Common Issues in Gas Chromatographs and Their Solutions!
发布时间: 2022-06-01 09:48
Gas chromatograph It is an organic multi-component analytical instrument with extremely broad applications. It boasts high separation efficiency, rapid analysis speed, and the capability to measure multiple components simultaneously. In chemical analysis, it occupies a pivotal position; nearly 80% of raw material and product analyses are performed using gas chromatographs. Consequently, troubleshooting common malfunctions of gas chromatographs is of paramount importance. Below, we will outline solutions for typical gas chromatograph issues.

I. Incomplete separation
(1) Several peaks overlap and cannot be resolved.
Remedial measures: Reduce the carrier gas flow rate, decrease the injection volume, and lower the column temperature. If a compound that was previously completely resolved can no longer be fully separated after a certain period, this indicates that the stationary phase has degraded and the chromatographic column has reached the end of its useful life; the stationary phase should be replaced.
(2) An excessively long separation time causes the late-eluting peak to become broad and flattened.
Remedy: This can be addressed by increasing the column temperature.
(3) The detector sensitivity is too low, resulting in the inability to detect components present at low concentrations.
Solution: This can be addressed by increasing the injection volume and enhancing detector sensitivity.
II. Irregular Peak Shape
(1) A trailing peak is observed.
Remedy: Use a highly polar stationary phase, eliminate the activity of the support material, and increase the column temperature.
(2) Exhibits a flat-topped shape or peak.
Remedy: Address the issue by reducing the injection volume, increasing the column temperature, and raising the carrier gas flow rate. Additionally, a flat‑topped peak may also arise when the amplifier input is saturated.
III. Effects of Detector Formation
Take the thermal conductivity detector (TCD) as an example: it exploits the difference in thermal conductivity between the carrier gas and the analyte. The resulting unbalanced voltage in the bridge circuit is directly proportional to the concentration of the component being measured, thereby enabling quantitative analysis of that component.
(1) The TCD detector may become contaminated, resulting in baseline drift or a step-like baseline, and may also exhibit high noise levels.
(2) The TCD heating wire has burned out, causing the baseline to drop to zero.
(3) The TCD power supply is unstable, resulting in irregular pulse interference peaks.
IV. The Influence of the Carrier Gas
The carrier gas carries the sample through the stationary phase, and the separated components are sequentially eluted from the chromatographic column by the carrier gas and directed to the detector for analysis. Operational parameters such as the carrier gas flow rate, its properties, and the carrier gas pressure all influence the efficiency of chromatographic separation.
(1) A low carrier gas flow rate can lead to increased retention times, reduced sensitivity, or the appearance of rounded peaks and tailing peaks.
(2) An excessively high carrier gas flow rate can cause excessive noise or poor peak separation.
(3) Unstable carrier gas control results in irregular baseline drift or wavy baseline drift.
In the above situation, check whether the pressure-reducing valve has exceeded its service limits; if necessary, replace the valve, and then verify that there are no leaks in the carrier gas line.
V. Circuit Issues
Circuit faults are generally relatively easy to diagnose, such as the power supply failing to start, the detector or injector not heating, or the temperature‑controlled oven unable to maintain a constant temperature. If the baseline exhibits a periodic sine wave, this is typically caused by a malfunction in the amplification circuit; the usual remedy is to replace the faulty electronic components.
In routine analyses, you may also encounter issues not covered above—for example, the hydrogen flame detector fails to ignite. First, verify that both hydrogen and air have been turned on, then check whether the combustion coil is functioning properly. If all three of these conditions are met yet ignition still does not occur, the problem might be a gas leak at the detector–column interface. As for situations where an inverted peak appears, this could indicate that the polarity of the instrument or the data processor has been reversed; in such cases, begin by inspecting the instrument’s polarity settings. Finally, if the injected sample amount does not match the integrated peak area, the output signal cable may have been connected incorrectly.
