Common issues encountered during ROHS testing and their solutions!

发布时间: 2022-08-19 10:51

RoHS detector As an instrument used for RoHS testing, we often encounter numerous issues during operation. Below, we will… RoHS detector manufacturer - DePu will explain the common issues encountered during ROHS testing and their corresponding solutions!

1. To what extent are the measurement results for cadmium, lead, and mercury affected? Under what circumstances might they be influenced, and where can such influences be observed? How should these findings be interpreted in practice?

Answer: In general, cadmium is subject to interference from the combined peaks of tin and bromine; lead is affected by the combined peaks of bismuth, arsenic, bromine, and iron; and mercury is interfered with by tungsten, as can be seen in the spectrum.

2. Is the measurement accurate when the product has an electroplated coating?

Answer: Inaccurate. Perform one test with the coating applied, then perform another test without the coating.

3. If a plastic product contains metal components that cannot be cut, how should RoHS compliance be assessed? Should the plastic analysis method or the metal analysis method be selected? And will the measurement results reflect the plastic’s compliance or the metal’s compliance?

Answer: These samples should be tested in two parts: the metal curve of the coating and the plastic curve after the coating has been removed.

4. Poor measurement repeatability—how can the variation be minimized?

Answer: The sample should be flat, have a certain thickness, and consist of a single substance.

5. When testing samples, if the material information is unknown in advance, how should one select the appropriate analysis group? For example, which analysis group should be chosen for testing solder paste, stainless steel, and similar materials? Can the instrument analyze liquid substances?

A: You can start with the EASY qualitative test, then select an appropriate testing method. Liquids can be measured, but the test must be conducted safely and without leakage.

6. Do metallic materials contain bromine (Br) and chlorine (Cl)? (Both Br and Cl are flame retardants.) If not, why do we select ferrous alloys when testing U‑iron, yet the test results still show the presence of Br?

Answer: The metal does not contain Br or Cl; this is due to interference from Br caused by the overlapping peaks of iron.

7. Why do the spectra of samples differ before testing with an RoHS analyzer?

Answer: ① The selected sample measurement type differs. ② The sample spectrum inadvertently overlaps with the sample inspection.

8. Why does the system prompt that the sample size is too small during testing?

A: This issue occurs in the analysis software; it does not appear in the RoHS software. It arises when “Control Count Rate” is selected under the “Device Parameters” option during measurement, for the following reasons:

(1) Error Curve

(2) The sample is not very good or too small.

(3) Error Message

9. Why is it necessary to measure the peak drift of lead, tin, and lead?

A: This is not a drift, but rather an apparent spectral artifact caused by the presence of nearby Sn–Pb–Bi-induced peaks.

10. Why is the count rate measured in accordance with the Restriction of Hazardous Substances (RoHS) directive consistently only on the order of tens to hundreds?

A: There may be debris obstructing the sensor or the collimator. Additionally, the instrument’s optical system could be malfunctioning, or there may be issues with the instrument’s settings.

11. After replacing the collimator on the RoHS tester, does the software crash, restart, and then resume normal operation?

A: This phenomenon is relatively rare and may affect the computer operating system on which the testing software crashes.

12. Why does data sometimes get lost after a RoHS test is completed?

Answer: (1) The save path is incorrect, (2) the system shut down unexpectedly, (3) special circumstances were used.

13. Why are the spectrum’s save path and the software installation path the same?

Answer: Software design requirements.

14. Why does the automatic count rate adjustment sometimes get selected and sometimes not?

Answer: Using test samples and measurement methods, the principle is that counting can be adjusted when the results are independent of the sample count rate.

15. Why is it necessary to preheat for 30 minutes?

A: Allow the instrument to warm up thoroughly to ensure the accuracy of test data. We recommend a 30-minute warm-up period, along with a rapid thermal cycling procedure.

16. Under the same conditions, is the deviation between two consecutive results large? In fact, which result do we choose?

