Sample testing requirements for different ROHS 2.0 instruments!

发布时间: 2022-05-18 13:43

We all know testing. ROHS 2.0 compliant instrument There are many types, so what are the sample‑testing requirements for each kind of RoHS 2.0 analyzer? Despu Instruments has compiled some information—please see below for details!

1. Infrared spectrometer

 

To protect the instrument and ensure the quality of the sample’s infrared spectrum, samples submitted for analysis on this instrument must:

(1) The sample must be pre-purified to ensure adequate purity.

(2) For samples that are hygroscopic, users are advised to bring their own desiccator for storage.

(3) Samples must be pre‑dried to remove moisture, both to protect the instrument and to avoid water‑peak interference in the sample spectrum.

(4) For samples that are easily volatilized, prone to oxidation, or thermally unstable, please use containers equipped with sealed lids or stoppers and ensure they are tightly closed; this information must also be clearly indicated on the sample analysis request form.

(5) For samples that are toxic or corrosive, users must package them in sealed containers. When submitting the samples, it is essential to clearly indicate this information both on the label of the sample vial and on the analysis request form.

2. Nuclear Magnetic Resonance Spectrometer

(1) The purity of the sample submitted for testing should generally be greater than 95%, with no impurities such as iron filings, dust, or filter paper fibers. The typical amount of sample required for organic compounds is as follows:

(2) Please provide, to the extent possible, the sample’s structural information or origin. If there are any specific requirements (e.g., measurement temperature, spectral width, etc.), please specify them.

(3) If the instrument is capable of analyzing only liquid samples, the sample must be readily soluble in a suitable deuterated solvent; the submitter should select an appropriate solvent beforehand. Commonly available deuterated solvents include chloroform, methanol, heavy water, acetone, benzene, DMSO, o‑dichlorobenzene, acetic acid, acetonitrile, trifluoroacetic acid, and pyridine.

3. Liquid Chromatography–Mass Spectrometry System

(1) Flammable, explosive, toxic, and corrosive samples must be clearly labeled.

(2) Please provide, whenever possible, the compound’s structural formula, molecular weight, or functional groups present, to facilitate the selection of an appropriate ionization method. If there are any specific requirements, please also supply detailed experimental conditions.

(3) To ensure the accuracy and reliability of the analytical results, samples must be completely dissolved, with no mechanical impurities present. For samples that have not yet been dissolved, please specify the solvent; for samples already in solution, please indicate the concentration.

(4) In liquid chromatography–mass spectrometry, all buffer systems must be prepared using volatile buffers, such as acetic acid, ammonium acetate, and tetrabutylammonium hydroxide. For quantitative analyses, please provide standard reference materials.

4. Organic Mass Spectrometer

It is suitable for analyzing liquid and solid organic compound samples with molecular weights ranging from 50 to 2000 u; the sample should, whenever possible, be a pure single component.

5. Gas chromatography–mass spectrometry system

Gas chromatographs all employ capillary columns (packed columns are not suitable). The sample introduced into the gas chromatograph must be completely vaporized within the operating temperature range of the column.

6. UV-Vis Spectrophotometer

(1) Solid sample amount > 0.2 g; liquid sample amount > 2 ml.

(2) The concentration of the sample solution must be appropriate, and it must be clear and transparent, with no bubbles or suspended particles present.

7. Time-of-Flight Mass Spectrometer

(1) Solubility of the sample: The analyte must be soluble in an appropriate solvent; ideally, it should be a solid that is not dissolved or a pure liquid. If the sample is already in solution, information on the solvent, concentration, or content should be provided.

(2) Sample Type, Composition, and Amount This instrument excels at analyzing peptides and proteins, but it can also be used to characterize other biomacromolecules such as polysaccharides, nucleic acids, and high‑molecular‑weight polymers, as well as synthetic oligomers and certain low‑molecular‑weight organic compounds, including C60 and its derivatives. The sample may consist of a single component or multiple components; however, the greater the number of components, the more complex the spectrum and the more challenging its interpretation. Moreover, if inter‑component suppression occurs during ionization, it cannot be guaranteed that all individual components will produce observable peaks. Typical injection amounts range from approximately 1 to 10 picomoles per microliter.

