Operating procedures for RoHS 2.0 instruments!
发布时间: 2022-08-23 10:06
RoHS 2.0 instrument It is a kind of RoHS testing instrument Many users are unsure how to operate it, so today we’ll walk through the operating guidelines for ROHS 2.0 instruments. If you’re interested, feel free to take a look!
1. Turn on the instrument.
1. First, check the wavelength displayed on the detector, e.g., 225; 2) This is for o‑phthalaldehyde detection; the range value is 1.
2. Pump priming: Place the suction tips of pumps A and B into the reagent bottles, and rotate the priming valves to an angle between 90° and 180°.
3. Press the “Purge” button again until no air bubbles remain in the liquid inlet tube. (Precautions for using the RoHS 2.0 analyzer: During purging, the drain valve must be open, and no liquid should flow into the inlet tube. When the drain valve is open, the liquid can be drawn out using a bulb syringe.)
II. Instrument Rinse and Column Equilibration
Every day, before starting the instrument to analyze samples, first add a measured volume of methanol to the reagent bottle, place the tips of pumps A and B into the bottle, and then follow these steps: press the Edit button on pump A, select Method 1, confirm, activate, and start the run; flush for at least 30 minutes. (Note: This is the column‑flushing procedure. If the column has already been equilibrated, you can open the software to verify that the baseline is stable and properly adjusted.) When shutting down, repeat the above steps using methanol as the flushing solvent (this step allows you to turn off the detector, thereby extending the lifespan of the UV lamp).
III. Flow Allocation
Pump A – Acetonitrile Bottle: Filter the acetonitrile reagent using a solvent filter and an organic membrane filter, then subject it to ultrasonic treatment for at least 30 minutes. Pump B – Water Bottle: Filter the purified water using a solvent filter and a water filtration membrane, then subject it to ultrasonic treatment for at least 30 minutes.
IV. Sample Pre-treatment for RoHS 2.0 Analyzers
Grind the sample into a fine powder or small particles with a diameter of less than 1 mm, and weigh out 0.2 g to an accuracy of 0.0001 g. Transfer the weighed sample into a 50 mL volumetric flask, add 40 mL of acetonitrile, and sonicate continuously for 70 minutes. After cooling to room temperature, dilute to the mark with solvent to a final volume of 50 mL, and allow the solution to equilibrate prior to analysis.
V. Software Usage
On the computer desktop, open the N2000 Online Workstation, select “Serial Port 1,” maximize the window, and on the “Data Acquisition” screen, click “View Baseline.” You will see a red area in the lower-left corner, with voltage and time values fluctuating. If these changes occur, it indicates that the computer and the instrument are not properly connected. In the method interface, when testing standard samples or actual samples, be sure to modify the file prefix before clicking “Apply.”
VI. Use of the Gradient Method
After the column is balanced, use the gradient method to fine‑tune it; invoke this method by pressing the Edit key, then the Confirm key, followed by the Activate key and the Run key. Both pumps operate simultaneously. If they do not run at the same time, the key has not been pressed—press it again. This is how you perform gradient elution and establish a baseline. The baseline run takes just over 20 minutes, and you can view its shape in the software.
VII. Testing of Standard Samples
After the gradient baseline begins running in step 6, we can flush the injection valve by rinsing it two to three times with cleaning solvent until the baseline stabilizes.
Measurement 4P example:
1) Click the “Methods” tab in the software, rename the file to “4P Standard,” then click “Apply.” Next, go to the baseline view to verify that the baseline is flat.
2) Inject the 4P mixed standard solution (using a microsyringe, at least 40 µL, ensuring no air bubbles are introduced) → Press the A/Z key to reset → Open the valve → Simultaneously press the “Run” key; (pay close attention to the pressure and flow rate changes of both pumps, and verify that the gradient elution starts normally). After completing the test, record the results.
VIII. Sample Testing with the RoHS 2.0 Analyzer
When testing samples each day, begin by analyzing the standard sample. If Step 7 has been completed, first initiate the gradient method, run the baseline, and use the cleaning needle to flush the injection valve. Once the baseline is stable, rename the file to the name of the sample being tested at that time (e.g., Sample 1), then click “Apply.” Next, use a syringe to draw up the ultrasonically treated solution prepared in Step 4, attach the filter tip (ensuring there are no air bubbles in the syringe), and inject it into the quantitative injection valve. Press A/Z to zero the system, open the valve, and simultaneously press the “Run” button. Monitor the pressure and flow rates of both pumps to confirm that the gradient program has started and is running normally. Finally, wait for the test results.
