The working principle of an X-ray fluorescence coating thickness gauge!
发布时间: 2022-06-10 11:24
We all know X-ray fluorescence. Coating thickness gauge It is commonly used to measure the thickness of metal coatings, and many people are likely curious about how an X-ray fluorescence coating thickness gauge works. Today, we’ll provide a brief explanation on this topic.
If an electron is ejected from an orbital, electrons from higher energy levels will naturally transition to its vacant position, thereby achieving a stable configuration. This interlevel transition releases energy in the form of X-rays. Since the energy levels of atoms differ for each element, the energy differences between orbitals also vary from one element to another.
The following can describe the characteristics of X-ray fluorescence: When X-ray fluorescence arises from the transfer of an electron into the K shell, and an electron in the K shell has previously been ionized, another electron takes its place; this is referred to as K‑radiation. Different transitions between energy levels yield different energies: for example, Kα radiation corresponds to an electron transitioning from the L shell to the K shell, whereas Kβ radiation results from an electron jumping from the M shell to the K shell—these two processes are distinct. If X-ray fluorescence involves an electron filling an empty orbital in the L shell, such emission is termed L‑radiation. Similarly, L‑radiation can be further subdivided into Lα radiation, which occurs when an electron from the M shell transitions to the L shell, and Lβ radiation, which arises when an electron from the N shell makes the transition to the L shell. Because the energy of Kβ radiation is approximately 11% that of Kα, and the energy of Lβ radiation is about 20% higher than that of Lα, Lα and Lβ can be readily distinguished on the basis of their energies.
The properties of an atom are determined by its atomic number, which corresponds to the number of protons or the number of electrons in its electron shell; as illustrated, there is a specific relationship between X-ray energy and atomic number. K‑radiation has much higher energy than L‑radiation, and different atomic numbers give rise to distinct energy differences. Specific X-rays can be detected using a proportional counter. When radiation strikes the counter, it is converted into electrical pulses. The pulse height output by the circuit is directly proportional to the energy of the incident radiation. X-rays emitted by particular materials can then be recorded by subsequent discrimination circuits. Thickness measurement based on X-ray fluorescence involves placing the sample within the instrument and exposing the region of interest to X-ray irradiation. At this point, characteristic X-rays are generated from the coating, the substrate, and any intermediate layers; the detection system converts these X-rays into proportional electrical signals, which are recorded by the instrument. By measuring the intensity of the X-rays, the thickness of the coating can be determined.
In certain cases—such as measuring IC leads on printed circuit boards, contact pins, and conductor components—higher measurement accuracy is required. Generally speaking, the measurement of coating thickness must essentially meet the following requirements:
1. High precision under non-destructive measurement.
2. Extremely small measurement area.
3. The intermediate coating and the material composition do not affect the measurement results.
4. Simultaneous, non-interfering measurement of both the top and intermediate coatings.
5. Simultaneously measure the coating thickness and composition of the bimetallic alloy.
In contrast, X-ray fluorescence spectroscopy can deliver highly precise measurements unaffected by the material or the presence of various intermediate layers.
The above is an overview of the working principle of X-ray fluorescence coating thickness gauges. If you have a need for this instrument, please contact or consult us online, and we will promptly provide you with a quotation and product information.
Latest News
Recommended Products
Frequently Asked Questions
What is RoHS?
RoHS is a mandatory standard established by EU legislation, officially known as the “Directive on the Restriction of the Use of Certain Hazardous Substances in Electrical and Electronic Equipment.” The standard entered into force on July 1, 2006, and primarily aims to regulate the materials and manufacturing processes of electrical and electronic products, thereby promoting human health and environmental protection.
The purpose of this standard is to eliminate six substances—lead, mercury, cadmium, hexavalent chromium, polybrominated biphenyls, and polybrominated diphenyl ethers (note: the correct Chinese term for PBDE is “polybrominated diphenyl ethers”; “polybrominated biphenyl ethers” is an incorrect designation)—and specifically stipulates that the cadmium content must not exceed 0.01%.
The latest RoHS Directive has added four new phthalate‑ester testing items: DEHP—di(2‑ethylhexyl) phthalate; BBP—butyl benzyl phthalate; DBP—dibutyl phthalate; and DIBP—diisobutyl phthalate.
What are the limit values for the ten hazardous substances under RoHS?
1. Cadmium: less than 100 ppm.
2. Lead: less than 1000 ppm.
3. Mercury: less than 1000 ppm.
4. Hexavalent chromium: less than 1000 ppm.
5. Polybrominated biphenyls (PBB): less than 1000 ppm.
6. Polybrominated diphenyl ethers (PBDEs): less than 1000 ppm.
7. Di(2-ethylhexyl) phthalate (DEHP): less than 1000 ppm.
8. Butyl benzyl phthalate (BBP): less than 1000 ppm.
9. Dibutyl phthalate (DBP): less than 1000 ppm.
10. Diisobutyl phthalate (DIBP): less than 1000 ppm.
What instruments are typically used for RoHS testing?
The commonly used analytical instrument for RoHS 1.0 is X-ray fluorescence spectroscopy (XRF). XRF analyzers are available in standalone, benchtop, and handheld configurations. Typically, a handheld XRF analyzer is employed first for portable on-site testing, with a focus on components of the product that pose a higher risk of containing restricted substances.
With the introduction of RoHS 2.0, four phthalate esters have been added to the RoHS Directive, necessitating additional testing to determine the levels of these compounds, which are extracted using solvents. The solvent extracts are then analyzed for the presence of phthalates by gas chromatography–mass spectrometry (GC/MS) or gas chromatography with flame ionization detection (GC/FID). Other elements and polybrominated diphenyl ethers/alkylated elastomers are analyzed using comparative methods and GC/MS, respectively.
Do the ten RoHS items require two instruments for testing?
The ten items specified in RoHS require two instruments for testing.
For the ten‑item RoHS testing, two instruments are required to complete the analysis; a single instrument cannot achieve this. Even if one instrument is used to measure heavy metal content, the other must be employed to analyze organic compounds.
X-ray fluorescence spectrometers are used to detect RoHS 1.0, while chromatographs, pyrolyzers, or hyphenated techniques such as GC–MS are employed to analyze the four phthalates specified in RoHS 2.0. At present, no single instrument can independently perform all the tests required for RoHS 2.0.
Can ROHS testing instruments detect metallic elements?
ROHS testing instruments are specialized devices used to detect the presence of hazardous substances in electronic components, household appliances, toys, and other products. These harmful substances can pose long-term risks to both human health and the environment, making the use of ROHS testing equipment of paramount importance.
Regarding whether ROHS testing instruments can detect metallic elements, the answer is yes. This is because ROHS testing covers not only hazardous substances but also the analysis of metallic constituents in compliant products. These metals include lead, cadmium, mercury, and others, which are widely used during the manufacturing process. Although these elements are essential, their excessive use poses significant risks to human health and the environment. Consequently, the advent of ROHS testing instruments has greatly facilitated the detection of metallic elements.