Calibration and Verification of Atomic Fluorescence Spectrometers!
发布时间: 2022-08-26 10:14
For hydride‑free dispersive atomic fluorescence spectrometers that use a hollow cathode lamp as the excitation source—covering single‑channel, dual‑channel, and multi‑channel instruments—calibration is typically performed in accordance with the “Atomic Fluorescence Spectrometer” standard. As for atoms… Fluorescence spectrometer Calibration and verification—let’s take a look!
1) General Requirements
① The instrument shall bear the following markings: instrument name, model number, manufacturer’s name, factory serial number and date of manufacture, as well as the mark and number of the manufacturing license, etc.;
② All fasteners on the instrument and its accessories shall be securely tightened, and moving parts shall operate smoothly and with free movement.
③ The instrument’s switches, knobs, and buttons shall function properly. For instruments controlled by a computer or microcomputer, the corresponding functions shall operate normally when commands are entered via the keyboard.
2) Calibration Items and Performance Requirements: The technical requirements that the instrument must meet during use or after maintenance are listed in the table.
3) Handling of calibration results: When all calibration items pass, affix a calibration‑pass mark; for instruments that fail calibration, suspend or restrict their use, affix a “calibration failed” mark indicating the specific failing item, and promptly report the calibration information to the manufacturer.
4) Calibration Period: The calibration period generally does not exceed one year. In this case, calibration should be performed whenever conditions change, measurement results are questionable, following fault repair or modification, or after relocation or transportation.
5) Regular verification, in accordance with the requirements of ISO/IEC 17025, “General Requirements for the Competence of Testing and Calibration Laboratories,” and CNAS‑CL01, “Guidelines for Accreditation of Testing and Calibration Laboratories,” is conducted to ensure the reliability of the testing and calibration laboratory’s capabilities. Regarding the calibration status of instruments, necessary periodic checks should be performed between successive verifications or calibration cycles.
In conjunction with mathematical statistics, the periodic verification methods applicable to atomic fluorescence spectrometers include: 1. Verification using certified reference materials or traceable standard substances that are within their validity period; 2. Verification using an atomic fluorescence spectrometer of the same accuracy class or with the same maximum permissible error; 3. Inter‑laboratory testing and comparison on the same sample; 4. Validation using stable retained samples; 5. Assessment of stability, detection limit, relative standard deviation, and other relevant parameters; 6. Use of the manufacturer’s operating manual for the atomic fluorescence spectrometer, along with the product specifications or applicable methods.
Determine whether the instrument can continue to be used based on the verification results. If the inspection results are abnormal, conduct an analysis to identify the cause, revise the inspection procedure, and add additional inspection points.
Latest News
Recommended Products
Frequently Asked Questions
What is RoHS?
RoHS is a mandatory standard established by EU legislation, officially known as the “Directive on the Restriction of the Use of Certain Hazardous Substances in Electrical and Electronic Equipment.” The standard entered into force on July 1, 2006, and primarily aims to regulate the materials and manufacturing processes of electrical and electronic products, thereby promoting human health and environmental protection.
The purpose of this standard is to eliminate six substances—lead, mercury, cadmium, hexavalent chromium, polybrominated biphenyls, and polybrominated diphenyl ethers (note: the correct Chinese term for PBDE is “polybrominated diphenyl ethers”; “polybrominated biphenyl ethers” is an incorrect designation)—and specifically stipulates that the cadmium content must not exceed 0.01%.
The latest RoHS Directive has added four new phthalate‑ester testing items: DEHP—di(2‑ethylhexyl) phthalate; BBP—butyl benzyl phthalate; DBP—dibutyl phthalate; and DIBP—diisobutyl phthalate.
What are the limit values for the ten hazardous substances under RoHS?
1. Cadmium: less than 100 ppm.
2. Lead: less than 1000 ppm.
3. Mercury: less than 1000 ppm.
4. Hexavalent chromium: less than 1000 ppm.
5. Polybrominated biphenyls (PBB): less than 1000 ppm.
6. Polybrominated diphenyl ethers (PBDEs): less than 1000 ppm.
7. Di(2-ethylhexyl) phthalate (DEHP): less than 1000 ppm.
8. Butyl benzyl phthalate (BBP): less than 1000 ppm.
9. Dibutyl phthalate (DBP): less than 1000 ppm.
10. Diisobutyl phthalate (DIBP): less than 1000 ppm.
What instruments are typically used for RoHS testing?
The commonly used analytical instrument for RoHS 1.0 is X-ray fluorescence spectroscopy (XRF). XRF analyzers are available in standalone, benchtop, and handheld configurations. Typically, a handheld XRF analyzer is employed first for portable on-site testing, with a focus on components of the product that pose a higher risk of containing restricted substances.
With the introduction of RoHS 2.0, four phthalate esters have been added to the RoHS Directive, necessitating additional testing to determine the levels of these compounds, which are extracted using solvents. The solvent extracts are then analyzed for the presence of phthalates by gas chromatography–mass spectrometry (GC/MS) or gas chromatography with flame ionization detection (GC/FID). Other elements and polybrominated diphenyl ethers/alkylated elastomers are analyzed using comparative methods and GC/MS, respectively.
Do the ten RoHS items require two instruments for testing?
The ten items specified in RoHS require two instruments for testing.
For the ten‑item RoHS testing, two instruments are required to complete the analysis; a single instrument cannot achieve this. Even if one instrument is used to measure heavy metal content, the other must be employed to analyze organic compounds.
X-ray fluorescence spectrometers are used to detect RoHS 1.0, while chromatographs, pyrolyzers, or hyphenated techniques such as GC–MS are employed to analyze the four phthalates specified in RoHS 2.0. At present, no single instrument can independently perform all the tests required for RoHS 2.0.
Can ROHS testing instruments detect metallic elements?
ROHS testing instruments are specialized devices used to detect the presence of hazardous substances in electronic components, household appliances, toys, and other products. These harmful substances can pose long-term risks to both human health and the environment, making the use of ROHS testing equipment of paramount importance.
Regarding whether ROHS testing instruments can detect metallic elements, the answer is yes. This is because ROHS testing covers not only hazardous substances but also the analysis of metallic constituents in compliant products. These metals include lead, cadmium, mercury, and others, which are widely used during the manufacturing process. Although these elements are essential, their excessive use poses significant risks to human health and the environment. Consequently, the advent of ROHS testing instruments has greatly facilitated the detection of metallic elements.