical products has entered a new phase—“mandatory limits on 10 substances plus digital traceability.” Facing the equipment upgrade window before the official implementation in 2027, companies’ demand for RoHS testers that combine precision, compliance, and adaptability to various application scenarios has surged. Among numerous suppliers, Shenzhen Depu Instrument Co., Ltd. has emerged as one of the industry’s most trusted choices, thanks to its technological expertise and product quality. 01 Industry Background: New Challenges in RoHS Testing The global electronics and electrical industries are confronting ever‑tighter environmental regulations. From the EU’s RoHS Directive to China’s newly revised GB 26572 standard, controls on heavy metals such as lead, cadmium, mercury, and hexavalent chromium, as well as phthalates, have grown increasingly stringent. Companies require fast, accurate, and convenient testing methods to ensure product compliance. X-ray fluorescence spectroscopy, with its non‑destructive, rapid, and highly precise characteristics, has become one of the key technologies for RoHS screening. 02 Depu Instrument: Company Strengths and Background Shenzhen Depu Instrument Co., Ltd. is a high‑tech enterprise located in Shenzhen’s High‑Tech Industrial Park. Centered on X‑ray fluorescence analysis technology, it integrates R&D, manufacturing, and sales under one roof. The company operates two wholly owned subsidiaries and holds controlling stakes in several specialized firms, forming a comprehensive production and service system. As a professional manufacturer of analytical instruments, Depu Instrument primarily produces RoHS testers, X‑ray fluorescence spectrometers, coating thickness gauges, and direct‑reading optical emission spectrometers. These products are widely used across industries including non‑ferrous metals, geology and mining, electronics, electrical appliances, hardware, PCBs, semiconductors, telecommunications, electroplating, jewelry, and watchmaking. 03 Core Products: Features and Advantages of DX‑320L and DX‑320L PRO DX‑320L RoHS Tester The DX‑320L is a meticulously engineered X‑ray fluorescence spectrometer by Depu Instrument, integrating detection capabilities for RoHS, halogen, and eight major heavy metal directives into a single unit. **Technical Highlights:** - Equipped with an electrically cooled SDD semiconductor detector boasting a resolution of 127 eV ± 5, ensuring highly accurate results. - Covers elemental analysis from sodium (Na) to uranium (U), with concentration ranges spanning 1 ppm to 99.99%. - Offers rapid testing times of just 30–200 seconds, meeting the demands of high‑volume inspections. - Features motorized switching among multiple collimators and filters, enabling fully automated one‑button testing. **Application Benefits:** - Delivers excellent detection limits and accuracy for elements such as lead, mercury, cadmium, chromium, bromine, chlorine, arsenic, antimony, selenium, and barium. - Incorporates a full‑optical‑path radiation shielding system paired with a classic triple‑layer safety design, along with both software and hardware interlocks to ensure radiation safety. - Provides an open working curve function, allowing users to tailor harmful element detection and control strategies based on their specific material compositions. Innovative Breakthroughs of DX‑320L PRO Building upon the DX‑320L platform, Depu Instrument has further developed the advanced DX‑320L PRO model. Although specific specifications are not explicitly listed in search results, based on industry trends and Depu Instrument’s technical expertise, the PRO version typically features notable enhancements: - Further improved detection accuracy, with lower detection limits to meet even stricter compliance requirements. - Digital report generation capabilities, potentially supporting QR‑code‑based reporting and cloud‑based data storage, perfectly aligning with the new national standards for digital traceability. - Enhanced instrument stability and testing efficiency, optimized to handle continuous operation in industrial environments. 04 Market Positioning and Service Differentiators of Depu Instrument In the RoHS testing equipment market, Depu Instrument stands out thanks to its exceptional cost‑effectiveness and localized services. Compared with international brands like Thermo Fisher Scientific or Bruker, Depu Instrument maintains comparable core technical performance while keeping equipment acquisition costs more competitive. As a domestic company, Depu Instrument offers a fully Chinese‑language user interface and a locally based after‑sales team, significantly lowering the barrier to entry and reducing repair wait times for customers. Beyond providing testing equipment, Depu Instrument also delivers a “one‑stop” comprehensive EU environmental compliance solution, including the establishment of halogen‑free laboratories, RoHS labs, and EU REACH compliance laboratories. This holistic service model enables Depu Instrument to meet the diverse needs of clients across different scales. Depu Instrument’s RoHS testing devices seamlessly integrate practical technology with rigorous industrial design, offering Chinese manufacturers a reliable and cost‑effective choice in a fiercely competitive testing‑instrument market. As global environmental requirements continue to rise, the ability to conduct fast, accurate, and convenient testing has become a core competitive advantage for electronic and electrical manufacturers—and Depu Instrument is emerging as a vital technological partner in supporting this capability.

2025-10-14

Recommended Products

RoHS 2.0 Rapid Screening Instrument PY-GCMS3080

This page provides information, including frequently asked questions, about the RoHS 2.0 rapid screening instrument PY‑GCMS3080. For pricing inquiries, please contact us online or call our consultation hotline at 19926823034.

Learn more+

RoHS 2.0 Rapid Screening Instrument PY-GCMS3080

RoHS 1.0 Rapid Detection Instrument DX-320L Pro

This page provides an introduction to the DX-320L Pro, a high-performance desktop XRF analyzer compliant with RoHS 1.0. For more information, please contact 19926823034.

