The principle and operating procedure of a spectrophotometer!
发布时间: 2022-06-13 11:18
Spectrometer Also known as Spectrophotometer It is a scientific instrument that separates light with complex components into its spectral lines. Today, we’ll introduce the principle and operating procedures of a spectrophotometer!
The measurement range of a spectrophotometer typically encompasses the visible region, with wavelengths from 380 to 780 nm, and the ultraviolet region, from 200 to 380 nm. Different light sources possess distinct emission spectra; accordingly, various luminous materials can serve as the instrument’s illumination source. Tungsten lamp emission spectrum: The spectrum emitted by a tungsten lamp, spanning 380–780 nm, is refracted by a prism, yielding a continuous color spectrum composed of red, orange, yellow, green, blue, indigo, and violet. This spectrum can be used as the visible-light source for a spectrophotometer.

I. Principle of the Spectrophotometer:
A spectrophotometer uses a light source capable of generating multiple wavelengths. Through a series of spectroscopic components, it produces light at specific wavelengths. After the light passes through the sample being analyzed, part of it is absorbed; the absorbance of the sample is then calculated and converted into its concentration. The absorbance of the sample is directly proportional to its concentration.
When monochromatic light passes through a solution of the substance being measured, the amount of light absorbed by the substance is directly proportional to both the concentration of the substance and the thickness of the liquid layer (the optical path length), as described by the following relationship:
A = -log(I/I0) = -logT = kLc
In the equation: A denotes absorbance;
I0 is the intensity of the incident monochromatic light;
I is the intensity of the transmitted monochromatic light;
T——the transmittance of the substance;
k is the molar absorptivity;
L is the optical path length of the substance to be analyzed, i.e., the side length of the cuvette.
c is the concentration of the substance;
The wavelengths at which a substance selectively absorbs light, along with the corresponding absorption coefficients, are physical constants of that substance. When the absorbance coefficient of a pure substance is known under specified conditions, the sample can be prepared as a solution under the same conditions, its absorbance measured, and the content of the target substance in the sample calculated using the aforementioned formula. In the visible region, while some substances absorb light, many do not exhibit intrinsic absorption; however, their presence can be detected by adding a color‑developing reagent or subjecting the sample to a color‑development procedure under appropriate conditions. This approach is therefore also referred to as colorimetric analysis. Because numerous factors can influence the intensity of the color developed, and instruments often employed in such analyses may have limited monochromatic light purity, it is essential to use standard or reference materials concurrently during measurements.
II. Operating Procedures for the Spectrophotometer:
1. Turn on the power, switch on the instrument, and open the dark-box cover of the sample chamber; allow it to warm up for 10 minutes.
2. Set the sensitivity switch to the “1” position (if the zero-adjustment knob is not set to “0,” select a higher range).
3. Rotate the wavelength selector knob to the desired wavelength.
4. Pour the blank solution and the sample solution into the cuvettes to approximately three-quarters of their volume, wipe the exterior surfaces with lens paper, place the cuvettes in the sample holder, and align the blank cuvette with the light path.
5. Adjust the zero-adjustment knob with the dark-box cover open, so that the dial pointer indicates t = 0.
6. Close the dark box lid, adjust the “100” knob until the transmittance of the blank tube reaches 100%, and once the pointer has stabilized, gradually pull out the sample slide to read and record the absorbance value of the measurement tube.
7. After completing the colorimetric measurement, turn off the power, remove the cuvette for cleaning, and wipe the sample chamber with a soft cloth or soft paper.
Latest News
Recommended Products
Frequently Asked Questions
What is RoHS?
RoHS is a mandatory standard established by EU legislation, officially known as the “Directive on the Restriction of the Use of Certain Hazardous Substances in Electrical and Electronic Equipment.” The standard entered into force on July 1, 2006, and primarily aims to regulate the materials and manufacturing processes of electrical and electronic products, thereby promoting human health and environmental protection.
The purpose of this standard is to eliminate six substances—lead, mercury, cadmium, hexavalent chromium, polybrominated biphenyls, and polybrominated diphenyl ethers (note: the correct Chinese term for PBDE is “polybrominated diphenyl ethers”; “polybrominated biphenyl ethers” is an incorrect designation)—and specifically stipulates that the cadmium content must not exceed 0.01%.
The latest RoHS Directive has added four new phthalate‑ester testing items: DEHP—di(2‑ethylhexyl) phthalate; BBP—butyl benzyl phthalate; DBP—dibutyl phthalate; and DIBP—diisobutyl phthalate.
What are the limit values for the ten hazardous substances under RoHS?
1. Cadmium: less than 100 ppm.
2. Lead: less than 1000 ppm.
3. Mercury: less than 1000 ppm.
4. Hexavalent chromium: less than 1000 ppm.
5. Polybrominated biphenyls (PBB): less than 1000 ppm.
6. Polybrominated diphenyl ethers (PBDEs): less than 1000 ppm.
7. Di(2-ethylhexyl) phthalate (DEHP): less than 1000 ppm.
8. Butyl benzyl phthalate (BBP): less than 1000 ppm.
9. Dibutyl phthalate (DBP): less than 1000 ppm.
10. Diisobutyl phthalate (DIBP): less than 1000 ppm.
What instruments are typically used for RoHS testing?
The commonly used analytical instrument for RoHS 1.0 is X-ray fluorescence spectroscopy (XRF). XRF analyzers are available in standalone, benchtop, and handheld configurations. Typically, a handheld XRF analyzer is employed first for portable on-site testing, with a focus on components of the product that pose a higher risk of containing restricted substances.
With the introduction of RoHS 2.0, four phthalate esters have been added to the RoHS Directive, necessitating additional testing to determine the levels of these compounds, which are extracted using solvents. The solvent extracts are then analyzed for the presence of phthalates by gas chromatography–mass spectrometry (GC/MS) or gas chromatography with flame ionization detection (GC/FID). Other elements and polybrominated diphenyl ethers/alkylated elastomers are analyzed using comparative methods and GC/MS, respectively.
Do the ten RoHS items require two instruments for testing?
The ten items specified in RoHS require two instruments for testing.
For the ten‑item RoHS testing, two instruments are required to complete the analysis; a single instrument cannot achieve this. Even if one instrument is used to measure heavy metal content, the other must be employed to analyze organic compounds.
X-ray fluorescence spectrometers are used to detect RoHS 1.0, while chromatographs, pyrolyzers, or hyphenated techniques such as GC–MS are employed to analyze the four phthalates specified in RoHS 2.0. At present, no single instrument can independently perform all the tests required for RoHS 2.0.
Can ROHS testing instruments detect metallic elements?
ROHS testing instruments are specialized devices used to detect the presence of hazardous substances in electronic components, household appliances, toys, and other products. These harmful substances can pose long-term risks to both human health and the environment, making the use of ROHS testing equipment of paramount importance.
Regarding whether ROHS testing instruments can detect metallic elements, the answer is yes. This is because ROHS testing covers not only hazardous substances but also the analysis of metallic constituents in compliant products. These metals include lead, cadmium, mercury, and others, which are widely used during the manufacturing process. Although these elements are essential, their excessive use poses significant risks to human health and the environment. Consequently, the advent of ROHS testing instruments has greatly facilitated the detection of metallic elements.