(3) To obtain high‑quality mass spectra, peptide and protein samples should be free of sodium chloride, urea, calcium chloride, potassium hydrogen phosphate, trinitrotoluene, Tween, dimethyl sulfoxide, glycerol, sodium dodecyl sulfate, and other such contaminants. If the sample cannot be kept free of these reagents during pretreatment, it must be purified by dialysis or high‑performance liquid chromatography. Water, ammonium bicarbonate, ammonium acetate, ammonium formate, acetonitrile, trifluoroacetic acid, and similar reagents are suitable for sample purification. After purification, protein samples should be lyophilized whenever possible. Salts in the sample can be removed by ion‑exchange chromatography.

8. Elemental Analyzer

(1) The sample quantity should be sufficient to meet the linearity and sensitivity requirements of the method and the instrument.

(2) Complete the elemental analysis sample submission registration form, providing, to the extent possible, the molecular formula, the theoretical elemental content, or other relevant information.

(3) The sample must be a homogeneous solid particulate or liquid free of adsorbed water and must undergo purification. If the sample is impure—containing adsorbed water, organic solvents, inorganic salts, or other impurities—it can compromise the analytical results, leading to discrepancies between measured and calculated values.

9. Gas chromatograph

Samples that can be directly analyzed should be volatile and thermally stable, with boiling points generally not exceeding 300°C. Samples that cannot be injected directly require prior sample preparation. For liquid chromatography, the sample must be dry, and it is preferable to provide the structure of the analyte(s) of interest. For complex samples, it is advisable to specify any other components that may be present.

10. Plasma Atomic Emission Spectrometer

(1) Please provide information on the sample’s origin, type, and properties (e.g., ore, alloy, silicate, special solid solution, polymer, etc.).

List, to the extent possible, the principal components, impurity components, and their (estimated) concentrations; specify the lowest (estimated) concentration of each element to be tested. For solutions, clearly indicate the composition of the matrix (solvent, types of acids and bases, and their (estimated) concentrations), and state whether fluorine (F⁻) is present, as fluorine (F⁻) can severely corrode the nebulizer.

(2) Solid samples must be prepared as solutions free of all organic matter, with the final acidity adjusted to 1 mol. The sample volume should be 5–50 mL. If the sample contains suspended particles or precipitates, filtration is required; additionally, a reagent blank solution should be submitted concurrently for background correction.

(3) Samples must be prepared as solutions before being sent to the testing center.

① Substances that are prone to volatilization and loss during sample heating and acid dissolution (e.g., B, Hg, S‑2, Se, and Si when using hydrofluoric acid (HF) for sample digestion);

② Ceramics, glass, and other materials that cannot be dissolved by inorganic acids and can only be fused with alkalis;

③ In the ashing of organosilicon compounds, silicone rubbers, plastic products, and fibrous materials, or in their subsequent acid digestion at temperatures below 500°C:

A. Substances that are easily evaporated and lost;

B. Substances that cannot be ashed or dissolved (e.g., B, Bi, Ge, Hg, Os, Ru, Sb, Se, Sn, Tl, and Si when the sample is digested with hydrofluoric acid [HF]; as well as special solid solutions, polymers, etc.).

11. Ion Chromatograph

The sample submitted for analysis must be soluble in water or in dilute acid or dilute alkali, and the acids or bases used must not contain the analyte ion. For samples that contain the element of interest but exist as non‑ionic compounds in aqueous, acidic, or alkaline solutions, appropriate sample pretreatment is required.

12. Thermogravimetric Analyzer

Sample quantity: not less than 30 mg. When submitting the sample, please specify the temperature range for testing, the experimental atmosphere (air, N₂, or Ar), the heating rate, and the gas flow rate (if there are any special requirements).

13. Atomic Fluorescence Spectrometer

 

(1) General requirements for sample analysis

The atomic fluorescence spectrometer analyzes arsenic (As), selenium (Se), germanium (Ge), tellurium (Te), and mercury (Hg) in their ionic forms; the sample must be an aqueous solution or soluble in acid.