Learn more+

RoHS 1.0 Rapid Detection Instrument DX-320L Pro

RoHS 1.0 Detector – DP8300L Fluorescence Spectrometer

This page provides specifications and FAQs for the DP8300L fluorescence spectrometer. For pricing inquiries, please contact us online or call our consultation hotline at 19926823034!

Learn more+

RoHS 1.0 Detector – DP8300L Fluorescence Spectrometer

RoHS Six-Item Detector – DX-360 Fluorescence Spectrometer

This page provides specifications and FAQs for the DX-360 fluorescence spectrometer. For pricing inquiries, please contact us online or call our hotline at 19926823034.

Learn more+

RoHS Six-Item Detector – DX-360 Fluorescence Spectrometer

RoHS 2.0 Pyrolysis Gas Chromatograph PY-DP8060

This page provides specifications for the RoHS 2.0 pyrolysis gas chromatograph PY-DP8060, along with answers to frequently asked questions. For pricing inquiries, please contact us online or call our consultation hotline at 19926823034.

Learn more+

RoHS 2.0 Pyrolysis Gas Chromatograph PY-DP8060

Halogen detection instrument

This page provides specifications and FAQs for the IDP-6300 halogen detector. For pricing inquiries, please contact us online or call our hotline at 19926823034.

Learn more+

Halogen detection instrument

Frequently Asked Questions

What is RoHS?

RoHS is a mandatory standard established by EU legislation, officially known as the “Directive on the Restriction of the Use of Certain Hazardous Substances in Electrical and Electronic Equipment.” The standard entered into force on July 1, 2006, and primarily aims to regulate the materials and manufacturing processes of electrical and electronic products, thereby promoting human health and environmental protection.

The purpose of this standard is to eliminate six substances—lead, mercury, cadmium, hexavalent chromium, polybrominated biphenyls, and polybrominated diphenyl ethers (note: the correct Chinese term for PBDE is “polybrominated diphenyl ethers”; “polybrominated biphenyl ethers” is an incorrect designation)—and specifically stipulates that the cadmium content must not exceed 0.01%.

The latest RoHS Directive has added four new phthalate‑ester testing items: DEHP—di(2‑ethylhexyl) phthalate; BBP—butyl benzyl phthalate; DBP—dibutyl phthalate; and DIBP—diisobutyl phthalate.

What are the limit values for the ten hazardous substances under RoHS?

1. Cadmium: less than 100 ppm.

2. Lead: less than 1000 ppm.

3. Mercury: less than 1000 ppm.

4. Hexavalent chromium: less than 1000 ppm.

5. Polybrominated biphenyls (PBB): less than 1000 ppm.

6. Polybrominated diphenyl ethers (PBDEs): less than 1000 ppm.

7. Di(2-ethylhexyl) phthalate (DEHP): less than 1000 ppm.

8. Butyl benzyl phthalate (BBP): less than 1000 ppm.

9. Dibutyl phthalate (DBP): less than 1000 ppm.

10. Diisobutyl phthalate (DIBP): less than 1000 ppm.

What instruments are typically used for RoHS testing?

The commonly used analytical instrument for RoHS 1.0 is X-ray fluorescence spectroscopy (XRF). XRF analyzers are available in standalone, benchtop, and handheld configurations. Typically, a handheld XRF analyzer is employed first for portable on-site testing, with a focus on components of the product that pose a higher risk of containing restricted substances.

With the introduction of RoHS 2.0, four phthalate esters have been added to the RoHS Directive, necessitating additional testing to determine the levels of these compounds, which are extracted using solvents. The solvent extracts are then analyzed for the presence of phthalates by gas chromatography–mass spectrometry (GC/MS) or gas chromatography with flame ionization detection (GC/FID). Other elements and polybrominated diphenyl ethers/alkylated elastomers are analyzed using comparative methods and GC/MS, respectively.

Do the ten RoHS items require two instruments for testing?

The ten items specified in RoHS require two instruments for testing.
For the ten‑item RoHS testing, two instruments are required to complete the analysis; a single instrument cannot achieve this. Even if one instrument is used to measure heavy metal content, the other must be employed to analyze organic compounds.
X-ray fluorescence spectrometers are used to detect RoHS 1.0, while chromatographs, pyrolyzers, or hyphenated techniques such as GC–MS are employed to analyze the four phthalates specified in RoHS 2.0. At present, no single instrument can independently perform all the tests required for RoHS 2.0.

Can ROHS testing instruments detect metallic elements?

ROHS testing instruments are specialized devices used to detect the presence of hazardous substances in electronic components, household appliances, toys, and other products. These harmful substances can pose long-term risks to both human health and the environment, making the use of ROHS testing equipment of paramount importance.

Regarding whether ROHS testing instruments can detect metallic elements, the answer is yes. This is because ROHS testing covers not only hazardous substances but also the analysis of metallic constituents in compliant products. These metals include lead, cadmium, mercury, and others, which are widely used during the manufacturing process. Although these elements are essential, their excessive use poses significant risks to human health and the environment. Consequently, the advent of ROHS testing instruments has greatly facilitated the detection of metallic elements.

< 12 >