Sample testing requirements for different ROHS 2.0 instruments!
发布时间: 2022-05-18 13:43
We all know testing. ROHS 2.0 compliant instrument There are many types, so what are the sample‑testing requirements for each kind of RoHS 2.0 analyzer? Despu Instruments has compiled some information—please see below for details!

1. Infrared spectrometer
To protect the instrument and ensure the quality of the sample’s infrared spectrum, samples submitted for analysis on this instrument must:
(1) The sample must be pre-purified to ensure adequate purity.
(2) For samples that are hygroscopic, users are advised to bring their own desiccator for storage.
(3) Samples must be pre‑dried to remove moisture, both to protect the instrument and to avoid water‑peak interference in the sample spectrum.
(4) For samples that are easily volatilized, prone to oxidation, or thermally unstable, please use containers equipped with sealed lids or stoppers and ensure they are tightly closed; this information must also be clearly indicated on the sample analysis request form.
(5) For samples that are toxic or corrosive, users must package them in sealed containers. When submitting the samples, it is essential to clearly indicate this information both on the label of the sample vial and on the analysis request form.
2. Nuclear Magnetic Resonance Spectrometer
(1) The purity of the sample submitted for testing should generally be greater than 95%, with no impurities such as iron filings, dust, or filter paper fibers. The typical amount of sample required for organic compounds is as follows:
(2) Please provide, to the extent possible, the sample’s structural information or origin. If there are any specific requirements (e.g., measurement temperature, spectral width, etc.), please specify them.
(3) If the instrument is capable of analyzing only liquid samples, the sample must be readily soluble in a suitable deuterated solvent; the submitter should select an appropriate solvent beforehand. Commonly available deuterated solvents include chloroform, methanol, heavy water, acetone, benzene, DMSO, o‑dichlorobenzene, acetic acid, acetonitrile, trifluoroacetic acid, and pyridine.
3. Liquid Chromatography–Mass Spectrometry System
(1) Flammable, explosive, toxic, and corrosive samples must be clearly labeled.
(2) Please provide, whenever possible, the compound’s structural formula, molecular weight, or functional groups present, to facilitate the selection of an appropriate ionization method. If there are any specific requirements, please also supply detailed experimental conditions.
(3) To ensure the accuracy and reliability of the analytical results, samples must be completely dissolved, with no mechanical impurities present. For samples that have not yet been dissolved, please specify the solvent; for samples already in solution, please indicate the concentration.
(4) In liquid chromatography–mass spectrometry, all buffer systems must be prepared using volatile buffers, such as acetic acid, ammonium acetate, and tetrabutylammonium hydroxide. For quantitative analyses, please provide standard reference materials.
4. Organic Mass Spectrometer
It is suitable for analyzing liquid and solid organic compound samples with molecular weights ranging from 50 to 2000 u; the sample should, whenever possible, be a pure single component.
5. Gas chromatography–mass spectrometry system
Gas chromatographs all employ capillary columns (packed columns are not suitable). The sample introduced into the gas chromatograph must be completely vaporized within the operating temperature range of the column.
6. UV-Vis Spectrophotometer
(1) Solid sample amount > 0.2 g; liquid sample amount > 2 ml.
(2) The concentration of the sample solution must be appropriate, and it must be clear and transparent, with no bubbles or suspended particles present.
7. Time-of-Flight Mass Spectrometer
(1) Solubility of the sample: The analyte must be soluble in an appropriate solvent; ideally, it should be a solid that is not dissolved or a pure liquid. If the sample is already in solution, information on the solvent, concentration, or content should be provided.
(2) Sample Type, Composition, and Amount This instrument excels at analyzing peptides and proteins, but it can also be used to characterize other biomacromolecules such as polysaccharides, nucleic acids, and high‑molecular‑weight polymers, as well as synthetic oligomers and certain low‑molecular‑weight organic compounds, including C60 and its derivatives. The sample may consist of a single component or multiple components; however, the greater the number of components, the more complex the spectrum and the more challenging its interpretation. Moreover, if inter‑component suppression occurs during ionization, it cannot be guaranteed that all individual components will produce observable peaks. Typical injection amounts range from approximately 1 to 10 picomoles per microliter.
(3) To obtain high‑quality mass spectra, peptide and protein samples should be free of sodium chloride, urea, calcium chloride, potassium hydrogen phosphate, trinitrotoluene, Tween, dimethyl sulfoxide, glycerol, sodium dodecyl sulfate, and other such contaminants. If the sample cannot be kept free of these reagents during pretreatment, it must be purified by dialysis or high‑performance liquid chromatography. Water, ammonium bicarbonate, ammonium acetate, ammonium formate, acetonitrile, trifluoroacetic acid, and similar reagents are suitable for sample purification. After purification, protein samples should be lyophilized whenever possible. Salts in the sample can be removed by ion‑exchange chromatography.