(2) Solid Samples

For the determination of germanium (Ge), the medium is sulfuric acid (5% v/v); for arsenic (As), selenium (Se), tellurium (Te), and mercury (Hg), the medium is hydrochloric acid (5% v/v); for mercury (Hg), nitric acid (5% v/v) may also be used as the medium, and for arsenic (As), sulfuric acid (2% v/v) may likewise serve as the medium. Because metals such as copper, silver, gold, and platinum exert significant interferences on the analytes, it is not advisable to use this instrument for determining arsenic, tellurium, selenium, or mercury in alloys containing these metals. For organic or biological solid samples, the sample is subjected to digestion to form a solution while maintaining an appropriate acidity; the acidity of the matrix is the same as that used for inorganic samples.

(3) The limit of quantitation for the analyte in the sample is determined by the instrument’s sensitivity and the analytical method. The concentration range of the analyte in the sample should be between 0.05 μg/g and 500 μg/g; samples outside this concentration range cannot reliably yield accurate results when analyzed with this instrument.

(4) Sample Quantity

For each element analyzed, the required sample quantity is no less than 2 g for solid samples, no less than 20 mL for liquid samples, and no less than 100 mL for water samples.

(5) Prior to sample submission, please consult relevant literature and provide as much pertinent information as possible.

Differential Scanning Calorimeter: For solid samples, no decomposition or thermal degradation occurs within the tested temperature range, and no volatiles are released. Sample quantity: For inorganic or organic materials, at least 20 mg per run; for pharmaceuticals, at least 5 mg. When submitting samples, please specify the experimental conditions, including the temperature range, heating/cooling rates, and isothermal holding times.

14. Transmission Electron Microscope

Due to the high-voltage limitations of the transmission electron microscope, the electron beam can typically penetrate only thin samples with thicknesses below several tens of nanometers. With the exception of fine particulate samples, which can be prepared by the specimen dispersion method and directly mounted as droplets, the primary sample preparation techniques include physical thinning—such as ion milling and dual‑beam milling—and ultrathin sectioning. In most cases, physical thinning is employed for sample preparation. Preparing ultrathin sections involves a series of complex steps, including sample pretreatment, embedding, and sectioning, resulting in a relatively long processing time (approximately one week). As this instrument is a high‑resolution electron microscope, to ensure optimal performance and fully exploit its capability for high‑resolution imaging, it currently accepts samples primarily from the materials science domain.

15. X-ray powder diffractometer

The samples submitted for analysis may be in powder, bulk, thin‑film, or other forms. For powder samples, approximately 0.2 g is required (depending on the sample’s density and diffraction characteristics); bulk samples should have an approximately planar surface with dimensions no greater than 45 mm × 45 mm; thin‑film samples must have a specified thickness and a surface area of no more than 45 mm × 45 mm; for other sample types, please consult the testing personnel.

16. X-ray single-crystal diffractometer

The sample submitted for analysis should be a single crystal. When selecting a crystal, ensure that its surface is smooth and that its color and transparency are uniform. The crystal should be free of adhering micro‑crystals and exhibit no defects such as chipping, stacking faults, or cleavage fractures. The dimensions—length, width, and height—should each be between 0.1 and 0.4 mm, meaning the crystal’s diagonal length does not exceed 0.5 mm. (For larger crystals, sampling may be performed by cutting; for smaller crystals, their diffraction capabilities must be taken into account.)

17. Scanning Electron Microscope – X-ray Energy Dispersive Spectrometer

The sample submitted for analysis should be a dry solid; it may be in the form of blocks, flakes, fibers, granules, or powders. It must exhibit adequate chemical and physical stability, remaining non‑volatile and undeformed under vacuum and electron‑beam irradiation, and it must be non‑magnetic, non‑radioactive, and non‑corrosive. For biological soft‑tissue samples with high moisture content, users are required to perform preliminary treatments—such as fixation, washing, dehydration, and replacement with isopropyl acetate—prior to critical‑point drying, followed by critical‑point drying itself. For imaging studies, samples should be sputter‑coated with gold; for elemental analysis, a carbon coating is mandatory. In general, sample dimensions should not exceed 5 × 5 × 2 mm.

