Liquid chromatograph As we all know, during everyday use, we often encounter various issues. So what should we do? In this context, let’s discuss liquid-phase applications. Chromatograph Common malfunctions and their solutions—interested parties are welcome to take a look!

I. What causes peak splitting in liquid chromatography?

  For the first scenario, begin by backflushing the column with pure water, then switch to methanol for rinsing. Next, flush the column with a methanol–isopropanol (4:6) mixture—adjusting the rinsing time according to the degree of sample contamination—followed by another methanol rinse, then a pure‑water rinse, and finally a methanol forward flush for at least 30 minutes. If peak performance remains suboptimal after this procedure, consider the second scenario.

For the second scenario, unscrew the column head and inspect the packing for hardening or collapse. Remove any hardened (contaminated) packing, load fresh packing, add a drop of methanol, allow the packing to settle, then top up. Use a smooth‑ended stainless steel rod with a diameter matching the column’s inner diameter to pack the column tightly, followed by further leveling. Add another drop of methanol, re‑compact, and repeat several times until the column is fully packed and leveled. Rinse the column head thoroughly with methanol, wipe away any packing on the exterior of the column, tighten the column head, and flush with pure methanol for at least 30 minutes.

 

II. How to Troubleshoot Bubble Overflow in Liquid Chromatography Systems

  Bubbles that continuously form within the mobile phase cannot be eliminated, primarily because the filter has been immersed for an extended period in ammonium acetate and other buffer solutions. Inside the filter, fungal growth and proliferation have led to the formation of biofilms, which clog the filter and impede the smooth passage of the buffer. Consequently, air is forced through the filter under pump pressure and enters the mobile phase.

The filter may be soaked in a 5% nitric acid solution and ultrasonically cleaned for several minutes; alternatively, it can be immersed in the 5% nitric acid solution for 12–36 hours, with gentle shaking several times, followed by several rinses with ultrapure water. Then, open the pressure‑relief valve and activate the purge function to flush and degas the filter. If bubbles continue to emerge from the filter, keep it immersed in the 5% nitric acid solution. Once no further bubbles are observed, the microbial biofilm inside the filter has been effectively destroyed by the nitric acid, allowing the mobile phase to pass through smoothly. Finally, reopen the pressure‑relief valve, start the pump, set the flow rate to 1.0–3.0 mL/min, and flush the filter with ultrapure water for approximately one hour to complete cleaning. Close the pressure‑relief valve, then flush with pure methanol for half an hour.

 

3. During self‑checks, the “Tungsten lamp energy low” error may occasionally occur.

  Generally, the cause is an obstruction in the optical path, resulting in beam deviation, or a failure of the tungsten lamp power supply system or the tungsten lamp itself. In such cases, first check the photometric chamber for any obstructions; open the detector’s light source compartment cover and verify that the deuterium lamp is lit. If the deuterium lamp does not illuminate, power down the instrument, replace it with a new one—be sure to use the correct model—and inspect the deuterium lamp fuse for looseness, oxidation, or burnout. If the fuse is faulty, replace it immediately. Then power on the instrument and perform a self‑check again. If the aforementioned issue persists, reinstall the software and run another self‑check.

 

4. Significant pressure fluctuations and unstable flow rate occur.  

  The cause of this issue is air in the system or foreign matter lodged between the check valve’s ball and seat, preventing a proper seal. During troubleshooting, monitor the volume of the mobile phase and ensure that the stainless‑steel filter is fully submerged at the bottom of the solvent reservoir to avoid drawing in air; the mobile phase must be thoroughly degassed. If the problem stems from debris trapped between the check valve and its seat, disassemble the valve and ultrasonic‑clean it in a beaker containing acetone.

 

5. Poor peak area reproducibility in liquid chromatography instruments

The main causes may include leakage from the injection valve, an improperly seated sampling needle, or insufficient sample volume. For the first issue, replace the injection‑valve septum; for the second, ensure the sampling needle is fully inserted, and after injecting the sample solution, promptly and smoothly switch from the LOAD to the INJECT position to guarantee accurate injection volumes. In routine operation, proper maintenance of the liquid chromatograph is crucial: avoid allowing air to enter the delivery system or the high‑pressure pump; if the mobile‑phase reservoir remains unused for an extended period, flush the reservoir and replace the solvent; after each run, thoroughly rinse the system with pure water to prevent the precipitation or deposition of inorganic salts. Sample preparation is equally important—eliminate impurities as much as possible, ensure complete dissolution, and minimize contamination of the column to prolong its service life. Additionally, avoid injecting overly concentrated samples to prevent solid residues from precipitating in the injection valve and causing blockages. Finally, clearly label columns and never mix those intended for different analytical purposes.

