News Center

News Center

Verification Procedure for Liquid Chromatographs!

发布时间: 2022-07-06 09:59

Liquid chromatograph It is our laboratory instrument, yet many people are unaware of the calibration procedures and how to operate it. For our lab personnel, this is unacceptable. So, what are the calibration protocols for liquid chromatographs? Let’s take a look together.

1. Filter the mobile phase, selecting an appropriate membrane filter as needed.

2. After vacuum filtration for 10–20 minutes, degas the mobile phase using ultrasonic treatment.

3. Turn on the HPLC workstation (including the computer software and the chromatograph), connect the mobile-phase tubing, and link the detection system.

4. Enter the HPLC control interface’s main menu, then click “Manual” to access the manual mode menu.

5. If the system has been idle for a period of time, or if a new mobile phase has been used, first flush the pump and the injection valve. Flush the pump directly at its outlet using a syringe, then draw the solvent through the system with a needle. To flush the injection valve, in the manual mode, first select “Purge” and then press “Start,” maintaining a flushing flow rate of no more than 10 mL/min.

6. Adjust the flow rate. When using a new mobile phase for the first time, you can start by checking the pressure. Higher flow rates correspond to higher pressures; typically, keep it below 2000 psi. Click “Harm,” select an appropriate flow rate, click again, and then acquire the baseline while monitoring it.

7. Design a sampling method. Click File, then select “Select Users and Methods” to choose from various pre-defined methods. To create a new method, click “New Method.” Select the required accessories as needed, including the injection valve, pump, detector, and others. After making your selections, click “Protocol.” A complete sampling method should include: A) flow stabilization prior to injection, typically 2–5 minutes; B) baseline zeroing; C) loading‑to‑injection switching of the injection valve; and D) the sampling time, which varies depending on the sample type.

8. Sample injection and post‑injection procedures. Select the alias method and click Start. Inject the samples, filtering all of them. Once the method is complete, click “Postrun” to record the data and mark it. After running all samples, use the aforementioned method to run the baseline and flush out any remaining sample residues.

9. Complete the registration log and have it signed by the person in charge.

10. The mobile phase should be chromatography-grade, and the water should be 20 mΩ·cm deionized water. Carefully shake the degassed mobile phase to minimize bubble formation.

11. The column is very fragile. The first method is to prevent the liquid from passing through the column.

12. All liquids passing through the column must be rigorously filtered.

13. Pressure should not be too high; keep it below 2,000 psi.

The above is an introduction to the verification procedures for liquid chromatographs. If you would like to learn more about… Liquid Chromatography Instrument Information Yes, you can continue browsing other pages, or if you’d like to learn more… Liquid chromatography instrument products You can consult us online or contact us by phone!

Latest News

Recommended Products

RoHS 2.0 Rapid Screening Instrument PY-GCMS3080

This page provides information, including frequently asked questions, about the RoHS 2.0 rapid screening instrument PY‑GCMS3080. For pricing inquiries, please contact us online or call our hotline at 19926823034!

Learn more+

RoHS 2.0 Rapid Screening Instrument PY-GCMS3080

RoHS 1.0 Rapid Testing Instrument DX-320L Pro

This page provides an introduction to the DX-320L Pro, a high-performance desktop XRF analyzer compliant with RoHS 1.0. For more information, please contact 19926823034.

Learn more+

RoHS 1.0 Rapid Testing Instrument DX-320L Pro

RoHS 1.0 Detector – IDP8300L Fluorescence Spectrometer

This page provides specifications and FAQs for the DP8300L fluorescence spectrometer. For pricing inquiries, please contact us online or call 19926823034!

Learn more+

RoHS 1.0 Detector – IDP8300L Fluorescence Spectrometer

RoHS Six-Item Detector – DX-360 Fluorescence Spectrometer

This page provides specifications and FAQs for the DX-360 fluorescence spectrometer. For pricing inquiries, please contact us online or call 19926823034!

Learn more+

RoHS Six-Item Detector – DX-360 Fluorescence Spectrometer

RoHS 2.0 Pyrolysis Gas Chromatograph PY-DP8060

This page provides specifications for the RoHS 2.0 pyrolysis gas chromatograph PY-DP8060, along with FAQs and other information. For a quote, please contact us online or call our hotline at 19926823034!