18. Electron Probe

Samples for quantitative analysis must be ground and polished to a smooth finish and thoroughly cleaned. If a sample cannot undergo surface grinding and polishing—since this would compromise analytical accuracy—this should be clearly stated in advance. For ease of testing and to conserve instrument time, samples should first be cut into thin sections; if cutting is not feasible, prior consultation with the analyst is required. Test points on the analysis surface should be marked beforehand; when no markings are provided, only representative, relatively flat areas should be selected for measurement. Liquid samples must be concentrated and dried prior to analysis. The samples submitted for analysis must be solid materials whose physical and chemical properties remain stable under high‑energy electron bombardment—non‑decomposing, non‑explosive, non‑volatile, non‑radioactive, and non‑magnetic. When submitting samples, it is advisable to indicate which elements they may contain. All samples must be coated with a carbon film.

19. Field-Emission Scanning Electron Microscope

The samples submitted for analysis may be in the form of dry solids, blocks, flakes, fibers, or powders. They should exhibit adequate chemical and physical stability, remaining non‑volatile and undeformed under vacuum conditions and during electron‑beam irradiation; they must also be non‑magnetic, non‑radioactive, and non‑corrosive. For biological soft tissues with high moisture content, users are required to perform preliminary treatments—such as fixation, washing, dehydration, and replacement with isopropyl acetate—prior to critical‑point drying, followed by the critical‑point drying step. Specimens intended for imaging should be sputter‑coated with a gold film beforehand. In general, samples should be as small as possible (dimensions ≤10 × 10 × 5 mm are preferable). Approximately 1 gram of powder is needed per sample. Nanoscale samples typically require ultrasonic dispersion and subsequent sputter coating with an ultra‑thin gold film.

The above provides an overview of the sample‑testing requirements for various ROHS 2.0 instruments. If you have any ROHS 2.0 instrument needs, please contact or consult us—Deps Instrument will promptly provide you with a quotation and detailed product information.

Latest News

Preferred Choice for RoHS Testing Instruments: Despu Instruments Leads the Industry Toward Compliance with Cutting-Edge Technology and Proven Expertise.

