Working principle of the RoHS detector!
发布时间: 2022-06-08 09:46
We all know RoHS instrument It is used to detect the six hazardous substances and four phthalates specified in the RoHS directive. So today, let’s talk about… RoHS detector The working principle—those who are interested can take a look!
After the sample analysis is completed, the ROHS instrument automatically generates a test report according to the report template configured in the calibration curve, and displays a “Print Report Preview” dialog box. The report can also be directly sent to a printer for hard‑copy archiving, with data saved automatically and historical records accessible on the computer.

1. Test Conditions
Environmental requirements: Maintain the ambient temperature between 10 and 30°C, with relative humidity below 85% to ensure proper operation and storage. Avoid high room temperatures, strong electromagnetic interference, and physical vibrations. 4.2 External Power Supply: A 1000 W regulated power supply is recommended.
Power-on procedure: First turn on the device and any external peripherals, then power on the computer.
Measurement preparation: Ensure the instrument is properly connected to the computer, and allow it to warm up for approximately 30 minutes after power-up.
2. Instruction Operation
Turn on the software: After powering on the computer, switch on the instrument’s power supply, then double-click the desktop icon to launch the program. The program will prompt you to log in.
Select the corresponding carrier and enter your password to log in.
Preheating: In the X-ray tube settings dialog box, select both the “Turn On High-Voltage Power Supply” and “Gradually Increase Tube Voltage and Tube Current” options, then click OK. Once the tube voltage and tube current reach their set values, the instrument will be preheated for 30 minutes.
Silver standard calibration adjustment: Calibrate the instrument using a silver standard calibration plate. Open the instrument’s top cover, place the silver standard calibration plate inside, and then click the “Fine Adjustment” button to perform the calibration. Once the adjustment is complete and verified, a dialog box will appear; click “OK.”
3. Comments
ROHS instruments generate ionizing radiation; during testing, do not point the instrument at people. Do not perform tests in open air, and do not handle samples with your hands. While the ROHS tester is operating, do not open the protective cover or aim the instrument at a human body. The ROHS instrument must be calibrated each time it is powered on and after every four consecutive hours of operation. Only authorized personnel are permitted to operate the instrument. All operations must be performed strictly in accordance with the prescribed procedures. Any improper use may compromise the instrument’s test results and shorten its service life.
4. Preventive Measures
After completing the test, click the “Set X-ray Tube” button, then in the dialog box that appears, click the “Turn On High-Voltage Power Supply” button, and finally click “OK.” Once both the tube voltage and tube current are within acceptable limits, you may proceed to the next step. Click the “Close” button in the upper-right corner of the operator interface; only after closing the software can you shut down the hardware.
The above explains the operating principle of RoHS detectors. If you have any needs for a RoHS detector, please contact us online or send us an inquiry—we will get back to you promptly and provide you with relevant preferential pricing, technical specifications, and other details.
Latest News
Recommended Products
Frequently Asked Questions
What is RoHS?
RoHS is a mandatory standard established by EU legislation, officially known as the “Directive on the Restriction of the Use of Certain Hazardous Substances in Electrical and Electronic Equipment.” The standard entered into force on July 1, 2006, and primarily aims to regulate the materials and manufacturing processes of electrical and electronic products, thereby promoting human health and environmental protection.
The purpose of this standard is to eliminate six substances—lead, mercury, cadmium, hexavalent chromium, polybrominated biphenyls, and polybrominated diphenyl ethers (note: the correct Chinese term for PBDE is “polybrominated diphenyl ethers”; “polybrominated biphenyl ethers” is an incorrect designation)—and specifically stipulates that the cadmium content must not exceed 0.01%.
The latest RoHS Directive has added four new phthalate‑ester testing items: DEHP—di(2‑ethylhexyl) phthalate; BBP—butyl benzyl phthalate; DBP—dibutyl phthalate; and DIBP—diisobutyl phthalate.
What are the limit values for the ten hazardous substances under RoHS?
1. Cadmium: less than 100 ppm.
2. Lead: less than 1000 ppm.
3. Mercury: less than 1000 ppm.
4. Hexavalent chromium: less than 1000 ppm.
5. Polybrominated biphenyls (PBB): less than 1000 ppm.
6. Polybrominated diphenyl ethers (PBDEs): less than 1000 ppm.
7. Di(2-ethylhexyl) phthalate (DEHP): less than 1000 ppm.
8. Butyl benzyl phthalate (BBP): less than 1000 ppm.
9. Dibutyl phthalate (DBP): less than 1000 ppm.
10. Diisobutyl phthalate (DIBP): less than 1000 ppm.
What instruments are typically used for RoHS testing?
The commonly used analytical instrument for RoHS 1.0 is X-ray fluorescence spectroscopy (XRF). XRF analyzers are available in standalone, benchtop, and handheld configurations. Typically, a handheld XRF analyzer is employed first for portable on-site testing, with a focus on components of the product that pose a higher risk of containing restricted substances.
With the introduction of RoHS 2.0, four phthalate esters have been added to the RoHS Directive, necessitating additional testing to determine the levels of these compounds, which are extracted using solvents. The solvent extracts are then analyzed for the presence of phthalates by gas chromatography–mass spectrometry (GC/MS) or gas chromatography with flame ionization detection (GC/FID). Other elements and polybrominated diphenyl ethers/alkylated elastomers are analyzed using comparative methods and GC/MS, respectively.
Do the ten RoHS items require two instruments for testing?
The ten items specified in RoHS require two instruments for testing.
For the ten‑item RoHS testing, two instruments are required to complete the analysis; a single instrument cannot achieve this. Even if one instrument is used to measure heavy metal content, the other must be employed to analyze organic compounds.
X-ray fluorescence spectrometers are used to detect RoHS 1.0, while chromatographs, pyrolyzers, or hyphenated techniques such as GC–MS are employed to analyze the four phthalates specified in RoHS 2.0. At present, no single instrument can independently perform all the tests required for RoHS 2.0.
Can ROHS testing instruments detect metallic elements?
ROHS testing instruments are specialized devices used to detect the presence of hazardous substances in electronic components, household appliances, toys, and other products. These harmful substances can pose long-term risks to both human health and the environment, making the use of ROHS testing equipment of paramount importance.
Regarding whether ROHS testing instruments can detect metallic elements, the answer is yes. This is because ROHS testing covers not only hazardous substances but also the analysis of metallic constituents in compliant products. These metals include lead, cadmium, mercury, and others, which are widely used during the manufacturing process. Although these elements are essential, their excessive use poses significant risks to human health and the environment. Consequently, the advent of ROHS testing instruments has greatly facilitated the detection of metallic elements.