8. Elemental Analyzer
(1) The sample quantity should be sufficient to meet the linearity and sensitivity requirements of the method and the instrument.
(2) Complete the elemental analysis sample submission registration form, providing, to the extent possible, the molecular formula, the theoretical elemental content, or other relevant information.
(3) The sample must be a homogeneous solid particulate or liquid free of adsorbed water and must undergo purification. If the sample is impure—containing adsorbed water, organic solvents, inorganic salts, or other impurities—it can compromise the analytical results, leading to discrepancies between measured and calculated values.
9. Gas chromatograph
Samples that can be directly analyzed should be volatile and thermally stable, with boiling points generally not exceeding 300°C. Samples that cannot be injected directly require prior sample preparation. For liquid chromatography, the sample must be dry, and it is preferable to provide the structure of the analyte(s) of interest. For complex samples, it is advisable to specify any other components that may be present.
10. Plasma Atomic Emission Spectrometer
(1) Please provide information on the sample’s origin, type, and properties (e.g., ore, alloy, silicate, special solid solution, polymer, etc.).
List, to the extent possible, the principal components, impurity components, and their (estimated) concentrations; specify the lowest (estimated) concentration of each element to be tested. For solutions, clearly indicate the composition of the matrix (solvent, types of acids and bases, and their (estimated) concentrations), and state whether fluorine (F⁻) is present, as fluorine (F⁻) can severely corrode the nebulizer.
(2) Solid samples must be prepared as solutions free of all organic matter, with the final acidity adjusted to 1 mol. The sample volume should be 5–50 mL. If the sample contains suspended particles or precipitates, filtration is required; additionally, a reagent blank solution should be submitted concurrently for background correction.
(3) Samples must be prepared as solutions before being sent to the testing center.
① Substances that are prone to volatilization and loss during sample heating and acid dissolution (e.g., B, Hg, S‑2, Se, and Si when using hydrofluoric acid (HF) for sample digestion);
② Ceramics, glass, and other materials that cannot be dissolved by inorganic acids and can only be fused with alkalis;
③ In the ashing of organosilicon compounds, silicone rubbers, plastic products, and fibrous materials, or in their subsequent acid digestion at temperatures below 500°C:
A. Substances that are easily evaporated and lost;
B. Substances that cannot be ashed or dissolved (e.g., B, Bi, Ge, Hg, Os, Ru, Sb, Se, Sn, Tl, and Si when the sample is digested with hydrofluoric acid [HF]; as well as special solid solutions, polymers, etc.).
11. Ion Chromatograph
The sample submitted for analysis must be soluble in water or in dilute acid or dilute alkali, and the acids or bases used must not contain the analyte ion. For samples that contain the element of interest but exist as non‑ionic compounds in aqueous, acidic, or alkaline solutions, appropriate sample pretreatment is required.
12. Thermogravimetric Analyzer
Sample quantity: not less than 30 mg. When submitting the sample, please specify the temperature range for testing, the experimental atmosphere (air, N₂, or Ar), the heating rate, and the gas flow rate (if there are any special requirements).
13. Atomic Fluorescence Spectrometer
(1) General requirements for sample analysis
The atomic fluorescence spectrometer analyzes arsenic (As), selenium (Se), germanium (Ge), tellurium (Te), and mercury (Hg) in their ionic forms; the sample must be an aqueous solution or soluble in acid.
(2) Solid Samples
For the determination of germanium (Ge), the medium is sulfuric acid (5% v/v); for arsenic (As), selenium (Se), tellurium (Te), and mercury (Hg), the medium is hydrochloric acid (5% v/v); for mercury (Hg), nitric acid (5% v/v) may also be used as the medium, and for arsenic (As), sulfuric acid (2% v/v) may likewise serve as the medium. Because metals such as copper, silver, gold, and platinum exert significant interferences on the analytes, it is not advisable to use this instrument for determining arsenic, tellurium, selenium, or mercury in alloys containing these metals. For organic or biological solid samples, the sample is subjected to digestion to form a solution while maintaining an appropriate acidity; the acidity of the matrix is the same as that used for inorganic samples.