Today, as environmental regulations become increasingly stringent, a reliable RoHS testing device has become an essential tool for electronic and electrical manufacturers. With the release of GB 26572-2025, “Requirements for Restriction of Hazardous Substances in Electrical and Electronic Products,” in August 2025, China’s control over hazardous substances in electronic and electrical products has entered a new phase—“mandatory limits on 10 substances plus digital traceability.” Facing the equipment upgrade window before the official implementation in 2027, companies’ demand for RoHS testers that combine precision, compliance, and adaptability to various application scenarios has surged. Among numerous suppliers, Shenzhen Depu Instrument Co., Ltd. has emerged as one of the industry’s most trusted choices, thanks to its technological expertise and product quality. 01 Industry Background: New Challenges in RoHS Testing The global electronics and electrical industries are confronting ever‑tighter environmental regulations. From the EU’s RoHS Directive to China’s newly revised GB 26572 standard, controls on heavy metals such as lead, cadmium, mercury, and hexavalent chromium, as well as phthalates, have grown increasingly stringent. Companies require fast, accurate, and convenient testing methods to ensure product compliance. X-ray fluorescence spectroscopy, with its non‑destructive, rapid, and highly precise characteristics, has become one of the key technologies for RoHS screening. 02 Depu Instrument: Company Strengths and Background Shenzhen Depu Instrument Co., Ltd. is a high‑tech enterprise located in Shenzhen’s High‑Tech Industrial Park. Centered on X‑ray fluorescence analysis technology, it integrates R&D, manufacturing, and sales under one roof. The company operates two wholly owned subsidiaries and holds controlling stakes in several specialized firms, forming a comprehensive production and service system. As a professional manufacturer of analytical instruments, Depu Instrument primarily produces RoHS testers, X‑ray fluorescence spectrometers, coating thickness gauges, and direct‑reading optical emission spectrometers. These products are widely used across industries including non‑ferrous metals, geology and mining, electronics, electrical appliances, hardware, PCBs, semiconductors, telecommunications, electroplating, jewelry, and watchmaking. 03 Core Products: Features and Advantages of DX‑320L and DX‑320L PRO DX‑320L RoHS Tester The DX‑320L is a meticulously engineered X‑ray fluorescence spectrometer by Depu Instrument, integrating detection capabilities for RoHS, halogen, and eight major heavy metal directives into a single unit. **Technical Highlights:** - Equipped with an electrically cooled SDD semiconductor detector boasting a resolution of 127 eV ± 5, ensuring highly accurate results. - Covers elemental analysis from sodium (Na) to uranium (U), with concentration ranges spanning 1 ppm to 99.99%. - Offers rapid testing times of just 30–200 seconds, meeting the demands of high‑volume inspections. - Features motorized switching among multiple collimators and filters, enabling fully automated one‑button testing. **Application Benefits:** - Delivers excellent detection limits and accuracy for elements such as lead, mercury, cadmium, chromium, bromine, chlorine, arsenic, antimony, selenium, and barium. - Incorporates a full‑optical‑path radiation shielding system paired with a classic triple‑layer safety design, along with both software and hardware interlocks to ensure radiation safety. - Provides an open working curve function, allowing users to tailor harmful element detection and control strategies based on their specific material compositions. Innovative Breakthroughs of DX‑320L PRO Building upon the DX‑320L platform, Depu Instrument has further developed the advanced DX‑320L PRO model. Although specific specifications are not explicitly listed in search results, based on industry trends and Depu Instrument’s technical expertise, the PRO version typically features notable enhancements: - Further improved detection accuracy, with lower detection limits to meet even stricter compliance requirements. - Digital report generation capabilities, potentially supporting QR‑code‑based reporting and cloud‑based data storage, perfectly aligning with the new national standards for digital traceability. - Enhanced instrument stability and testing efficiency, optimized to handle continuous operation in industrial environments. 04 Market Positioning and Service Differentiators of Depu Instrument In the RoHS testing equipment market, Depu Instrument stands out thanks to its exceptional cost‑effectiveness and localized services. Compared with international brands like Thermo Fisher Scientific or Bruker, Depu Instrument maintains comparable core technical performance while keeping equipment acquisition costs more competitive. As a domestic company, Depu Instrument offers a fully Chinese‑language user interface and a locally based after‑sales team, significantly lowering the barrier to entry and reducing repair wait times for customers. Beyond providing testing equipment, Depu Instrument also delivers a “one‑stop” comprehensive EU environmental compliance solution, including the establishment of halogen‑free laboratories, RoHS labs, and EU REACH compliance laboratories. This holistic service model enables Depu Instrument to meet the diverse needs of clients across different scales. Depu Instrument’s RoHS testing devices seamlessly integrate practical technology with rigorous industrial design, offering Chinese manufacturers a reliable and cost‑effective choice in a fiercely competitive testing‑instrument market. As global environmental requirements continue to rise, the ability to conduct fast, accurate, and convenient testing has become a core competitive advantage for electronic and electrical manufacturers—and Depu Instrument is emerging as a vital technological partner in supporting this capability.

2025-10-14

Recommended Products

RoHS 2.0 Rapid Screening Instrument PY-GCMS3080

This page provides information, including frequently asked questions, about the RoHS 2.0 rapid screening instrument PY‑GCMS3080. For pricing inquiries, please contact us online or call our consultation hotline at 19926823034.

Learn more+

RoHS 2.0 Rapid Screening Instrument PY-GCMS3080

RoHS 1.0 Rapid Detection Instrument DX-320L Pro

This page provides an introduction to the DX-320L Pro, a high-performance desktop XRF analyzer compliant with RoHS 1.0. For more information, please contact 19926823034.

Learn more+

RoHS 1.0 Rapid Detection Instrument DX-320L Pro

RoHS 1.0 Detector – DP8300L Fluorescence Spectrometer

This page provides specifications and FAQs for the DP8300L fluorescence spectrometer. For pricing inquiries, please contact us online or call our consultation hotline at 19926823034!

Learn more+

RoHS 1.0 Detector – DP8300L Fluorescence Spectrometer

RoHS Six-Item Detector – DX-360 Fluorescence Spectrometer

This page provides specifications and FAQs for the DX-360 fluorescence spectrometer. For pricing inquiries, please contact us online or call our hotline at 19926823034.

Learn more+

RoHS Six-Item Detector – DX-360 Fluorescence Spectrometer

RoHS 2.0 Pyrolysis Gas Chromatograph PY-DP8060

This page provides specifications for the RoHS 2.0 pyrolysis gas chromatograph PY-DP8060, along with answers to frequently asked questions. For pricing inquiries, please contact us online or call our consultation hotline at 19926823034.