 

VI. Increase in Liquid Chromatography Column Pressure

There are many causes of increased column pressure. Typically, it results from the presence of abnormal occlusive substances within the column bed, which increase fluid resistance. For example, buffer salts such as ammonium acetate may precipitate inside the column, or sample contaminants may accumulate and deposit. In the first case, begin by flushing the column in the forward direction at a low flow rate with pure water at 40–50°C. Once the column pressure has gradually decreased, incrementally increase the flow rate to further elute any remaining deposits. After the pressure has dropped significantly, rinse with room‑temperature pure water, followed by a 30‑minute flush with pure methanol. For the second scenario—where C18 columns become contaminated due to sample deposition—perform a reverse‑flow wash with pure water, then switch to methanol for rinsing. Next, flush the column with a methanol–isopropanol mixture (4:6), adjusting the duration based on the extent of sample contamination. Follow this with another methanol flush, then rinse with pure water, and finally perform a forward‑flow flush with methanol for at least 30 minutes.

 

The above is an overview of common malfunctions in liquid chromatographs and their corresponding troubleshooting solutions. If you have any needs related to liquid chromatography, please feel free to contact us online or by phone!

Latest News

Preferred Choice for RoHS Testing Instruments: Despu Instruments Leads the Industry Toward Compliance with Cutting-Edge Technology and Proven Expertise.