The above outlines several common issues encountered in gas chromatographs and their corresponding solutions. However, in practical applications, it is not uncommon for multiple malfunctions to occur simultaneously. In such cases, a careful analysis and diagnosis based on the specific symptoms are essential, followed by the systematic application of the aforementioned troubleshooting methods to restore the instrument to normal operation. If you have any gas chromatography equipment needs, please feel free to contact or consult us online—we will promptly provide you with a detailed quotation and relevant technical specifications.
Latest News
Recommended Products
Frequently Asked Questions
What is RoHS?
RoHS is a mandatory standard established by EU legislation, officially known as the “Directive on the Restriction of the Use of Certain Hazardous Substances in Electrical and Electronic Equipment.” The standard entered into force on July 1, 2006, and primarily aims to regulate the materials and manufacturing processes of electrical and electronic products, thereby promoting human health and environmental protection.
The purpose of this standard is to eliminate six substances—lead, mercury, cadmium, hexavalent chromium, polybrominated biphenyls, and polybrominated diphenyl ethers (note: the correct Chinese term for PBDE is “polybrominated diphenyl ethers”; “polybrominated biphenyl ethers” is an incorrect designation)—and specifically stipulates that the cadmium content must not exceed 0.01%.
The latest RoHS Directive has added four new phthalate‑ester testing items: DEHP—di(2‑ethylhexyl) phthalate; BBP—butyl benzyl phthalate; DBP—dibutyl phthalate; and DIBP—diisobutyl phthalate.
What are the limit values for the ten hazardous substances under RoHS?
1. Cadmium: less than 100 ppm.
2. Lead: less than 1000 ppm.
3. Mercury: less than 1000 ppm.
4. Hexavalent chromium: less than 1000 ppm.
5. Polybrominated biphenyls (PBB): less than 1000 ppm.
6. Polybrominated diphenyl ethers (PBDEs): less than 1000 ppm.
7. Di(2-ethylhexyl) phthalate (DEHP): less than 1000 ppm.
8. Butyl benzyl phthalate (BBP): less than 1000 ppm.
9. Dibutyl phthalate (DBP): less than 1000 ppm.
10. Diisobutyl phthalate (DIBP): less than 1000 ppm.
What instruments are typically used for RoHS testing?
The commonly used analytical instrument for RoHS 1.0 is X-ray fluorescence spectroscopy (XRF). XRF analyzers are available in standalone, benchtop, and handheld configurations. Typically, a handheld XRF analyzer is employed first for portable on-site testing, with a focus on components of the product that pose a higher risk of containing restricted substances.
With the introduction of RoHS 2.0, four phthalate esters have been added to the RoHS Directive, necessitating additional testing to determine the levels of these compounds, which are extracted using solvents. The solvent extracts are then analyzed for the presence of phthalates by gas chromatography–mass spectrometry (GC/MS) or gas chromatography with flame ionization detection (GC/FID). Other elements and polybrominated diphenyl ethers/alkylated elastomers are analyzed using comparative methods and GC/MS, respectively.
Do the ten RoHS items require two instruments for testing?
The ten items specified in RoHS require two instruments for testing.
For the ten‑item RoHS testing, two instruments are required to complete the analysis; a single instrument cannot achieve this. Even if one instrument is used to measure heavy metal content, the other must be employed to analyze organic compounds.
X-ray fluorescence spectrometers are used to detect RoHS 1.0, while chromatographs, pyrolyzers, or hyphenated techniques such as GC–MS are employed to analyze the four phthalates specified in RoHS 2.0. At present, no single instrument can independently perform all the tests required for RoHS 2.0.
Can ROHS testing instruments detect metallic elements?
ROHS testing instruments are specialized devices used to detect the presence of hazardous substances in electronic components, household appliances, toys, and other products. These harmful substances can pose long-term risks to both human health and the environment, making the use of ROHS testing equipment of paramount importance.
Regarding whether ROHS testing instruments can detect metallic elements, the answer is yes. This is because ROHS testing covers not only hazardous substances but also the analysis of metallic constituents in compliant products. These metals include lead, cadmium, mercury, and others, which are widely used during the manufacturing process. Although these elements are essential, their excessive use poses significant risks to human health and the environment. Consequently, the advent of ROHS testing instruments has greatly facilitated the detection of metallic elements.