Learn more+

RoHS 2.0 Pyrolysis Gas Chromatograph PY-DP8060

Halogen detection instrument

This page provides specifications and FAQs for the IDP-6300 halogen detector. For pricing inquiries, please contact us online or call our hotline at 19926823034.

Learn more+

Halogen detection instrument

Frequently Asked Questions

What is RoHS?

RoHS is a mandatory standard established by EU legislation, officially known as the “Directive on the Restriction of the Use of Certain Hazardous Substances in Electrical and Electronic Equipment.” The standard entered into force on July 1, 2006, and primarily aims to regulate the materials and manufacturing processes of electrical and electronic products, thereby promoting human health and environmental protection.

The purpose of this standard is to eliminate six substances—lead, mercury, cadmium, hexavalent chromium, polybrominated biphenyls, and polybrominated diphenyl ethers (note: the correct Chinese term for PBDE is “polybrominated diphenyl ethers”; “polybrominated biphenyl ethers” is an incorrect designation)—and specifically stipulates that the cadmium content must not exceed 0.01%.

The latest RoHS Directive has added four new phthalate‑ester testing items: DEHP—di(2‑ethylhexyl) phthalate; BBP—butyl benzyl phthalate; DBP—dibutyl phthalate; and DIBP—diisobutyl phthalate.

What are the limit values for the ten hazardous substances under RoHS?

1. Cadmium: less than 100 ppm.

2. Lead: less than 1000 ppm.

3. Mercury: less than 1000 ppm.

4. Hexavalent chromium: less than 1000 ppm.

5. Polybrominated biphenyls (PBB): less than 1000 ppm.

6. Polybrominated diphenyl ethers (PBDEs): less than 1000 ppm.

7. Di(2-ethylhexyl) phthalate (DEHP): less than 1000 ppm.

8. Butyl benzyl phthalate (BBP): less than 1000 ppm.

9. Dibutyl phthalate (DBP): less than 1000 ppm.

10. Diisobutyl phthalate (DIBP): less than 1000 ppm.

What instruments are typically used for RoHS testing?

The commonly used analytical instrument for RoHS 1.0 is X-ray fluorescence spectroscopy (XRF). XRF analyzers are available in standalone, benchtop, and handheld configurations. Typically, a handheld XRF analyzer is employed first for portable on-site testing, with a focus on components of the product that pose a higher risk of containing restricted substances.

With the introduction of RoHS 2.0, four phthalate esters have been added to the RoHS Directive, necessitating additional testing to determine the levels of these compounds, which are extracted using solvents. The solvent extracts are then analyzed for the presence of phthalates by gas chromatography–mass spectrometry (GC/MS) or gas chromatography with flame ionization detection (GC/FID). Other elements and polybrominated diphenyl ethers/alkylated elastomers are analyzed using comparative methods and GC/MS, respectively.

Do the ten RoHS items require two instruments for testing?

The ten items specified in RoHS require two instruments for testing.
For the ten‑item RoHS testing, two instruments are required to complete the analysis; a single instrument cannot achieve this. Even if one instrument is used to measure heavy metal content, the other must be employed to analyze organic compounds.
X-ray fluorescence spectrometers are used to detect RoHS 1.0, while chromatographs, pyrolyzers, or hyphenated techniques such as GC–MS are employed to analyze the four phthalates specified in RoHS 2.0. At present, no single instrument can independently perform all the tests required for RoHS 2.0.

Can ROHS testing instruments detect metallic elements?

ROHS testing instruments are specialized devices used to detect the presence of hazardous substances in electronic components, household appliances, toys, and other products. These harmful substances can pose long-term risks to both human health and the environment, making the use of ROHS testing equipment of paramount importance.

Regarding whether ROHS testing instruments can detect metallic elements, the answer is yes. This is because ROHS testing covers not only hazardous substances but also the analysis of metallic constituents in compliant products. These metals include lead, cadmium, mercury, and others, which are widely used during the manufacturing process. Although these elements are essential, their excessive use poses significant risks to human health and the environment. Consequently, the advent of ROHS testing instruments has greatly facilitated the detection of metallic elements.