Factors Affecting the Measurement Accuracy of Coating Thickness Gauges!
发布时间: 2022-07-08 11:00
All metals are prone to oxidation and corrosion when exposed to air over time, which is why they undergo one or more surface treatments. These treatments serve two purposes: first, to better protect the metal’s service life; and second, to enhance the aesthetic appeal of the object. However, coating thickness measurements after metal treatment are typically applied in… Coating thickness gauge Above, it affects the coating. Thickness gauge What are the causes of measurement accuracy?
The main factors affecting the measurement accuracy of coating thickness gauges are: Base material magnetism, substrate thickness, edge effects, curvature, surface roughness, external magnetic field, adhering substances, probe pressure, probe position, specimen deformation, and so on.
Factors Affecting Measurement Accuracy and Relevant Explanations:
1. Thickness measurements using the magnetic‑method on base materials are affected by changes in the material’s magnetic properties (in practical applications, variations in the magnetic properties of low‑carbon steel can be considered negligible). To minimize the influence of heat treatment and cold working, the instrument should be calibrated against standard metal sheets with identical characteristics. Calibration may also be performed directly on the test specimens to be coated.
2. Metal Thickness: Each instrument has a critical thickness of the base metal. Beyond this thickness, the measurement is no longer affected by the thickness of the base metal.
3. Edge effect: The instrument is highly sensitive to abrupt changes in the sample’s surface geometry. Consequently, measurements taken near the sample edges or at internal corners are unreliable.
4. The curvature of the specimen affects the measurement. This effect invariably becomes more pronounced as the radius of curvature decreases.
5. Surface roughness: The surface roughness of both the base material and the coating layer affects measurement results. As roughness increases, the influence also increases. Rough surfaces can introduce systematic and random errors; therefore, each measurement should be repeated at multiple locations to mitigate such random errors. If the base material is rough, the instrument’s zero point must be calibrated at several positions on uncoated base‑material specimens with comparable roughness.
6. Strong magnetic fields generated by various electrical equipment in the vicinity can severely interfere with magnetic thickness measurements.
7. Adhesive substances are sensitive to those that prevent the probe from making close contact with the surface of the coating layer. Therefore, adhering materials must be removed to ensure direct contact between the instrument’s probe and the specimen’s surface.
Latest News
Recommended Products
Frequently Asked Questions
What is RoHS?
RoHS is a mandatory standard established by EU legislation, officially known as the “Directive on the Restriction of the Use of Certain Hazardous Substances in Electrical and Electronic Equipment.” The standard entered into force on July 1, 2006, and primarily aims to regulate the materials and manufacturing processes of electrical and electronic products, thereby promoting human health and environmental protection.
The purpose of this standard is to eliminate six substances—lead, mercury, cadmium, hexavalent chromium, polybrominated biphenyls, and polybrominated diphenyl ethers (note: the correct Chinese term for PBDE is “polybrominated diphenyl ethers”; “polybrominated biphenyl ethers” is an incorrect designation)—and specifically stipulates that the cadmium content must not exceed 0.01%.
The latest RoHS Directive has added four new phthalate‑ester testing items: DEHP—di(2‑ethylhexyl) phthalate; BBP—butyl benzyl phthalate; DBP—dibutyl phthalate; and DIBP—diisobutyl phthalate.
What are the limit values for the ten hazardous substances under RoHS?
1. Cadmium: less than 100 ppm.
2. Lead: less than 1000 ppm.
3. Mercury: less than 1000 ppm.
4. Hexavalent chromium: less than 1000 ppm.
5. Polybrominated biphenyls (PBB): less than 1000 ppm.
6. Polybrominated diphenyl ethers (PBDEs): less than 1000 ppm.
7. Di(2-ethylhexyl) phthalate (DEHP): less than 1000 ppm.
8. Butyl benzyl phthalate (BBP): less than 1000 ppm.
9. Dibutyl phthalate (DBP): less than 1000 ppm.
10. Diisobutyl phthalate (DIBP): less than 1000 ppm.
What instruments are typically used for RoHS testing?
The commonly used analytical instrument for RoHS 1.0 is X-ray fluorescence spectroscopy (XRF). XRF analyzers are available in standalone, benchtop, and handheld configurations. Typically, a handheld XRF analyzer is employed first for portable on-site testing, with a focus on components of the product that pose a higher risk of containing restricted substances.
With the introduction of RoHS 2.0, four phthalate esters have been added to the RoHS Directive, necessitating additional testing to determine the levels of these compounds, which are extracted using solvents. The solvent extracts are then analyzed for the presence of phthalates by gas chromatography–mass spectrometry (GC/MS) or gas chromatography with flame ionization detection (GC/FID). Other elements and polybrominated diphenyl ethers/alkylated elastomers are analyzed using comparative methods and GC/MS, respectively.
Do the ten RoHS items require two instruments for testing?
The ten items specified in RoHS require two instruments for testing.
For the ten‑item RoHS testing, two instruments are required to complete the analysis; a single instrument cannot achieve this. Even if one instrument is used to measure heavy metal content, the other must be employed to analyze organic compounds.
X-ray fluorescence spectrometers are used to detect RoHS 1.0, while chromatographs, pyrolyzers, or hyphenated techniques such as GC–MS are employed to analyze the four phthalates specified in RoHS 2.0. At present, no single instrument can independently perform all the tests required for RoHS 2.0.
Can ROHS testing instruments detect metallic elements?
ROHS testing instruments are specialized devices used to detect the presence of hazardous substances in electronic components, household appliances, toys, and other products. These harmful substances can pose long-term risks to both human health and the environment, making the use of ROHS testing equipment of paramount importance.
Regarding whether ROHS testing instruments can detect metallic elements, the answer is yes. This is because ROHS testing covers not only hazardous substances but also the analysis of metallic constituents in compliant products. These metals include lead, cadmium, mercury, and others, which are widely used during the manufacturing process. Although these elements are essential, their excessive use poses significant risks to human health and the environment. Consequently, the advent of ROHS testing instruments has greatly facilitated the detection of metallic elements.