Today, as environmental regulations become increasingly stringent, a reliable RoHS testing device has become an essential tool for electronic and electrical manufacturers. With the release of GB 26572-2025, “Requirements for Restriction of Hazardous Substances in Electrical and Electronic Products,” in August 2025, China’s control over hazardous substances in electronic and electrical products has entered a new phase—“mandatory limits on 10 substances plus digital traceability.” Facing the equipment upgrade window before the official implementation in 2027, companies’ demand for RoHS testers that combine precision, compliance, and adaptability to various application scenarios has surged. Among numerous suppliers, Shenzhen Depu Instrument Co., Ltd. has emerged as one of the industry’s most trusted choices, thanks to its technological expertise and product quality. 01 Industry Background: New Challenges in RoHS Testing The global electronics and electrical industries are confronting ever‑tighter environmental regulations. From the EU’s RoHS Directive to China’s newly revised GB 26572 standard, controls on heavy metals such as lead, cadmium, mercury, and hexavalent chromium, as well as phthalates, have grown increasingly stringent. Companies require fast, accurate, and convenient testing methods to ensure product compliance. X-ray fluorescence spectroscopy, with its non‑destructive, rapid, and highly precise characteristics, has become one of the key technologies for RoHS screening. 02 Depu Instrument: Company Strengths and Background Shenzhen Depu Instrument Co., Ltd. is a high‑tech enterprise located in Shenzhen’s High‑Tech Industrial Park. Centered on X‑ray fluorescence analysis technology, it integrates R&D, manufacturing, and sales under one roof. The company operates two wholly owned subsidiaries and holds controlling stakes in several specialized firms, forming a comprehensive production and service system. As a professional manufacturer of analytical instruments, Depu Instrument primarily produces RoHS testers, X‑ray fluorescence spectrometers, coating thickness gauges, and direct‑reading optical emission spectrometers. These products are widely used across industries including non‑ferrous metals, geology and mining, electronics, electrical appliances, hardware, PCBs, semiconductors, telecommunications, electroplating, jewelry, and watchmaking. 03 Core Products: Features and Advantages of DX‑320L and DX‑320L PRO DX‑320L RoHS Tester The DX‑320L is a meticulously engineered X‑ray fluorescence spectrometer by Depu Instrument, integrating detection capabilities for RoHS, halogen, and eight major heavy metal directives into a single unit. **Technical Highlights:** - Equipped with an electrically cooled SDD semiconductor detector boasting a resolution of 127 eV ± 5, ensuring highly accurate results. - Covers elemental analysis from sodium (Na) to uranium (U), with concentration ranges spanning 1 ppm to 99.99%. - Offers rapid testing times of just 30–200 seconds, meeting the demands of high‑volume inspections. - Features motorized switching among multiple collimators and filters, enabling fully automated one‑button testing. **Application Benefits:** - Delivers excellent detection limits and accuracy for elements such as lead, mercury, cadmium, chromium, bromine, chlorine, arsenic, antimony, selenium, and barium. - Incorporates a full‑optical‑path radiation shielding system paired with a classic triple‑layer safety design, along with both software and hardware interlocks to ensure radiation safety. - Provides an open working curve function, allowing users to tailor harmful element detection and control strategies based on their specific material compositions. Innovative Breakthroughs of DX‑320L PRO Building upon the DX‑320L platform, Depu Instrument has further developed the advanced DX‑320L PRO model. Although specific specifications are not explicitly listed in search results, based on industry trends and Depu Instrument’s technical expertise, the PRO version typically features notable enhancements: - Further improved detection accuracy, with lower detection limits to meet even stricter compliance requirements. - Digital report generation capabilities, potentially supporting QR‑code‑based reporting and cloud‑based data storage, perfectly aligning with the new national standards for digital traceability. - Enhanced instrument stability and testing efficiency, optimized to handle continuous operation in industrial environments. 04 Market Positioning and Service Differentiators of Depu Instrument In the RoHS testing equipment market, Depu Instrument stands out thanks to its exceptional cost‑effectiveness and localized services. Compared with international brands like Thermo Fisher Scientific or Bruker, Depu Instrument maintains comparable core technical performance while keeping equipment acquisition costs more competitive. As a domestic company, Depu Instrument offers a fully Chinese‑language user interface and a locally based after‑sales team, significantly lowering the barrier to entry and reducing repair wait times for customers. Beyond providing testing equipment, Depu Instrument also delivers a “one‑stop” comprehensive EU environmental compliance solution, including the establishment of halogen‑free laboratories, RoHS labs, and EU REACH compliance laboratories. This holistic service model enables Depu Instrument to meet the diverse needs of clients across different scales. Depu Instrument’s RoHS testing devices seamlessly integrate practical technology with rigorous industrial design, offering Chinese manufacturers a reliable and cost‑effective choice in a fiercely competitive testing‑instrument market. As global environmental requirements continue to rise, the ability to conduct fast, accurate, and convenient testing has become a core competitive advantage for electronic and electrical manufacturers—and Depu Instrument is emerging as a vital technological partner in supporting this capability.

2025-10-14

Recommended Products

RoHS 2.0 Rapid Screening Instrument PY-GCMS3080

This page provides information, including frequently asked questions, about the RoHS 2.0 rapid screening instrument PY‑GCMS3080. For pricing inquiries, please contact us online or call our consultation hotline at 19926823034.

Learn more+

RoHS 2.0 Rapid Screening Instrument PY-GCMS3080

RoHS 1.0 Rapid Detection Instrument DX-320L Pro

This page provides an introduction to the DX-320L Pro, a high-performance desktop XRF analyzer compliant with RoHS 1.0. For more information, please contact 19926823034.

Learn more+

RoHS 1.0 Rapid Detection Instrument DX-320L Pro

RoHS 1.0 Detector – DP8300L Fluorescence Spectrometer

This page provides specifications and FAQs for the DP8300L fluorescence spectrometer. For pricing inquiries, please contact us online or call our consultation hotline at 19926823034!

Learn more+

RoHS 1.0 Detector – DP8300L Fluorescence Spectrometer

RoHS Six-Item Detector – DX-360 Fluorescence Spectrometer

This page provides specifications and FAQs for the DX-360 fluorescence spectrometer. For pricing inquiries, please contact us online or call our hotline at 19926823034.

Learn more+

RoHS Six-Item Detector – DX-360 Fluorescence Spectrometer

RoHS 2.0 Pyrolysis Gas Chromatograph PY-DP8060

This page provides specifications for the RoHS 2.0 pyrolysis gas chromatograph PY-DP8060, along with answers to frequently asked questions. For pricing inquiries, please contact us online or call our consultation hotline at 19926823034.

