Operating Procedure for the Spark Direct-Reading Spectrometer!
发布时间: 2022-09-14 11:10
About Spark Direct-Reading Spectrometer Many people, upon receiving the instrument, are eager to start using it right away. Today, we’ll walk through the operating procedures for a spark‑direct‑reading optical emission spectrometer—interested readers are welcome to take a look!
I. Preparations: Turn on the power.
Power-on sequence: First turn on the printer, then the monitor, and finally the spectrometer.
II. Instrument Warm-Up
To obtain stable analytical results, the instrument’s spectroscopic chamber must be kept at a constant temperature. After powering on, allow at least 30 minutes—depending on the ambient temperature—for the system to reach thermal equilibrium before it can operate normally.
III. Preparation of Samples for Spectral Testing
1. After the workshop has melted the alloy according to the preliminary formulation, one to two samples are taken from each heat for chemical composition analysis.
2. The surface of the test specimen must be machined flat on a lathe, with a surface roughness not exceeding Ra 3.0. Edges shall be free of burrs, and the surface must be free of porosity and blowholes; furthermore, the interior must exhibit no unevenness or irregularities.
3. The surface of the test specimen must be free from contamination, fingerprints, and oxidation; otherwise, accurate analytical results cannot be obtained.
IV. Connect the Argon Gas Supply
Open the argon cylinder, adjust the pressure regulator, and set the argon outlet pressure to approximately 0.30–0.40 MPa.
V. Daily Analysis
1. Inspire
Place the prepared test specimen flat on the spark stand, ensuring that the negative electrode clamp firmly holds the sample. Enter the specimen’s relevant product information—such as furnace number, grade, and specification name—into the computer software. Then, press the instrument’s excitation button; when a test icon appears on the software screen, it indicates that the instrument is performing the analysis, and you will hear a “zzz” sound of electrode discharge inside the device. After a brief wait, the test results will be displayed on the computer screen.
2. Change a little bit
To analyze a sample, 1–3 spots must be excited, and the average of these measurements is taken as the analytical result for that sample, ensuring the accuracy of the analysis. Before changing the measurement spot, thoroughly clean the electrode with the electrode brush; to minimize the ingress of ambient air into the excitation chamber, the spot change should be performed as quickly as possible. For unsatisfactory results, you can delete the corresponding excitation data by clicking the “scissors” icon in the software.
3. Issue the report
After completing the testing and analysis of a sample, we can issue test reports in various formats, such as a “Material Analysis Report” and a “Material Certification Report,” upon request.
4. Test another sample.
Once the analysis of a test sample is complete, you can click the “New Analysis” icon in the software to begin analyzing the next sample; the subsequent steps are identical to those described earlier.
VI. Shutting Down the Spectrometer
1. If the instrument will not be used for a specified period, you can exit the software by clicking the “EXIT” icon. The software will close, returning the computer to the desktop standby mode. At this point, the instrument enters standby, with a low‑flow argon gas stream maintaining the argon atmosphere in the excitation chamber, ensuring the instrument is ready for immediate use. To resume operation, simply open the spectrometer’s testing software.
2. Shutdown sequence: First turn off the argon gas supply and the instrument’s power, then turn off the computer monitor’s power.
Precautions: When the instrument is not in use, a flat‑surface sample should also be placed over the spark‑stand excitation port to keep the excitation chamber clean and minimize air ingress.
The above is an introduction to the operating procedures for spark‑direct‑reading optical spectrometers, provided for reference only. Since spark‑direct‑reading optical spectrometers from different manufacturers may vary slightly in design and configuration, operational details could differ accordingly. If you would like to learn more… Spark Direct-Reading Spectrometer Information Yes, you can continue to visit our website, or contact us directly!
Latest News
Recommended Products
Frequently Asked Questions
What is RoHS?
RoHS is a mandatory standard established by EU legislation, officially known as the “Directive on the Restriction of the Use of Certain Hazardous Substances in Electrical and Electronic Equipment.” The standard entered into force on July 1, 2006, and primarily aims to regulate the materials and manufacturing processes of electrical and electronic products, thereby promoting human health and environmental protection.
The purpose of this standard is to eliminate six substances—lead, mercury, cadmium, hexavalent chromium, polybrominated biphenyls, and polybrominated diphenyl ethers (note: the correct Chinese term for PBDE is “polybrominated diphenyl ethers”; “polybrominated biphenyl ethers” is an incorrect designation)—and specifically stipulates that the cadmium content must not exceed 0.01%.
The latest RoHS Directive has added four new phthalate‑ester testing items: DEHP—di(2‑ethylhexyl) phthalate; BBP—butyl benzyl phthalate; DBP—dibutyl phthalate; and DIBP—diisobutyl phthalate.
What are the limit values for the ten hazardous substances under RoHS?
1. Cadmium: less than 100 ppm.
2. Lead: less than 1000 ppm.
3. Mercury: less than 1000 ppm.
4. Hexavalent chromium: less than 1000 ppm.
5. Polybrominated biphenyls (PBB): less than 1000 ppm.
6. Polybrominated diphenyl ethers (PBDEs): less than 1000 ppm.
7. Di(2-ethylhexyl) phthalate (DEHP): less than 1000 ppm.
8. Butyl benzyl phthalate (BBP): less than 1000 ppm.
9. Dibutyl phthalate (DBP): less than 1000 ppm.
10. Diisobutyl phthalate (DIBP): less than 1000 ppm.
What instruments are typically used for RoHS testing?
The commonly used analytical instrument for RoHS 1.0 is X-ray fluorescence spectroscopy (XRF). XRF analyzers are available in standalone, benchtop, and handheld configurations. Typically, a handheld XRF analyzer is employed first for portable on-site testing, with a focus on components of the product that pose a higher risk of containing restricted substances.
With the introduction of RoHS 2.0, four phthalate esters have been added to the RoHS Directive, necessitating additional testing to determine the levels of these compounds, which are extracted using solvents. The solvent extracts are then analyzed for the presence of phthalates by gas chromatography–mass spectrometry (GC/MS) or gas chromatography with flame ionization detection (GC/FID). Other elements and polybrominated diphenyl ethers/alkylated elastomers are analyzed using comparative methods and GC/MS, respectively.
Do the ten RoHS items require two instruments for testing?
The ten items specified in RoHS require two instruments for testing.
For the ten‑item RoHS testing, two instruments are required to complete the analysis; a single instrument cannot achieve this. Even if one instrument is used to measure heavy metal content, the other must be employed to analyze organic compounds.
X-ray fluorescence spectrometers are used to detect RoHS 1.0, while chromatographs, pyrolyzers, or hyphenated techniques such as GC–MS are employed to analyze the four phthalates specified in RoHS 2.0. At present, no single instrument can independently perform all the tests required for RoHS 2.0.
Can ROHS testing instruments detect metallic elements?
ROHS testing instruments are specialized devices used to detect the presence of hazardous substances in electronic components, household appliances, toys, and other products. These harmful substances can pose long-term risks to both human health and the environment, making the use of ROHS testing equipment of paramount importance.
Regarding whether ROHS testing instruments can detect metallic elements, the answer is yes. This is because ROHS testing covers not only hazardous substances but also the analysis of metallic constituents in compliant products. These metals include lead, cadmium, mercury, and others, which are widely used during the manufacturing process. Although these elements are essential, their excessive use poses significant risks to human health and the environment. Consequently, the advent of ROHS testing instruments has greatly facilitated the detection of metallic elements.