Answer: Under normal circumstances, there are two main reasons for significant deviations under identical conditions: ① imbalance; ② comparison with data from the modified equipment. Additionally, the instrument should be free from external interference—for example, test results affected by power‑supply noise. Furthermore, certain hardware malfunctions may not actually constitute true hardware failures; such cases can be confirmed through RoHS testing.

The above is an overview of the issues encountered during ROHS detector testing and their corresponding solutions. If you have any other questions about ROHS detectors, please continue browsing. ROHS detector information If you have a need for an ROHS testing instrument, please leave us an online message or contact us by phone!

Latest News

Preferred Choice for RoHS Testing Instruments: Despu Instruments Leads the Industry Toward Compliance with Cutting-Edge Technology and Proven Expertise.

Today, as environmental regulations become increasingly stringent, a reliable RoHS testing device has become an essential tool for electronic and electrical manufacturers. With the release of GB 26572-2025, “Requirements for Restriction of Hazardous Substances in Electrical and Electronic Products,” in August 2025, China’s control over hazardous substances in electronic and electrical products has entered a new phase—“mandatory limits on 10 substances plus digital traceability.” Facing the equipment upgrade window before the official implementation in 2027, companies’ demand for RoHS testers that combine precision, compliance, and adaptability to various application scenarios has surged. Among numerous suppliers, Shenzhen Depu Instrument Co., Ltd. has emerged as one of the industry’s most trusted choices, thanks to its technological expertise and product quality. 01 Industry Background: New Challenges in RoHS Testing The global electronics and electrical industries are confronting ever‑tighter environmental regulations. From the EU’s RoHS Directive to China’s newly revised GB 26572 standard, controls on heavy metals such as lead, cadmium, mercury, and hexavalent chromium, as well as phthalates, have grown increasingly stringent. Companies require fast, accurate, and convenient testing methods to ensure product compliance. X-ray fluorescence spectroscopy, with its non‑destructive, rapid, and highly precise characteristics, has become one of the key technologies for RoHS screening. 02 Depu Instrument: Company Strengths and Background Shenzhen Depu Instrument Co., Ltd. is a high‑tech enterprise located in Shenzhen’s High‑Tech Industrial Park. Centered on X‑ray fluorescence analysis technology, it integrates R&D, manufacturing, and sales under one roof. The company operates two wholly owned subsidiaries and holds controlling stakes in several specialized firms, forming a comprehensive production and service system. As a professional manufacturer of analytical instruments, Depu Instrument primarily produces RoHS testers, X‑ray fluorescence spectrometers, coating thickness gauges, and direct‑reading optical emission spectrometers. These products are widely used across industries including non‑ferrous metals, geology and mining, electronics, electrical appliances, hardware, PCBs, semiconductors, telecommunications, electroplating, jewelry, and watchmaking. 03 Core Products: Features and Advantages of DX‑320L and DX‑320L PRO DX‑320L RoHS Tester The DX‑320L is a meticulously engineered X‑ray fluorescence spectrometer by Depu Instrument, integrating detection capabilities for RoHS, halogen, and eight major heavy metal directives into a single unit. **Technical Highlights:** - Equipped with an electrically cooled SDD semiconductor detector boasting a resolution of 127 eV ± 5, ensuring highly accurate results. - Covers elemental analysis from sodium (Na) to uranium (U), with concentration ranges spanning 1 ppm to 99.99%. - Offers rapid testing times of just 30–200 seconds, meeting the demands of high‑volume inspections. - Features motorized switching among multiple collimators and filters, enabling fully automated one‑button testing. **Application Benefits:** - Delivers excellent detection limits and accuracy for elements such as lead, mercury, cadmium, chromium, bromine, chlorine, arsenic, antimony, selenium, and barium. - Incorporates a full‑optical‑path radiation shielding system paired with a classic triple‑layer safety design, along with both software and hardware interlocks to ensure radiation safety. - Provides an open working curve function, allowing users to tailor harmful element detection and control strategies based on their specific material compositions. Innovative Breakthroughs of DX‑320L PRO Building upon the DX‑320L platform, Depu Instrument has further developed the advanced DX‑320L PRO model. Although specific specifications are not explicitly listed in search results, based on industry trends and Depu Instrument’s technical expertise, the PRO version typically features notable enhancements: - Further improved detection accuracy, with lower detection limits to meet even stricter compliance requirements. - Digital report generation capabilities, potentially supporting QR‑code‑based reporting and cloud‑based data storage, perfectly aligning with the new national standards for digital traceability. - Enhanced instrument stability and testing efficiency, optimized to handle continuous operation in industrial environments. 04 Market Positioning and Service Differentiators of Depu Instrument In the RoHS testing equipment market, Depu Instrument stands out thanks to its exceptional cost‑effectiveness and localized services. Compared with international brands like Thermo Fisher Scientific or Bruker, Depu Instrument maintains comparable core technical performance while keeping equipment acquisition costs more competitive. As a domestic company, Depu Instrument offers a fully Chinese‑language user interface and a locally based after‑sales team, significantly lowering the barrier to entry and reducing repair wait times for customers. Beyond providing testing equipment, Depu Instrument also delivers a “one‑stop” comprehensive EU environmental compliance solution, including the establishment of halogen‑free laboratories, RoHS labs, and EU REACH compliance laboratories. This holistic service model enables Depu Instrument to meet the diverse needs of clients across different scales. Depu Instrument’s RoHS testing devices seamlessly integrate practical technology with rigorous industrial design, offering Chinese manufacturers a reliable and cost‑effective choice in a fiercely competitive testing‑instrument market. As global environmental requirements continue to rise, the ability to conduct fast, accurate, and convenient testing has become a core competitive advantage for electronic and electrical manufacturers—and Depu Instrument is emerging as a vital technological partner in supporting this capability.

