Technical principles of metrological calibration for RoHS testers!
发布时间: 2022-09-19 10:12
RoHS tester It is an instrument used to meet the EU’s RoHS testing standards. Simply put, it performs metrological calibration of devices that detect six hazardous substances: lead (Pb), cadmium (Cd), mercury (Hg), hexavalent chromium (Cr6+), polybrominated diphenyl ethers (PBDEs), and polybrominated biphenyls (PBBs). Currently, the most common type of RoHS tester on the market is the X-ray fluorescence analyzer, which comes in two variants: energy-dispersive and wavelength-dispersive. Due to its simpler technical principles and structure compared to wavelength-dispersive models, the energy-dispersive type is more prevalent today. Now, let’s explain the underlying technical principles behind the metrological calibration of RoHS testers.
I. Characteristics of X-rays
X-rays or gamma rays emitted by a radioactive nuclide source or an X-ray generator interact with the atoms in the sample, ejecting inner-shell electrons. When the outer-shell electrons are left behind, they emit characteristic X-rays whose energy corresponds to the original kinetic energy of the ejected inner-shell electrons.
II. Correlation Between Elements and Characteristic X-ray Compressive Strength
X-ray energies vary among different elements, enabling qualitative analysis. Based on the characteristic X-ray intensities, qualitative identification can be performed. This can be expressed as: C = f(k1I1, k2I2, k3I3, ...), where Kn (n = 1, 2, 3, ...) denotes the undetermined coefficient for the nth element, and In (n = 1, 2, 3, ...) represents the characteristic X-ray intensity emitted by element n. Thus, qualitative elemental analysis in a chemical sample can only be carried out after the coefficients Kn have been determined through calibration.
Throughout the entire lifecycle of electronic information technology products—manufacturing, sales, and import—within the territory of the People’s Republic of China, this approach can be employed to manage and mitigate environmental pollution and other ecological risks posed by such products. However, it does not apply to the manufacturing of exported goods. The RoHS testing instrument is designed to more effectively regulate the levels of harmful heavy metals in products manufactured by companies. For example, central air conditioners, automatic washing machines, vacuum cleaners, electric water heaters, and mobile phones are commonly encountered in everyday life; consequently, heavy‑metal contamination can pose serious health risks. It is therefore recommended that manufacturers strictly control the content of heavy metals exceeding regulatory limits, thereby safeguarding the physical and mental well‑being of consumers.
III. Principle of Analysis
An ROHS tester is an X-ray fluorescence spectrometer, and its analytical principle is based on that of X-ray fluorescence spectroscopy. X-ray fluorescence spectrometers are generally classified into two types: wavelength-dispersive X-ray fluorescence spectrometers (WDXRF) and energy-dispersive X-ray fluorescence spectrometers (EDXRF). The key components of a wavelength-dispersive spectrometer include the excitation source, the spectrometer’s crystal and goniometer, and the detector, whereas an energy-dispersive spectrometer requires only an excitation source, a detector, and associated electronic components and a control system, making it relatively simpler.
Wavelength-dispersive X-ray fluorescence spectrometers use analyzing crystals to identify the analytical spectral lines of the elements being measured. Based on Bragg’s law and the measured diffraction angle, the wavelength of the element’s spectral line can be determined:
nλ = 2d sin θ (n = 1, 2, 3, …), where λ is the wavelength of the spectral line under analysis; d is the interplanar spacing of the crystal lattice; θ is the diffraction angle; and n is the order of diffraction. By using a goniometer to measure the diffraction angle of the spectral line of interest, one can determine the wavelength of the element being analyzed via the above equation, thereby obtaining characteristic information about the element in question.
Energy-dispersive X-ray fluorescence spectrometers employ kinetic-energy detectors; based on the charge collected by the detector, the kinetic energy of the characteristic X-rays emitted by the element under analysis can be determined in real time.
Q = kE, where k is the photon energy of the emitted radiation; q is the relative charge induced by the detector; and k is a key response parameter for various kinetic-energy detectors. The charge is positively correlated with the kinetic energy of the emitted radiation; therefore, the characteristic information of the element under analysis can be derived from the measured charge.