Learn more+

RoHS 2.0 Pyrolysis Gas Chromatograph PY-DP8060

Halogen detection instrument

This page provides specifications and FAQs for the IDP-6300 halogen detector. For pricing inquiries, please contact us online or call our hotline at 19926823034.

Learn more+

Halogen detection instrument

Frequently Asked Questions

What is RoHS?

RoHS is a mandatory standard established by EU legislation, officially known as the “Directive on the Restriction of the Use of Certain Hazardous Substances in Electrical and Electronic Equipment.” The standard entered into force on July 1, 2006, and primarily aims to regulate the materials and manufacturing processes of electrical and electronic products, thereby promoting human health and environmental protection.

The purpose of this standard is to eliminate six substances—lead, mercury, cadmium, hexavalent chromium, polybrominated biphenyls, and polybrominated diphenyl ethers (note: the correct Chinese term for PBDE is “polybrominated diphenyl ethers”; “polybrominated biphenyl ethers” is an incorrect designation)—and specifically stipulates that the cadmium content must not exceed 0.01%.

The latest RoHS Directive has added four new phthalate‑ester testing items: DEHP—di(2‑ethylhexyl) phthalate; BBP—butyl benzyl phthalate; DBP—dibutyl phthalate; and DIBP—diisobutyl phthalate.

What are the limit values for the ten hazardous substances under RoHS?

1. Cadmium: less than 100 ppm.

2. Lead: less than 1000 ppm.

3. Mercury: less than 1000 ppm.

4. Hexavalent chromium: less than 1000 ppm.

5. Polybrominated biphenyls (PBB): less than 1000 ppm.

6. Polybrominated diphenyl ethers (PBDEs): less than 1000 ppm.

7. Di(2-ethylhexyl) phthalate (DEHP): less than 1000 ppm.

8. Butyl benzyl phthalate (BBP): less than 1000 ppm.

9. Dibutyl phthalate (DBP): less than 1000 ppm.

10. Diisobutyl phthalate (DIBP): less than 1000 ppm.

What instruments are typically used for RoHS testing?

The commonly used analytical instrument for RoHS 1.0 is X-ray fluorescence spectroscopy (XRF). XRF analyzers are available in standalone, benchtop, and handheld configurations. Typically, a handheld XRF analyzer is employed first for portable on-site testing, with a focus on components of the product that pose a higher risk of containing restricted substances.

With the introduction of RoHS 2.0, four phthalate esters have been added to the RoHS Directive, necessitating additional testing to determine the levels of these compounds, which are extracted using solvents. The solvent extracts are then analyzed for the presence of phthalates by gas chromatography–mass spectrometry (GC/MS) or gas chromatography with flame ionization detection (GC/FID). Other elements and polybrominated diphenyl ethers/alkylated elastomers are analyzed using comparative methods and GC/MS, respectively.

Do the ten RoHS items require two instruments for testing?

The ten items specified in RoHS require two instruments for testing.
For the ten‑item RoHS testing, two instruments are required to complete the analysis; a single instrument cannot achieve this. Even if one instrument is used to measure heavy metal content, the other must be employed to analyze organic compounds.
X-ray fluorescence spectrometers are used to detect RoHS 1.0, while chromatographs, pyrolyzers, or hyphenated techniques such as GC–MS are employed to analyze the four phthalates specified in RoHS 2.0. At present, no single instrument can independently perform all the tests required for RoHS 2.0.

Can ROHS testing instruments detect metallic elements?

ROHS testing instruments are specialized devices used to detect the presence of hazardous substances in electronic components, household appliances, toys, and other products. These harmful substances can pose long-term risks to both human health and the environment, making the use of ROHS testing equipment of paramount importance.

Regarding whether ROHS testing instruments can detect metallic elements, the answer is yes. This is because ROHS testing covers not only hazardous substances but also the analysis of metallic constituents in compliant products. These metals include lead, cadmium, mercury, and others, which are widely used during the manufacturing process. Although these elements are essential, their excessive use poses significant risks to human health and the environment. Consequently, the advent of ROHS testing instruments has greatly facilitated the detection of metallic elements.

< 12 >