(3) The limit of quantitation for the analyte in the sample is determined by the instrument’s sensitivity and the analytical method. The concentration range of the analyte in the sample should be between 0.05 μg/g and 500 μg/g; samples outside this concentration range cannot reliably yield accurate results when analyzed with this instrument.
(4) Sample Quantity
For each element analyzed, the required sample quantity is no less than 2 g for solid samples, no less than 20 mL for liquid samples, and no less than 100 mL for water samples.
(5) Prior to sample submission, please consult relevant literature and provide as much pertinent information as possible.
Differential Scanning Calorimeter: For solid samples, no decomposition or thermal degradation occurs within the tested temperature range, and no volatiles are released. Sample quantity: For inorganic or organic materials, at least 20 mg per run; for pharmaceuticals, at least 5 mg. When submitting samples, please specify the experimental conditions, including the temperature range, heating/cooling rates, and isothermal holding times.
14. Transmission Electron Microscope
Due to the high-voltage limitations of the transmission electron microscope, the electron beam can typically penetrate only thin samples with thicknesses below several tens of nanometers. With the exception of fine particulate samples, which can be prepared by the specimen dispersion method and directly mounted as droplets, the primary sample preparation techniques include physical thinning—such as ion milling and dual‑beam milling—and ultrathin sectioning. In most cases, physical thinning is employed for sample preparation. Preparing ultrathin sections involves a series of complex steps, including sample pretreatment, embedding, and sectioning, resulting in a relatively long processing time (approximately one week). As this instrument is a high‑resolution electron microscope, to ensure optimal performance and fully exploit its capability for high‑resolution imaging, it currently accepts samples primarily from the materials science domain.
15. X-ray powder diffractometer
The samples submitted for analysis may be in powder, bulk, thin‑film, or other forms. For powder samples, approximately 0.2 g is required (depending on the sample’s density and diffraction characteristics); bulk samples should have an approximately planar surface with dimensions no greater than 45 mm × 45 mm; thin‑film samples must have a specified thickness and a surface area of no more than 45 mm × 45 mm; for other sample types, please consult the testing personnel.
16. X-ray single-crystal diffractometer
The sample submitted for analysis should be a single crystal. When selecting a crystal, ensure that its surface is smooth and that its color and transparency are uniform. The crystal should be free of adhering micro‑crystals and exhibit no defects such as chipping, stacking faults, or cleavage fractures. The dimensions—length, width, and height—should each be between 0.1 and 0.4 mm, meaning the crystal’s diagonal length does not exceed 0.5 mm. (For larger crystals, sampling may be performed by cutting; for smaller crystals, their diffraction capabilities must be taken into account.)
17. Scanning Electron Microscope – X-ray Energy Dispersive Spectrometer
The sample submitted for analysis should be a dry solid; it may be in the form of blocks, flakes, fibers, granules, or powders. It must exhibit adequate chemical and physical stability, remaining non‑volatile and undeformed under vacuum and electron‑beam irradiation, and it must be non‑magnetic, non‑radioactive, and non‑corrosive. For biological soft‑tissue samples with high moisture content, users are required to perform preliminary treatments—such as fixation, washing, dehydration, and replacement with isopropyl acetate—prior to critical‑point drying, followed by critical‑point drying itself. For imaging studies, samples should be sputter‑coated with gold; for elemental analysis, a carbon coating is mandatory. In general, sample dimensions should not exceed 5 × 5 × 2 mm.
18. Electron Probe
Samples for quantitative analysis must be ground and polished to a smooth finish and thoroughly cleaned. If a sample cannot undergo surface grinding and polishing—since this would compromise analytical accuracy—this should be clearly stated in advance. For ease of testing and to conserve instrument time, samples should first be cut into thin sections; if cutting is not feasible, prior consultation with the analyst is required. Test points on the analysis surface should be marked beforehand; when no markings are provided, only representative, relatively flat areas should be selected for measurement. Liquid samples must be concentrated and dried prior to analysis. The samples submitted for analysis must be solid materials whose physical and chemical properties remain stable under high‑energy electron bombardment—non‑decomposing, non‑explosive, non‑volatile, non‑radioactive, and non‑magnetic. When submitting samples, it is advisable to indicate which elements they may contain. All samples must be coated with a carbon film.
