Causes of Errors in XRF Coating Thickness Gauges!
发布时间: 2022-06-14 14:28
XRF is a relative analytical technique, which means that to obtain results, the data collected from an unknown sample must be compared with standard reference data on the instrument and correlated with established physical equations. Although this technique enjoys widespread acceptance and can be mastered by trained users, certain factors can affect the outcomes and introduce errors. Today, we will introduce… XRF coating thickness measurement Causes of instrument error!

1. Sample out of focus
This is a critical step in measuring parts. Focus must be maintained to keep a constant distance between the X-ray tube, the part, and the detector. X-ray intensity decreases with distance, so setting the tube–detector distance too far results in underestimation of thickness. Conversely, positioning the tube and detector too close to the sample leads to overestimation of thickness—this effect is even more pronounced in multilayer coatings, as incorrect distance data can introduce significant errors into the calculations.
2. The part orientation is incorrect.
For flat parts, the rotation angle is not an issue, as it does not affect the XRF coating thickness gauge signal. However, for curved parts, it is crucial to align the part’s axis with the axis of the X-ray tube and detector. This facilitates easier alignment and improved data reproducibility, ensuring that the X-ray beam strikes the top of convex features or the bottom of concave features rather than the sidewalls. As in the previous focusing scenario, misalignment also alters the distance between the X-ray tube, the sample, and the detector. In extreme cases, misalignment can prevent any XRF coating thickness gauge signal from reaching the detector.
3. Substrate Changes
Depending on the materials in both the coating and the substrate, elements present in the substrate can affect the performance of XRF coating thickness gauges. This is a well‑known fact; therefore, it is best to establish calibration procedures using materials that closely resemble those of the parts being measured. If the substrate of the part differs from the substrate used for the calibration standards, the results may be inaccurate. For example, consider a part with nickel (Ni) and gold (Au) coatings deposited on a bronze (CuSn) substrate. The tin (Sn) in the bronze substrate can effectively act as an auxiliary X‑ray source, enhancing the X‑ray fluorescence signal from the coating. If calibration is performed using a pure copper (Cu) substrate, the secondary fluorescence effect of tin (Sn) will not be properly accounted for in the calculation model, leading to incorrect measurements for nickel and gold.
4. Measurement values exceeding the calibration range
The relationship between coating thickness or composition and intensity (as measured by an XRF coating thickness gauge) is linear over a narrow range but may become nonlinear at larger ranges. Consequently, calibration curves are optimized to operate within a limited range of thicknesses and compositions, rather than spanning the entire analytical range. The optimal range is determined by the regression settings and the standards used to construct the calibration curve. Users can collaborate with the XRF coating thickness gauge manufacturer to define the appropriate calibration curve range and configure warning alerts in the user software when measured values fall outside this range.
5. Do not perform routine instrument adjustments or adjust the instrument too frequently.
XRF coating thickness gauges incorporate one or more methods to monitor instrument condition and automatically compensate for minute variations in the characteristics of the X-ray tube, detector, and electronics. It is crucial to perform these routine instrument adjustments at the intervals recommended by the manufacturer. If daily adjustments are advised but only performed monthly, the instrument may undergo continuous drift throughout that period. During adjustment runs, test results may exhibit gradual shifts. Running adjustments more frequently than recommended can have the opposite effect—causing the instrument to attempt numerous small, unnecessary corrections that, when accumulated, result in noticeable deviations. This phenomenon is known as “over‑correction.”
6. Adverse environmental conditions
In addition to variations in the analyzer’s temperature and humidity, other environmental factors can also affect XRF performance. Dust and corrosive chemicals in the atmosphere can interfere with XRF coating‑thickness measurements and prematurely degrade the performance of both the instrument itself and the components of the PC that controls it. The XRF coating‑thickness analyzer’s mains power supply influences the operation of electronic components, including the X‑ray tube power supply and detector electronics, potentially introducing measurement errors. In areas with unstable power supplies, it is advisable to install a line conditioner or an uninterruptible power supply (UPS). Whenever possible, position the analyzer in a space with a stable, reliable, and controlled power source, and maintain adequate clearance from other equipment on the shop floor to prevent personnel, workpieces, and mobile machinery from colliding with the instrument.
