Operation and Precautions for Spectrometers!
发布时间: 2022-06-17 13:49
As is well known, a spectrometer is an analytical instrument, yet many users overlook the essential precautions when operating it. Today, we’ll briefly outline the proper procedures and key safety considerations for using a spectrometer—interested readers are encouraged to take a look!
1. The purity of the supplied argon gas must be ≥99.999%.
2. When the pressure inside the argon cylinder falls below 2 MPa, the cylinder must be replaced.
3. When replacing the argon gas, first switch both the “Pure Gas” and “Original Gas” valves to the closed position.
4. After the purifier is powered on, the main power indicator light will illuminate in orange-red. Turn on the “Catalytic Heating” switch, and the “Catalytic Temperature Control” indicator panel will begin to heat up. Do not operate with either “feed gas” or “pure gas” until the temperature exceeds 240°C, to avoid damaging the purifier’s purification system.
5. When the “Catalytic Temperature Control” pointer reaches the “red dot,” it indicates that the catalytic system is functioning normally, and the “Feed Gas” and “Pure Gas” switches may be turned on. At this point, the feed gas is routed into the purifier for purification, and the purified gas is sequentially delivered to the main unit.
6. The purified argon gas must be introduced into the main unit before the unit can be powered on. The sequence is as follows: ① Turn the red wrench knob to “ON” to start the main unit; ② Press “SOURCE” to activate the light source (the button’s red indicator lights up); ③ Press “STANDBY” to turn on the controller (the button’s red indicator lights up). At this point, a faint humming sound will be audible inside the main unit.
7. The computer’s power supply is provided by the system unit. After the system unit is powered on, press the power button on the computer case; the indicator light will illuminate, and the computer will begin its self‑diagnostic startup sequence.
8. After the computer boots up normally and enters the Windows 2000 desktop, the “SP” icon on the desktop is a shortcut to the SPECTRO control software. Double-click it, press Enter, or right-click and select “Open” to launch the software.
9. Machines calibrated according to “standardization” can directly test ground samples. F2 is an excitation test; once the test conditions are met, it outputs the test results.
10. If excitation fails, first press the F3 command to stop the host, then investigate the cause of the failure.
11. After each excitation, thoroughly clean any oxide scale remaining on the excitation needle. Once cleaning is complete, resume testing or shut down the computer and main unit in the reverse order of startup.
The above is an overview of the spectrometer’s operation and key precautions. If you have any spectrometer requirements, please contact us online or reach out for assistance—we’ll promptly provide you with a quotation and other relevant product information.
Latest News
Recommended Products
Frequently Asked Questions
What is RoHS?
RoHS is a mandatory standard established by EU legislation, officially known as the “Directive on the Restriction of the Use of Certain Hazardous Substances in Electrical and Electronic Equipment.” The standard entered into force on July 1, 2006, and primarily aims to regulate the materials and manufacturing processes of electrical and electronic products, thereby promoting human health and environmental protection.
The purpose of this standard is to eliminate six substances—lead, mercury, cadmium, hexavalent chromium, polybrominated biphenyls, and polybrominated diphenyl ethers (note: the correct Chinese term for PBDE is “polybrominated diphenyl ethers”; “polybrominated biphenyl ethers” is an incorrect designation)—and specifically stipulates that the cadmium content must not exceed 0.01%.
The latest RoHS Directive has added four new phthalate‑ester testing items: DEHP—di(2‑ethylhexyl) phthalate; BBP—butyl benzyl phthalate; DBP—dibutyl phthalate; and DIBP—diisobutyl phthalate.
What are the limit values for the ten hazardous substances under RoHS?
1. Cadmium: less than 100 ppm.
2. Lead: less than 1000 ppm.
3. Mercury: less than 1000 ppm.
4. Hexavalent chromium: less than 1000 ppm.
5. Polybrominated biphenyls (PBB): less than 1000 ppm.
6. Polybrominated diphenyl ethers (PBDEs): less than 1000 ppm.
7. Di(2-ethylhexyl) phthalate (DEHP): less than 1000 ppm.
8. Butyl benzyl phthalate (BBP): less than 1000 ppm.
9. Dibutyl phthalate (DBP): less than 1000 ppm.
10. Diisobutyl phthalate (DIBP): less than 1000 ppm.
What instruments are typically used for RoHS testing?
The commonly used analytical instrument for RoHS 1.0 is X-ray fluorescence spectroscopy (XRF). XRF analyzers are available in standalone, benchtop, and handheld configurations. Typically, a handheld XRF analyzer is employed first for portable on-site testing, with a focus on components of the product that pose a higher risk of containing restricted substances.
With the introduction of RoHS 2.0, four phthalate esters have been added to the RoHS Directive, necessitating additional testing to determine the levels of these compounds, which are extracted using solvents. The solvent extracts are then analyzed for the presence of phthalates by gas chromatography–mass spectrometry (GC/MS) or gas chromatography with flame ionization detection (GC/FID). Other elements and polybrominated diphenyl ethers/alkylated elastomers are analyzed using comparative methods and GC/MS, respectively.
Do the ten RoHS items require two instruments for testing?
The ten items specified in RoHS require two instruments for testing.
For the ten‑item RoHS testing, two instruments are required to complete the analysis; a single instrument cannot achieve this. Even if one instrument is used to measure heavy metal content, the other must be employed to analyze organic compounds.
X-ray fluorescence spectrometers are used to detect RoHS 1.0, while chromatographs, pyrolyzers, or hyphenated techniques such as GC–MS are employed to analyze the four phthalates specified in RoHS 2.0. At present, no single instrument can independently perform all the tests required for RoHS 2.0.
Can ROHS testing instruments detect metallic elements?
ROHS testing instruments are specialized devices used to detect the presence of hazardous substances in electronic components, household appliances, toys, and other products. These harmful substances can pose long-term risks to both human health and the environment, making the use of ROHS testing equipment of paramount importance.
Regarding whether ROHS testing instruments can detect metallic elements, the answer is yes. This is because ROHS testing covers not only hazardous substances but also the analysis of metallic constituents in compliant products. These metals include lead, cadmium, mercury, and others, which are widely used during the manufacturing process. Although these elements are essential, their excessive use poses significant risks to human health and the environment. Consequently, the advent of ROHS testing instruments has greatly facilitated the detection of metallic elements.