The above is an overview of the operating procedures for RoHS 2.0 instruments, provided for reference only. If you have any requirements for RoHS 2.0 instruments, please feel free to leave us a message online or contact us by phone. RoHS 2.0 instrument manufacturer -De Pu!
Latest News
Recommended Products
Frequently Asked Questions
What is RoHS?
RoHS is a mandatory standard established by EU legislation, officially known as the “Directive on the Restriction of the Use of Certain Hazardous Substances in Electrical and Electronic Equipment.” The standard entered into force on July 1, 2006, and primarily aims to regulate the materials and manufacturing processes of electrical and electronic products, thereby promoting human health and environmental protection.
The purpose of this standard is to eliminate six substances—lead, mercury, cadmium, hexavalent chromium, polybrominated biphenyls, and polybrominated diphenyl ethers (note: the correct Chinese term for PBDE is “polybrominated diphenyl ethers”; “polybrominated biphenyl ethers” is an incorrect designation)—and specifically stipulates that the cadmium content must not exceed 0.01%.
The latest RoHS Directive has added four new phthalate‑ester testing items: DEHP—di(2‑ethylhexyl) phthalate; BBP—butyl benzyl phthalate; DBP—dibutyl phthalate; and DIBP—diisobutyl phthalate.
What are the limit values for the ten hazardous substances under RoHS?
1. Cadmium: less than 100 ppm.
2. Lead: less than 1000 ppm.
3. Mercury: less than 1000 ppm.
4. Hexavalent chromium: less than 1000 ppm.
5. Polybrominated biphenyls (PBB): less than 1000 ppm.
6. Polybrominated diphenyl ethers (PBDEs): less than 1000 ppm.
7. Di(2-ethylhexyl) phthalate (DEHP): less than 1000 ppm.
8. Butyl benzyl phthalate (BBP): less than 1000 ppm.
9. Dibutyl phthalate (DBP): less than 1000 ppm.
10. Diisobutyl phthalate (DIBP): less than 1000 ppm.
What instruments are typically used for RoHS testing?
The commonly used analytical instrument for RoHS 1.0 is X-ray fluorescence spectroscopy (XRF). XRF analyzers are available in standalone, benchtop, and handheld configurations. Typically, a handheld XRF analyzer is employed first for portable on-site testing, with a focus on components of the product that pose a higher risk of containing restricted substances.
With the introduction of RoHS 2.0, four phthalate esters have been added to the RoHS Directive, necessitating additional testing to determine the levels of these compounds, which are extracted using solvents. The solvent extracts are then analyzed for the presence of phthalates by gas chromatography–mass spectrometry (GC/MS) or gas chromatography with flame ionization detection (GC/FID). Other elements and polybrominated diphenyl ethers/alkylated elastomers are analyzed using comparative methods and GC/MS, respectively.
Do the ten RoHS items require two instruments for testing?
The ten items specified in RoHS require two instruments for testing.
For the ten‑item RoHS testing, two instruments are required to complete the analysis; a single instrument cannot achieve this. Even if one instrument is used to measure heavy metal content, the other must be employed to analyze organic compounds.
X-ray fluorescence spectrometers are used to detect RoHS 1.0, while chromatographs, pyrolyzers, or hyphenated techniques such as GC–MS are employed to analyze the four phthalates specified in RoHS 2.0. At present, no single instrument can independently perform all the tests required for RoHS 2.0.
Can ROHS testing instruments detect metallic elements?
ROHS testing instruments are specialized devices used to detect the presence of hazardous substances in electronic components, household appliances, toys, and other products. These harmful substances can pose long-term risks to both human health and the environment, making the use of ROHS testing equipment of paramount importance.
Regarding whether ROHS testing instruments can detect metallic elements, the answer is yes. This is because ROHS testing covers not only hazardous substances but also the analysis of metallic constituents in compliant products. These metals include lead, cadmium, mercury, and others, which are widely used during the manufacturing process. Although these elements are essential, their excessive use poses significant risks to human health and the environment. Consequently, the advent of ROHS testing instruments has greatly facilitated the detection of metallic elements.