The characteristic spectral lines of the element under analysis can only be obtained by selecting a specific excitation source. At present, the primary excitation sources are X-ray fluorescence tubes and radioactive isotope sources. To extract sufficient information for qualitative and quantitative analysis of the sample, in addition to the spectrometer, appropriate sample preparation techniques must be employed, along with suitable spectral analysis and data‑processing methods applied to the acquired intensity measurements.
The above is an introduction to the technical principles underlying the metrological calibration of RoHS testing instruments. If you would like to learn more about… ROHS tester information Yes, you can continue browsing our website or contact us directly online!
Latest News
Recommended Products
Frequently Asked Questions
What is RoHS?
RoHS is a mandatory standard established by EU legislation, officially known as the “Directive on the Restriction of the Use of Certain Hazardous Substances in Electrical and Electronic Equipment.” The standard entered into force on July 1, 2006, and primarily aims to regulate the materials and manufacturing processes of electrical and electronic products, thereby promoting human health and environmental protection.
The purpose of this standard is to eliminate six substances—lead, mercury, cadmium, hexavalent chromium, polybrominated biphenyls, and polybrominated diphenyl ethers (note: the correct Chinese term for PBDE is “polybrominated diphenyl ethers”; “polybrominated biphenyl ethers” is an incorrect designation)—and specifically stipulates that the cadmium content must not exceed 0.01%.
The latest RoHS Directive has added four new phthalate‑ester testing items: DEHP—di(2‑ethylhexyl) phthalate; BBP—butyl benzyl phthalate; DBP—dibutyl phthalate; and DIBP—diisobutyl phthalate.
What are the limit values for the ten hazardous substances under RoHS?
1. Cadmium: less than 100 ppm.
2. Lead: less than 1000 ppm.
3. Mercury: less than 1000 ppm.
4. Hexavalent chromium: less than 1000 ppm.
5. Polybrominated biphenyls (PBB): less than 1000 ppm.
6. Polybrominated diphenyl ethers (PBDEs): less than 1000 ppm.
7. Di(2-ethylhexyl) phthalate (DEHP): less than 1000 ppm.
8. Butyl benzyl phthalate (BBP): less than 1000 ppm.
9. Dibutyl phthalate (DBP): less than 1000 ppm.
10. Diisobutyl phthalate (DIBP): less than 1000 ppm.
What instruments are typically used for RoHS testing?
The commonly used analytical instrument for RoHS 1.0 is X-ray fluorescence spectroscopy (XRF). XRF analyzers are available in standalone, benchtop, and handheld configurations. Typically, a handheld XRF analyzer is employed first for portable on-site testing, with a focus on components of the product that pose a higher risk of containing restricted substances.
With the introduction of RoHS 2.0, four phthalate esters have been added to the RoHS Directive, necessitating additional testing to determine the levels of these compounds, which are extracted using solvents. The solvent extracts are then analyzed for the presence of phthalates by gas chromatography–mass spectrometry (GC/MS) or gas chromatography with flame ionization detection (GC/FID). Other elements and polybrominated diphenyl ethers/alkylated elastomers are analyzed using comparative methods and GC/MS, respectively.
Do the ten RoHS items require two instruments for testing?
The ten items specified in RoHS require two instruments for testing.
For the ten‑item RoHS testing, two instruments are required to complete the analysis; a single instrument cannot achieve this. Even if one instrument is used to measure heavy metal content, the other must be employed to analyze organic compounds.
X-ray fluorescence spectrometers are used to detect RoHS 1.0, while chromatographs, pyrolyzers, or hyphenated techniques such as GC–MS are employed to analyze the four phthalates specified in RoHS 2.0. At present, no single instrument can independently perform all the tests required for RoHS 2.0.
Can ROHS testing instruments detect metallic elements?
ROHS testing instruments are specialized devices used to detect the presence of hazardous substances in electronic components, household appliances, toys, and other products. These harmful substances can pose long-term risks to both human health and the environment, making the use of ROHS testing equipment of paramount importance.
Regarding whether ROHS testing instruments can detect metallic elements, the answer is yes. This is because ROHS testing covers not only hazardous substances but also the analysis of metallic constituents in compliant products. These metals include lead, cadmium, mercury, and others, which are widely used during the manufacturing process. Although these elements are essential, their excessive use poses significant risks to human health and the environment. Consequently, the advent of ROHS testing instruments has greatly facilitated the detection of metallic elements.