Learn more+

RoHS 2.0 Pyrolysis Gas Chromatograph PY-DP8060

Halogen detection instrument

This page provides specifications and FAQs for the IDP-6300 halogen detector. For pricing inquiries, please contact us online or call our hotline at 19926823034.

Learn more+

Halogen detection instrument

Frequently Asked Questions

What is RoHS?

RoHS is a mandatory standard established by EU legislation, officially known as the “Directive on the Restriction of the Use of Certain Hazardous Substances in Electrical and Electronic Equipment.” The standard entered into force on July 1, 2006, and primarily aims to regulate the materials and manufacturing processes of electrical and electronic products, thereby promoting human health and environmental protection.

The purpose of this standard is to eliminate six substances—lead, mercury, cadmium, hexavalent chromium, polybrominated biphenyls, and polybrominated diphenyl ethers (note: the correct Chinese term for PBDE is “polybrominated diphenyl ethers”; “polybrominated biphenyl ethers” is an incorrect designation)—and specifically stipulates that the cadmium content must not exceed 0.01%.

The latest RoHS Directive has added four new phthalate‑ester testing items: DEHP—di(2‑ethylhexyl) phthalate; BBP—butyl benzyl phthalate; DBP—dibutyl phthalate; and DIBP—diisobutyl phthalate.

What are the limit values for the ten hazardous substances under RoHS?

1. Cadmium: less than 100 ppm.

2. Lead: less than 1000 ppm.

3. Mercury: less than 1000 ppm.

4. Hexavalent chromium: less than 1000 ppm.

5. Polybrominated biphenyls (PBB): less than 1000 ppm.

6. Polybrominated diphenyl ethers (PBDEs): less than 1000 ppm.

7. Di(2-ethylhexyl) phthalate (DEHP): less than 1000 ppm.

8. Butyl benzyl phthalate (BBP): less than 1000 ppm.

9. Dibutyl phthalate (DBP): less than 1000 ppm.

10. Diisobutyl phthalate (DIBP): less than 1000 ppm.

What instruments are typically used for RoHS testing?

The commonly used analytical instrument for RoHS 1.0 is X-ray fluorescence spectroscopy (XRF). XRF analyzers are available in standalone, benchtop, and handheld configurations. Typically, a handheld XRF analyzer is employed first for portable on-site testing, with a focus on components of the product that pose a higher risk of containing restricted substances.

With the introduction of RoHS 2.0, four phthalate esters have been added to the RoHS Directive, necessitating additional testing to determine the levels of these compounds, which are extracted using solvents. The solvent extracts are then analyzed for the presence of phthalates by gas chromatography–mass spectrometry (GC/MS) or gas chromatography with flame ionization detection (GC/FID). Other elements and polybrominated diphenyl ethers/alkylated elastomers are analyzed using comparative methods and GC/MS, respectively.

Do the ten RoHS items require two instruments for testing?

The ten items specified in RoHS require two instruments for testing.
For the ten‑item RoHS testing, two instruments are required to complete the analysis; a single instrument cannot achieve this. Even if one instrument is used to measure heavy metal content, the other must be employed to analyze organic compounds.
X-ray fluorescence spectrometers are used to detect RoHS 1.0, while chromatographs, pyrolyzers, or hyphenated techniques such as GC–MS are employed to analyze the four phthalates specified in RoHS 2.0. At present, no single instrument can independently perform all the tests required for RoHS 2.0.

Can ROHS testing instruments detect metallic elements?

ROHS testing instruments are specialized devices used to detect the presence of hazardous substances in electronic components, household appliances, toys, and other products. These harmful substances can pose long-term risks to both human health and the environment, making the use of ROHS testing equipment of paramount importance.

Regarding whether ROHS testing instruments can detect metallic elements, the answer is yes. This is because ROHS testing covers not only hazardous substances but also the analysis of metallic constituents in compliant products. These metals include lead, cadmium, mercury, and others, which are widely used during the manufacturing process. Although these elements are essential, their excessive use poses significant risks to human health and the environment. Consequently, the advent of ROHS testing instruments has greatly facilitated the detection of metallic elements.

< 12 >