Operating Procedure for a Certain Coating Thickness Gauge!
发布时间: 2022-06-23 11:05
A coating thickness gauge measures the thickness of metallic films such as coatings by detecting the intensity of the second‑order X‑rays reflected from the sample after irradiation. Since it does not come into contact with the sample and delivers only about 45–75 W of X‑ray power, it causes no damage to the specimen. Moreover, measurements can be completed in as little as 10 seconds to a few minutes.
I. Operating Procedures for Coating Thickness Gauges!
1. Start
Turn on the power to the coating thickness gauge host, the computer, the monitor, and the printer.
2. Enter the program
After the computer boots up, the SmarkLink KP program will automatically display a password prompt; you must enter the correct password. If you exit the program, you’ll need to log in from scratch—simply double-click the SmarkLink FP icon on your desktop, and the same dialog box will appear.
3. Step-up transformation
Once the machine has completed initialization, the gray GO button at the bottom of the screen turns green, indicating that the boosting process is finished.
4. Heat Engine
If the device has been powered off for more than 3 hours, the following steps must be performed: Click “Count,” insert the Spect Cal into the instrument, align the Ag point with the center of the crosshair, focus the laser, set the measurement time to 6 seconds, and repeat the measurement 30–50 times. Then, click the green “Go” button and wait for the automatic continuous measurements to complete.
5. Spectral Calibration
If the instrument remains on for more than 3 hours without being shut down, or if it is shut down, it must be restarted to allow the device to perform self‑compensation and adjustment. First, click the “Spectrum Calibration” icon in the taskbar, then insert the spectrum calibration chart (“Spect Cal”) into the instrument and press the Go button to begin measurement. When the red STOP button changes to a green GO button, center the pure Ag spot on the crosshair, focus the laser, and click the Go button again to take the measurement. Once completed, a “Spectrum Calibration Successful” dialog box will appear; click OK. If the calibration fails, verify that no errors were made, that the process was not interrupted, and that the spectrum was not displaced. In such cases, the spectrum calibration must be repeated.
6. Measurement
On the measurement interface, click “Measure 123” at the top, then click the yellow folder on the right, select the appropriate file corresponding to the coating layer, open the file, and place the sample to be measured inside. Once the measurement is complete, you can view the results directly in the display box at the upper right corner.
7. Print
Click the “Print” option in the “Statistics” drop-down menu on the top taskbar, select either “All” or “Individual Data” for printing, and after exiting, click “Confirm.”
8. Shut down
First, remove the sample from the sample stage, exit the program, and then turn off the monitor, the host computer, the printer, and other devices.
Latest News
Recommended Products
Frequently Asked Questions
What is RoHS?
RoHS is a mandatory standard established by EU legislation, officially known as the “Directive on the Restriction of the Use of Certain Hazardous Substances in Electrical and Electronic Equipment.” The standard entered into force on July 1, 2006, and primarily aims to regulate the materials and manufacturing processes of electrical and electronic products, thereby promoting human health and environmental protection.
The purpose of this standard is to eliminate six substances—lead, mercury, cadmium, hexavalent chromium, polybrominated biphenyls, and polybrominated diphenyl ethers (note: the correct Chinese term for PBDE is “polybrominated diphenyl ethers”; “polybrominated biphenyl ethers” is an incorrect designation)—and specifically stipulates that the cadmium content must not exceed 0.01%.
The latest RoHS Directive has added four new phthalate‑ester testing items: DEHP—di(2‑ethylhexyl) phthalate; BBP—butyl benzyl phthalate; DBP—dibutyl phthalate; and DIBP—diisobutyl phthalate.
What are the limit values for the ten hazardous substances under RoHS?
1. Cadmium: less than 100 ppm.
2. Lead: less than 1000 ppm.
3. Mercury: less than 1000 ppm.
4. Hexavalent chromium: less than 1000 ppm.
5. Polybrominated biphenyls (PBB): less than 1000 ppm.
6. Polybrominated diphenyl ethers (PBDEs): less than 1000 ppm.
7. Di(2-ethylhexyl) phthalate (DEHP): less than 1000 ppm.
8. Butyl benzyl phthalate (BBP): less than 1000 ppm.
9. Dibutyl phthalate (DBP): less than 1000 ppm.
10. Diisobutyl phthalate (DIBP): less than 1000 ppm.
What instruments are typically used for RoHS testing?
The commonly used analytical instrument for RoHS 1.0 is X-ray fluorescence spectroscopy (XRF). XRF analyzers are available in standalone, benchtop, and handheld configurations. Typically, a handheld XRF analyzer is employed first for portable on-site testing, with a focus on components of the product that pose a higher risk of containing restricted substances.
With the introduction of RoHS 2.0, four phthalate esters have been added to the RoHS Directive, necessitating additional testing to determine the levels of these compounds, which are extracted using solvents. The solvent extracts are then analyzed for the presence of phthalates by gas chromatography–mass spectrometry (GC/MS) or gas chromatography with flame ionization detection (GC/FID). Other elements and polybrominated diphenyl ethers/alkylated elastomers are analyzed using comparative methods and GC/MS, respectively.
Do the ten RoHS items require two instruments for testing?
The ten items specified in RoHS require two instruments for testing.
For the ten‑item RoHS testing, two instruments are required to complete the analysis; a single instrument cannot achieve this. Even if one instrument is used to measure heavy metal content, the other must be employed to analyze organic compounds.
X-ray fluorescence spectrometers are used to detect RoHS 1.0, while chromatographs, pyrolyzers, or hyphenated techniques such as GC–MS are employed to analyze the four phthalates specified in RoHS 2.0. At present, no single instrument can independently perform all the tests required for RoHS 2.0.
Can ROHS testing instruments detect metallic elements?
ROHS testing instruments are specialized devices used to detect the presence of hazardous substances in electronic components, household appliances, toys, and other products. These harmful substances can pose long-term risks to both human health and the environment, making the use of ROHS testing equipment of paramount importance.
Regarding whether ROHS testing instruments can detect metallic elements, the answer is yes. This is because ROHS testing covers not only hazardous substances but also the analysis of metallic constituents in compliant products. These metals include lead, cadmium, mercury, and others, which are widely used during the manufacturing process. Although these elements are essential, their excessive use poses significant risks to human health and the environment. Consequently, the advent of ROHS testing instruments has greatly facilitated the detection of metallic elements.