[DePu Instruments] – A manufacturer specializing in testing instruments for RoHS compliance, heavy metals, alloy analysis, halogens, and more!
Product Inquiry:19926823034(WeChat ID is the same)
- All
- Product Management
- News and Information
- Introduction Content
- Corporate outlets
- Frequently Asked Questions
- Corporate Video
- Corporate Portfolio
RoHS 2.0 instrument RoHS 1.0 instrument
Alloy analyzerCoating Analyzer
What does an X-ray fluorescence spectrometer measure?
发布时间: 2022-08-25 10:03
X-ray fluorescence spectrometer It can quantitatively analyze the elemental composition of a wide variety of samples, including pressed pellets, fused beads, powdered or liquid materials, and even large solid specimens. By employing a high‑power X‑ray tube, it achieves low detection limits and short measurement times. Precise detection of light elements is also made possible through an optimized combination of excitation, detection, and vacuum modes, resulting in low operating costs. So, what does an X‑ray fluorescence spectrometer measure? Let’s take a brief look!
X-ray fluorescence analyzer measurement range: It can measure elements with atomic numbers ranging from 9 to 92 (from fluorine, F, to uranium, U). Concentration range: ppm to 99.99%.
Test objects: uncompressed standard powders, as well as solids, liquids, and powders that can be placed in standard sample cups;
Detector: Ultra-large, ultra-thin SDD detector, 5 eV; optional high-speed detector up to 125 eV.
Excitation source: The tube can be equipped with either a rhodium target or a tungsten target, depending on customer requirements.
Logging environment: Ultra‑high vacuum system; evacuation to below 10 Pa within 10 seconds; both atmospheric air and helium can be collimated and filtered.
Environmental Management Substance Measurement Software Ver. 2
Optimize measurement time using precision management software (feature available in Ver. 1);
Use a centralized database to manage measurement data across all branches (search, browse, analyze, edit, print, and generate reports).
For the same sample that has been measured previously, the results of two measurements can be compared.
In addition to RoHS analysis, halogen analysis, toy safety testing, and analyses for antimony (Sb) and tin (Sn), you can also freely add any elements that require monitoring.
Utilizing modern electronic technology, the instrument is entirely microprocessor‑controlled. All control boards are equipped with microprocessors that manage the sample injector, the vacuum level in the spectrometer chamber, temperature, the detector, the X‑ray tube and its generator, water cooling, the goniometer, and more. A status indicator promptly reports any issues. The green mode reduces power consumption and extends the X‑ray tube’s service life: when the instrument is about to be idle for one hour or several hours, the X‑ray tube’s power automatically switches to standby within a set time interval. Upon resuming operation, the system automatically restores the required working conditions.
The above is an introduction to what X-ray fluorescence spectrometers can detect. If you have any other questions, please continue browsing its… Spectrometer Information If you have any needs for an X-ray fluorescence spectrometer, please contact us online!
Latest News
Recommended Products
Frequently Asked Questions
What is RoHS?
RoHS is a mandatory standard established by EU legislation, officially known as the “Directive on the Restriction of the Use of Certain Hazardous Substances in Electrical and Electronic Equipment.” The standard entered into force on July 1, 2006, and primarily aims to regulate the materials and manufacturing processes of electrical and electronic products, thereby promoting human health and environmental protection.
The purpose of this standard is to eliminate six substances—lead, mercury, cadmium, hexavalent chromium, polybrominated biphenyls, and polybrominated diphenyl ethers (note: the correct Chinese term for PBDE is “polybrominated diphenyl ethers”; “polybrominated biphenyl ethers” is an incorrect designation)—and specifically stipulates that the cadmium content must not exceed 0.01%.
The latest RoHS Directive has added four new phthalate‑ester testing items: DEHP—di(2‑ethylhexyl) phthalate; BBP—butyl benzyl phthalate; DBP—dibutyl phthalate; and DIBP—diisobutyl phthalate.
What are the limit values for the ten hazardous substances under RoHS?
1. Cadmium: less than 100 ppm.
2. Lead: less than 1000 ppm.
3. Mercury: less than 1000 ppm.
4. Hexavalent chromium: less than 1000 ppm.
5. Polybrominated biphenyls (PBB): less than 1000 ppm.
6. Polybrominated diphenyl ethers (PBDEs): less than 1000 ppm.
7. Di(2-ethylhexyl) phthalate (DEHP): less than 1000 ppm.
8. Butyl benzyl phthalate (BBP): less than 1000 ppm.
9. Dibutyl phthalate (DBP): less than 1000 ppm.
10. Diisobutyl phthalate (DIBP): less than 1000 ppm.
What instruments are typically used for RoHS testing?
The commonly used analytical instrument for RoHS 1.0 is X-ray fluorescence spectroscopy (XRF). XRF analyzers are available in standalone, benchtop, and handheld configurations. Typically, a handheld XRF analyzer is employed first for portable on-site testing, with a focus on components of the product that pose a higher risk of containing restricted substances.
With the introduction of RoHS 2.0, four phthalate esters have been added to the RoHS Directive, necessitating additional testing to determine the levels of these compounds, which are extracted using solvents. The solvent extracts are then analyzed for the presence of phthalates by gas chromatography–mass spectrometry (GC/MS) or gas chromatography with flame ionization detection (GC/FID). Other elements and polybrominated diphenyl ethers/alkylated elastomers are analyzed using comparative methods and GC/MS, respectively.
Do the ten RoHS items require two instruments for testing?
The ten items specified in RoHS require two instruments for testing.
For the ten‑item RoHS testing, two instruments are required to complete the analysis; a single instrument cannot achieve this. Even if one instrument is used to measure heavy metal content, the other must be employed to analyze organic compounds.
X-ray fluorescence spectrometers are used to detect RoHS 1.0, while chromatographs, pyrolyzers, or hyphenated techniques such as GC–MS are employed to analyze the four phthalates specified in RoHS 2.0. At present, no single instrument can independently perform all the tests required for RoHS 2.0.
Can ROHS testing instruments detect metallic elements?
ROHS testing instruments are specialized devices used to detect the presence of hazardous substances in electronic components, household appliances, toys, and other products. These harmful substances can pose long-term risks to both human health and the environment, making the use of ROHS testing equipment of paramount importance.
Regarding whether ROHS testing instruments can detect metallic elements, the answer is yes. This is because ROHS testing covers not only hazardous substances but also the analysis of metallic constituents in compliant products. These metals include lead, cadmium, mercury, and others, which are widely used during the manufacturing process. Although these elements are essential, their excessive use poses significant risks to human health and the environment. Consequently, the advent of ROHS testing instruments has greatly facilitated the detection of metallic elements.