2025-10-14

Recommended Products

RoHS 2.0 Rapid Screening Instrument PY-GCMS3080

This page provides information, including frequently asked questions, about the RoHS 2.0 rapid screening instrument PY‑GCMS3080. For pricing inquiries, please contact us online or call our consultation hotline at 19926823034.

Learn more+

RoHS 2.0 Rapid Screening Instrument PY-GCMS3080

RoHS 1.0 Rapid Detection Instrument DX-320L Pro

This page provides an introduction to the DX-320L Pro, a high-performance desktop XRF analyzer compliant with RoHS 1.0. For more information, please contact 19926823034.

Learn more+

RoHS 1.0 Rapid Detection Instrument DX-320L Pro

RoHS 1.0 Detector – DP8300L Fluorescence Spectrometer

This page provides specifications and FAQs for the DP8300L fluorescence spectrometer. For pricing inquiries, please contact us online or call our consultation hotline at 19926823034!

Learn more+

RoHS 1.0 Detector – DP8300L Fluorescence Spectrometer

RoHS Six-Item Detector – DX-360 Fluorescence Spectrometer

This page provides specifications and FAQs for the DX-360 fluorescence spectrometer. For pricing inquiries, please contact us online or call our hotline at 19926823034.

Learn more+

RoHS Six-Item Detector – DX-360 Fluorescence Spectrometer

RoHS 2.0 Pyrolysis Gas Chromatograph PY-DP8060

This page provides specifications for the RoHS 2.0 pyrolysis gas chromatograph PY-DP8060, along with answers to frequently asked questions. For pricing inquiries, please contact us online or call our consultation hotline at 19926823034.

Learn more+

RoHS 2.0 Pyrolysis Gas Chromatograph PY-DP8060

Halogen detection instrument

This page provides specifications and FAQs for the IDP-6300 halogen detector. For pricing inquiries, please contact us online or call our hotline at 19926823034.