Latest News
Recommended Products
Frequently Asked Questions
What is RoHS?
RoHS is a mandatory standard established by EU legislation, officially known as the “Directive on the Restriction of the Use of Certain Hazardous Substances in Electrical and Electronic Equipment.” The standard entered into force on July 1, 2006, and primarily aims to regulate the materials and manufacturing processes of electrical and electronic products, thereby promoting human health and environmental protection.
The purpose of this standard is to eliminate six substances—lead, mercury, cadmium, hexavalent chromium, polybrominated biphenyls, and polybrominated diphenyl ethers (note: the correct Chinese term for PBDE is “polybrominated diphenyl ethers”; “polybrominated biphenyl ethers” is an incorrect designation)—and specifically stipulates that the cadmium content must not exceed 0.01%.
The latest RoHS Directive has added four new phthalate‑ester testing items: DEHP—di(2‑ethylhexyl) phthalate; BBP—butyl benzyl phthalate; DBP—dibutyl phthalate; and DIBP—diisobutyl phthalate.
What are the limit values for the ten hazardous substances under RoHS?
1. Cadmium: less than 100 ppm.
2. Lead: less than 1000 ppm.
3. Mercury: less than 1000 ppm.
4. Hexavalent chromium: less than 1000 ppm.
5. Polybrominated biphenyls (PBB): less than 1000 ppm.
6. Polybrominated diphenyl ethers (PBDEs): less than 1000 ppm.
7. Di(2-ethylhexyl) phthalate (DEHP): less than 1000 ppm.
8. Butyl benzyl phthalate (BBP): less than 1000 ppm.
9. Dibutyl phthalate (DBP): less than 1000 ppm.
10. Diisobutyl phthalate (DIBP): less than 1000 ppm.
What instruments are typically used for RoHS testing?
The commonly used analytical instrument for RoHS 1.0 is X-ray fluorescence spectroscopy (XRF). XRF analyzers are available in standalone, benchtop, and handheld configurations. Typically, a handheld XRF analyzer is employed first for portable on-site testing, with a focus on components of the product that pose a higher risk of containing restricted substances.
With the introduction of RoHS 2.0, four phthalate esters have been added to the RoHS Directive, necessitating additional testing to determine the levels of these compounds, which are extracted using solvents. The solvent extracts are then analyzed for the presence of phthalates by gas chromatography–mass spectrometry (GC/MS) or gas chromatography with flame ionization detection (GC/FID). Other elements and polybrominated diphenyl ethers/alkylated elastomers are analyzed using comparative methods and GC/MS, respectively.
Do the ten RoHS items require two instruments for testing?
The ten items specified in RoHS require two instruments for testing.
For the ten‑item RoHS testing, two instruments are required to complete the analysis; a single instrument cannot achieve this. Even if one instrument is used to measure heavy metal content, the other must be employed to analyze organic compounds.
X-ray fluorescence spectrometers are used to detect RoHS 1.0, while chromatographs, pyrolyzers, or hyphenated techniques such as GC–MS are employed to analyze the four phthalates specified in RoHS 2.0. At present, no single instrument can independently perform all the tests required for RoHS 2.0.
Can ROHS testing instruments detect metallic elements?
ROHS testing instruments are specialized devices used to detect the presence of hazardous substances in electronic components, household appliances, toys, and other products. These harmful substances can pose long-term risks to both human health and the environment, making the use of ROHS testing equipment of paramount importance.
Regarding whether ROHS testing instruments can detect metallic elements, the answer is yes. This is because ROHS testing covers not only hazardous substances but also the analysis of metallic constituents in compliant products. These metals include lead, cadmium, mercury, and others, which are widely used during the manufacturing process. Although these elements are essential, their excessive use poses significant risks to human health and the environment. Consequently, the advent of ROHS testing instruments has greatly facilitated the detection of metallic elements.