19. Field-Emission Scanning Electron Microscope
The samples submitted for analysis may be in the form of dry solids, blocks, flakes, fibers, or powders. They should exhibit adequate chemical and physical stability, remaining non‑volatile and undeformed under vacuum conditions and during electron‑beam irradiation; they must also be non‑magnetic, non‑radioactive, and non‑corrosive. For biological soft tissues with high moisture content, users are required to perform preliminary treatments—such as fixation, washing, dehydration, and replacement with isopropyl acetate—prior to critical‑point drying, followed by the critical‑point drying step. Specimens intended for imaging should be sputter‑coated with a gold film beforehand. In general, samples should be as small as possible (dimensions ≤10 × 10 × 5 mm are preferable). Approximately 1 gram of powder is needed per sample. Nanoscale samples typically require ultrasonic dispersion and subsequent sputter coating with an ultra‑thin gold film.
The above provides an overview of the sample‑testing requirements for various ROHS 2.0 instruments. If you have any ROHS 2.0 instrument needs, please contact or consult us—Deps Instrument will promptly provide you with a quotation and detailed product information.
Latest News
Recommended Products
Frequently Asked Questions
What is RoHS?
RoHS is a mandatory standard established by EU legislation, officially known as the “Directive on the Restriction of the Use of Certain Hazardous Substances in Electrical and Electronic Equipment.” The standard entered into force on July 1, 2006, and primarily aims to regulate the materials and manufacturing processes of electrical and electronic products, thereby promoting human health and environmental protection.
The purpose of this standard is to eliminate six substances—lead, mercury, cadmium, hexavalent chromium, polybrominated biphenyls, and polybrominated diphenyl ethers (note: the correct Chinese term for PBDE is “polybrominated diphenyl ethers”; “polybrominated biphenyl ethers” is an incorrect designation)—and specifically stipulates that the cadmium content must not exceed 0.01%.
The latest RoHS Directive has added four new phthalate‑ester testing items: DEHP—di(2‑ethylhexyl) phthalate; BBP—butyl benzyl phthalate; DBP—dibutyl phthalate; and DIBP—diisobutyl phthalate.
What are the limit values for the ten hazardous substances under RoHS?
1. Cadmium: less than 100 ppm.
2. Lead: less than 1000 ppm.
3. Mercury: less than 1000 ppm.
4. Hexavalent chromium: less than 1000 ppm.
5. Polybrominated biphenyls (PBB): less than 1000 ppm.
6. Polybrominated diphenyl ethers (PBDEs): less than 1000 ppm.
7. Di(2-ethylhexyl) phthalate (DEHP): less than 1000 ppm.
8. Butyl benzyl phthalate (BBP): less than 1000 ppm.
9. Dibutyl phthalate (DBP): less than 1000 ppm.
10. Diisobutyl phthalate (DIBP): less than 1000 ppm.
What instruments are typically used for RoHS testing?
The commonly used analytical instrument for RoHS 1.0 is X-ray fluorescence spectroscopy (XRF). XRF analyzers are available in standalone, benchtop, and handheld configurations. Typically, a handheld XRF analyzer is employed first for portable on-site testing, with a focus on components of the product that pose a higher risk of containing restricted substances.
With the introduction of RoHS 2.0, four phthalate esters have been added to the RoHS Directive, necessitating additional testing to determine the levels of these compounds, which are extracted using solvents. The solvent extracts are then analyzed for the presence of phthalates by gas chromatography–mass spectrometry (GC/MS) or gas chromatography with flame ionization detection (GC/FID). Other elements and polybrominated diphenyl ethers/alkylated elastomers are analyzed using comparative methods and GC/MS, respectively.
Do the ten RoHS items require two instruments for testing?
The ten items specified in RoHS require two instruments for testing.
For the ten‑item RoHS testing, two instruments are required to complete the analysis; a single instrument cannot achieve this. Even if one instrument is used to measure heavy metal content, the other must be employed to analyze organic compounds.
X-ray fluorescence spectrometers are used to detect RoHS 1.0, while chromatographs, pyrolyzers, or hyphenated techniques such as GC–MS are employed to analyze the four phthalates specified in RoHS 2.0. At present, no single instrument can independently perform all the tests required for RoHS 2.0.
Can ROHS testing instruments detect metallic elements?
ROHS testing instruments are specialized devices used to detect the presence of hazardous substances in electronic components, household appliances, toys, and other products. These harmful substances can pose long-term risks to both human health and the environment, making the use of ROHS testing equipment of paramount importance.
Regarding whether ROHS testing instruments can detect metallic elements, the answer is yes. This is because ROHS testing covers not only hazardous substances but also the analysis of metallic constituents in compliant products. These metals include lead, cadmium, mercury, and others, which are widely used during the manufacturing process. Although these elements are essential, their excessive use poses significant risks to human health and the environment. Consequently, the advent of ROHS testing instruments has greatly facilitated the detection of metallic elements.