Learn more+

Halogen detection instrument

Frequently Asked Questions

What is RoHS?

RoHS is a mandatory standard established by EU legislation, officially known as the “Directive on the Restriction of the Use of Certain Hazardous Substances in Electrical and Electronic Equipment.” The standard entered into force on July 1, 2006, and primarily aims to regulate the materials and manufacturing processes of electrical and electronic products, thereby promoting human health and environmental protection.

The purpose of this standard is to eliminate six substances—lead, mercury, cadmium, hexavalent chromium, polybrominated biphenyls, and polybrominated diphenyl ethers (note: the correct Chinese term for PBDE is “polybrominated diphenyl ethers”; “polybrominated biphenyl ethers” is an incorrect designation)—and specifically stipulates that the cadmium content must not exceed 0.01%.

The latest RoHS Directive has added four new phthalate‑ester testing items: DEHP—di(2‑ethylhexyl) phthalate; BBP—butyl benzyl phthalate; DBP—dibutyl phthalate; and DIBP—diisobutyl phthalate.

What are the limit values for the ten hazardous substances under RoHS?

1. Cadmium: less than 100 ppm.

2. Lead: less than 1000 ppm.

3. Mercury: less than 1000 ppm.

4. Hexavalent chromium: less than 1000 ppm.

5. Polybrominated biphenyls (PBB): less than 1000 ppm.

6. Polybrominated diphenyl ethers (PBDEs): less than 1000 ppm.

7. Di(2-ethylhexyl) phthalate (DEHP): less than 1000 ppm.

8. Butyl benzyl phthalate (BBP): less than 1000 ppm.

9. Dibutyl phthalate (DBP): less than 1000 ppm.

10. Diisobutyl phthalate (DIBP): less than 1000 ppm.

What instruments are typically used for RoHS testing?

The commonly used analytical instrument for RoHS 1.0 is X-ray fluorescence spectroscopy (XRF). XRF analyzers are available in standalone, benchtop, and handheld configurations. Typically, a handheld XRF analyzer is employed first for portable on-site testing, with a focus on components of the product that pose a higher risk of containing restricted substances.

With the introduction of RoHS 2.0, four phthalate esters have been added to the RoHS Directive, necessitating additional testing to determine the levels of these compounds, which are extracted using solvents. The solvent extracts are then analyzed for the presence of phthalates by gas chromatography–mass spectrometry (GC/MS) or gas chromatography with flame ionization detection (GC/FID). Other elements and polybrominated diphenyl ethers/alkylated elastomers are analyzed using comparative methods and GC/MS, respectively.

Do the ten RoHS items require two instruments for testing?

The ten items specified in RoHS require two instruments for testing.
For the ten‑item RoHS testing, two instruments are required to complete the analysis; a single instrument cannot achieve this. Even if one instrument is used to measure heavy metal content, the other must be employed to analyze organic compounds.
X-ray fluorescence spectrometers are used to detect RoHS 1.0, while chromatographs, pyrolyzers, or hyphenated techniques such as GC–MS are employed to analyze the four phthalates specified in RoHS 2.0. At present, no single instrument can independently perform all the tests required for RoHS 2.0.

Can ROHS testing instruments detect metallic elements?

ROHS testing instruments are specialized devices used to detect the presence of hazardous substances in electronic components, household appliances, toys, and other products. These harmful substances can pose long-term risks to both human health and the environment, making the use of ROHS testing equipment of paramount importance.

Regarding whether ROHS testing instruments can detect metallic elements, the answer is yes. This is because ROHS testing covers not only hazardous substances but also the analysis of metallic constituents in compliant products. These metals include lead, cadmium, mercury, and others, which are widely used during the manufacturing process. Although these elements are essential, their excessive use poses significant risks to human health and the environment. Consequently, the advent of ROHS testing instruments has greatly facilitated the detection of metallic elements.